Difference between revisions of "R1S6 SL1 BurnIn"
Jump to navigation
Jump to search
(40 intermediate revisions by 2 users not shown) | |||
Line 194: | Line 194: | ||
|- | |- | ||
| 04/30/13 || 09:54|| 536 || 1260 || -750 || .1 || .2|| | | 04/30/13 || 09:54|| 536 || 1260 || -750 || .1 || .2|| | ||
+ | |- | ||
+ | | 04/30/13 || 10:44|| 536 || 1270 || -750 || .2 || .3|| | ||
+ | |- | ||
+ | | 04/30/13 || 11:43|| 536 || 1270 || -750 || .1 || .2|| | ||
+ | |- | ||
+ | | 04/30/13 || 11:43|| 536 || 1270 || -750 || .1 || .2|| | ||
+ | |- | ||
+ | | 04/30/13 || 17:53|| 536 || 1270 || -750 || .07 || .155|| | ||
+ | |- | ||
+ | | 05/01/13 || 10:33|| 536 || 1270 || -750 || 0.000 || 0.095|| | ||
+ | |- | ||
+ | | 05/01/13 || 10:54|| 536 || 1280 || -750 || 0.01 || 0.1|| | ||
+ | |- | ||
+ | | 05/01/13 || 11:38|| 536 || 1290 || -750 || 0.02 || 0.1|| | ||
+ | |- | ||
+ | | 05/01/13 || 16:35|| 536 || 1290 || -750 || 0.01 || 0.110|| | ||
+ | |- | ||
+ | | 05/02/13 || 09:27|| 536 || 1290 || -750 || 0.00 || 0.08|| | ||
+ | |- | ||
+ | | 05/02/13 || 09:33|| 536 || 1310 || -750 || 0.00 || 0.1|| | ||
+ | |- | ||
+ | | 05/02/13 || 14:10|| 536 || 1310 || -750 || 0.00 || 0.1|| | ||
+ | |- | ||
+ | | 05/02/13 || 14:30|| 536 || 1320 || -750 || 0.010 || 0.105|| | ||
+ | |- | ||
+ | | 05/02/13 || 14:34|| 536 || 1320 || -750 || 0.015 || 0.115|| | ||
+ | |- | ||
+ | | 05/02/13 || 14:56|| 536 || 1320 || -750 || 0.010 || 0.110|| | ||
+ | |- | ||
+ | | 05/02/13 || 14:58|| 536 || 1330 || -750 || 0.035 || 0.125|| Current is fluctuating by around 50% of this measurement. | ||
+ | |- | ||
+ | | 05/02/13 || 15:12|| 536 || 1330 || -750 || 0.030 || 0.125|| Current reading appears stable within 5% over a 1 minute period. | ||
+ | |- | ||
+ | | 05/02/13 || 15:12|| 536 || 1320 || -750 || 0.020 || 0.115|| Turned down voltage for the weekend. | ||
+ | |- | ||
+ | | 05/03/13 || 10:45|| 536 || 1300 || -750 || 0.00 || 0.08|| Found HV for Field wires in tripped state, lowered Sense voltage | ||
+ | |- | ||
+ | | 05/04/13 || 10:46|| 536 || 1300 || -750 || 0.00 || 0.06|| Found HV for Field wires in tripped state, lowered Sense voltage | ||
+ | |- | ||
+ | | 05/04/13 || 10:47|| 536 || 1320 || -750 || 0.00 || 0.07|| Found HV for Field wires in tripped state, lowered Sense voltage | ||
+ | |- | ||
+ | | 05/06/13 || 07:35|| 536 || 1320 || -750 || 0.00 || 0.07|| | ||
+ | |- | ||
+ | | 05/06/13 || 07:37|| 536 || 1280 || -750 || 0.00 || 0.07|| HV 110 voltage was interrupted and power supply was reset | ||
+ | |- | ||
+ | | 05/06/13 || 11:19|| 536 || 1350 || -750 || 0.04 || 0.13|| | ||
+ | |- | ||
+ | | 05/06/13 || 13:38|| 536 || 1350 || -750 || 0.07 || 0.16|| | ||
+ | |- | ||
+ | | 05/07/13 || 07:37|| 536 || 1360 || -750 || 0.1 || 0.2|| | ||
+ | |- | ||
+ | | 05/07/13 || 09:55|| 536 || 1360 || -750 || 0.13 || 0.22|| | ||
+ | |- | ||
+ | | 05/08/13 || 09:01|| 536 || 1360 || -750 || 0.04 || 0.13|| | ||
+ | |- | ||
+ | | 05/08/13 || 09:01|| 536 || 1380 || -750 || 0.1 || 0.2|| | ||
+ | |- | ||
+ | | 05/08/13 || 12:45|| 536 || 1380 || -750 || 0.12 || 0.22|| | ||
+ | |- | ||
+ | | 05/08/13 || 17:31|| 536 || 1380 || -750 || 0.25 || 0.34|| | ||
+ | |- | ||
+ | | 05/09/13 || 17:20|| 536 || 1380 || -750 || 0.2 || 0.3|| | ||
+ | |- | ||
+ | | 05/09/13 || 17:21|| 536 || 1500 || -750 || 2.3 || 2.4|| pictures have been taken, lets see if voltage will hold or trip | ||
+ | |- | ||
+ | | 05/09/13 || 17:26|| 536 || 1500 || -750 || 2.8 || 2.9|| no current spikes but the current did increase | ||
+ | |- | ||
+ | | 05/09/13 || 17:31|| 536 || 1500 || -750 || 2.7 || 2.8|| | ||
+ | |- | ||
+ | | 05/09/13 || 17:52|| 536 || 1500 || -750 || 2.4 || 2.5|| | ||
+ | |- | ||
+ | | 05/13/13 || 07:43|| 536 || 1500 || -750 || 0.67 || 0.77|| | ||
+ | |- | ||
+ | | 05/14/13 || 07:38|| 536 || 1500 || -750 || 0.35 || 0.44|| | ||
+ | |- |