Line 144: | Line 144: | ||
|- | |- | ||
| 04/25/13 || 12:33|| 536 || 1160 || -750 || 0.000 || 0.055 || | | 04/25/13 || 12:33|| 536 || 1160 || -750 || 0.000 || 0.055 || | ||
+ | |- | ||
+ | | 04/25/13 || 16:30|| 536 || 1160 || -750 || 0.000 || 0.055 || | ||
+ | |- | ||
+ | | 04/25/13 || 16:35|| 536 || 1190 || -750 || 3.2 || 3.3 || Unstable current started when going up to 1200V. Averaged 3-13 microamps at 1200V. Average reading is closer to 3 at 1190V. | ||
+ | |- | ||
+ | | 04/25/13 || 16:35|| 536 || 1170 || -750 || 10 || 10 || Current went up to 12-20 microamps at 1190V after a couple minutes. Scaled back to 1170V. | ||
+ | |- | ||
+ | | 04/25/13 || 16:40|| 536 || 1180 || -750 || 0.000 || 0.080 || Current stabilized to almost 0 at 1170. Went back up to 1180. Slight rise in field current from 1160V. | ||
+ | |- | ||
+ | | 04/25/13 || 16:55|| 536 || 1190 || -750 || 0.000 || 0.095 || Temporary spiking above 0.095 on fields. Average reading is stable. | ||
+ | |- | ||
+ | | 04/25/13 || 17:07|| 536 || 1190 || -750 || 2.1 || 2.2 || Average reading is higher now. | ||
+ | |- | ||
+ | | 04/25/13 || 17:20|| 536 || 1190 || -750 || 10 || 10 || Average reading got higher, appeared unstable - was more around 15-16, occasional spikes to 20 and 70 microamps. Seems to be reaching a lower average now. | ||
+ | |- | ||
+ | | 04/25/13 || 17:23|| 536 || 1190 || -750 || 0.000 || 0.080 || Average reading is now stable. No observed spikes for a couple minutes. | ||
+ | |- | ||
+ | | 04/25/13 || 17:30|| 536 || 1190 || -750 || 0.000 || 0.075 || | ||
+ | |- | ||
+ | | 04/25/13 || 17:32|| 536 || 1195 || -750 || 0.000 || 0.080 || | ||
+ | |- | ||
+ | | 04/25/13 || 17:39|| 536 || 1195 || -750 || 0.000 || 0.085 || | ||
+ | |- | ||
+ | | 04/25/13 || 17:41|| 536 || 1200 || -750 || 9-30 || 9-30 || Reading is unstable, average ranges from 9-30 microamps. | ||
+ | |- | ||
+ | | 04/25/13 || 17:47|| 536 || 1200 || -750 || 0.000 || 0.085 || Reading has been stable for 3 minutes at 0.085-0.090 on fields. | ||
+ | |- | ||
+ | | 04/25/13 || 17:50|| 536 || 1205 || -750 || 0.000 || 0.090 || | ||
+ | |- | ||
+ | | 04/25/13 || 17:56|| 536 || 1205 || -750 || 0.000 || 0.090 || Reading has been stable for 6 minutes. Leaving at this voltage for the night. | ||
+ | |- | ||
+ | | 04/26/13 || 10:05|| 536 || 1240 || -750 || 0.000 || 0.09 || current was 0.5 , increased voltage from 1205 to 1220 in 3, 5 volt steps and 1V/s increases | ||
+ | |- | ||
+ | | 04/26/13 || 16:02|| 536 || 1230 || -750 || 0.000 || 0.09 || the HV did not trip but the current was fluctuating up to 2 muA so I lowered the HV to 1230 and the current went down to 0.09 on Field. | ||
+ | |- | ||
+ | | 04/27/13 || 15:40|| 536 || 1240 || -750 || 0.000 || 0.07 || I started seeing what may be signals on the scope at this HV but they look too narrow [[File:042713_S6_1240HV.png| 50 px]] | ||
+ | |- | ||
+ | | 04/29/13 || 07:40|| 536 || 1250 || -750 || 0.000 || 0.07 || HV stayed up 1240 over the weekend, increase another 10 volts. Signal rate went up and saw wider pulses[[File:042913_S6_1250HV.png| 50 px]] | ||
+ | |- | ||
+ | | 04/29/13 || 09:19|| 536 || 1260 || -750 || 0.000 || 0.08 || | ||
+ | |- | ||
+ | | 04/29/13 || 10:54|| 536 || 1270 || -750 || 0.000 || 0.09 || | ||
+ | |- | ||
+ | | 04/29/13 || 16:56|| 536 || 1500 || -750 || 11.30 || 11.4 || had to turn off power supply for a bit to work on soldering HVB-4, I tried to increase voltage to 1500 first to check if it would trip at 50 muA, The current was 11 uA and no current spike, lets see how long it will stay here. | ||
+ | |- | ||
+ | | 04/29/13 || 17:41|| 536 || 1500 || -750 || 9.8 || 9.9 || Guard Current 0muA, Field Current 9.55muA, Sense Current 9.43muA. | ||
+ | |- | ||
+ | | 04/30/13 || 07:43|| 536 || 1240 || -750 || 3 || 3|| Field HV was in tripped state, current fluctuations are large at 1240. | ||
+ | |- | ||
+ | | 04/30/13 || 09:54|| 536 || 1260 || -750 || .1 || .2|| | ||
+ | |- | ||
+ | | 04/30/13 || 10:44|| 536 || 1270 || -750 || .2 || .3|| | ||
+ | |- | ||
+ | | 04/30/13 || 11:43|| 536 || 1270 || -750 || .1 || .2|| | ||
+ | |- | ||
+ | | 04/30/13 || 11:43|| 536 || 1270 || -750 || .1 || .2|| | ||
+ | |- | ||
+ | | 04/30/13 || 17:53|| 536 || 1270 || -750 || .07 || .155|| | ||
+ | |- | ||
+ | | 05/01/13 || 10:33|| 536 || 1270 || -750 || 0.000 || 0.095|| | ||
+ | |- | ||
+ | | 05/01/13 || 10:54|| 536 || 1280 || -750 || 0.01 || 0.1|| | ||
+ | |- | ||
+ | | 05/01/13 || 11:38|| 536 || 1290 || -750 || 0.02 || 0.1|| | ||
+ | |- | ||
+ | | 05/01/13 || 16:35|| 536 || 1290 || -750 || 0.01 || 0.110|| | ||
+ | |- | ||
+ | | 05/02/13 || 09:27|| 536 || 1290 || -750 || 0.00 || 0.08|| | ||
+ | |- | ||
+ | | 05/02/13 || 09:33|| 536 || 1310 || -750 || 0.00 || 0.1|| | ||
+ | |- | ||
+ | | 05/02/13 || 14:10|| 536 || 1310 || -750 || 0.00 || 0.1|| | ||
+ | |- | ||
+ | | 05/02/13 || 14:30|| 536 || 1320 || -750 || 0.010 || 0.105|| | ||
+ | |- | ||
+ | | 05/02/13 || 14:34|| 536 || 1320 || -750 || 0.015 || 0.115|| | ||
+ | |- | ||
+ | | 05/02/13 || 14:56|| 536 || 1320 || -750 || 0.010 || 0.110|| | ||
+ | |- | ||
+ | | 05/02/13 || 14:58|| 536 || 1330 || -750 || 0.035 || 0.125|| Current is fluctuating by around 50% of this measurement. | ||
+ | |- | ||
+ | | 05/02/13 || 15:12|| 536 || 1330 || -750 || 0.030 || 0.125|| Current reading appears stable within 5% over a 1 minute period. | ||
+ | |- | ||
+ | | 05/02/13 || 15:12|| 536 || 1320 || -750 || 0.020 || 0.115|| Turned down voltage for the weekend. | ||
+ | |- | ||
+ | | 05/03/13 || 10:45|| 536 || 1300 || -750 || 0.00 || 0.08|| Found HV for Field wires in tripped state, lowered Sense voltage | ||
+ | |- | ||
+ | | 05/04/13 || 10:46|| 536 || 1300 || -750 || 0.00 || 0.06|| Found HV for Field wires in tripped state, lowered Sense voltage | ||
+ | |- | ||
+ | | 05/04/13 || 10:47|| 536 || 1320 || -750 || 0.00 || 0.07|| Found HV for Field wires in tripped state, lowered Sense voltage | ||
+ | |- | ||
+ | | 05/06/13 || 07:35|| 536 || 1320 || -750 || 0.00 || 0.07|| | ||
+ | |- | ||
+ | | 05/06/13 || 07:37|| 536 || 1280 || -750 || 0.00 || 0.07|| HV 110 voltage was interrupted and power supply was reset | ||
+ | |- | ||
+ | | 05/06/13 || 11:19|| 536 || 1350 || -750 || 0.04 || 0.13|| | ||
+ | |- | ||
+ | | 05/06/13 || 13:38|| 536 || 1350 || -750 || 0.07 || 0.16|| | ||
+ | |- | ||
+ | | 05/07/13 || 07:37|| 536 || 1360 || -750 || 0.1 || 0.2|| | ||
+ | |- | ||
+ | | 05/07/13 || 09:55|| 536 || 1360 || -750 || 0.13 || 0.22|| | ||
+ | |- | ||
+ | | 05/08/13 || 09:01|| 536 || 1360 || -750 || 0.04 || 0.13|| | ||
+ | |- | ||
+ | | 05/08/13 || 09:01|| 536 || 1380 || -750 || 0.1 || 0.2|| | ||
+ | |- | ||
+ | | 05/08/13 || 12:45|| 536 || 1380 || -750 || 0.12 || 0.22|| | ||
+ | |- | ||
+ | | 05/08/13 || 17:31|| 536 || 1380 || -750 || 0.25 || 0.34|| | ||
+ | |- | ||
+ | | 05/09/13 || 17:20|| 536 || 1380 || -750 || 0.2 || 0.3|| | ||
+ | |- | ||
+ | | 05/09/13 || 17:21|| 536 || 1500 || -750 || 2.3 || 2.4|| pictures have been taken, lets see if voltage will hold or trip | ||
+ | |- | ||
+ | | 05/09/13 || 17:26|| 536 || 1500 || -750 || 2.8 || 2.9|| no current spikes but the current did increase | ||
+ | |- | ||
+ | | 05/09/13 || 17:31|| 536 || 1500 || -750 || 2.7 || 2.8|| | ||
+ | |- | ||
+ | | 05/09/13 || 17:52|| 536 || 1500 || -750 || 2.4 || 2.5|| | ||
+ | |- | ||
+ | | 05/13/13 || 07:43|| 536 || 1500 || -750 || 0.67 || 0.77|| | ||
+ | |- | ||
+ | | 05/14/13 || 07:38|| 536 || 1500 || -750 || 0.35 || 0.44|| | ||
+ | |- |