Difference between revisions of "PAA CoincidenceCounting"
Jump to navigation
Jump to search
Line 11: | Line 11: | ||
==First pict with ADC gate== | ==First pict with ADC gate== | ||
+ | |||
+ | Two HpGe detectors were used to detect coincidence gammas from Co-60 on 6/14/2013 | ||
+ | |||
[[File:A061413_CPAA_coin.png | 200 px]][[File:B061413_CPAA_coin.png | 200 px]][[File:C061413_CPAA_coin.png | 200 px]] | [[File:A061413_CPAA_coin.png | 200 px]][[File:B061413_CPAA_coin.png | 200 px]][[File:C061413_CPAA_coin.png | 200 px]] |
Revision as of 16:43, 17 June 2013
ADC non-linearity
Canberra Model 9615 < 0.04% (integral) over 2usec
Canberra Model 9633 < 0.025% (integral) < 0.9% (differential)
CAEN V792 < 0.1% (integral)
Coinc Scope pictures
First pict with ADC gate
Two HpGe detectors were used to detect coincidence gammas from Co-60 on 6/14/2013