Difference between revisions of "R1S6 SL1 BurnIn"
Jump to navigation
Jump to search
(125 intermediate revisions by 3 users not shown) | |||
Line 52: | Line 52: | ||
|- | |- | ||
| 04/10/13 || 12:30 || 536 || 1050 || -750 || 0.2 || 0.3 || | | 04/10/13 || 12:30 || 536 || 1050 || -750 || 0.2 || 0.3 || | ||
+ | |- | ||
+ | | 04/10/13 || 13:45 || 536 || 1070 || -750 || 0.15 || 0.25 || Current instability was observed when taking voltage up to 1075. Backed off to 1040, current stabilized to less than 0.01 <math>\mu</math>A. Raised back up to 1070V again, current appears stable for the moment. | ||
+ | |- | ||
+ | | 04/10/13 || 15:15 || 536 || 1070 || -750 || 0.065 || 0.115 || | ||
+ | |- | ||
+ | | 04/11/13 || 08:30 || 536 || 1120 || -750 || 0.065 || 0.115 || Currently raising voltage only 1-3 volts at a time and monitoring sense current for a few seconds before raising again. If sense current stays above 0.075 for more than a few seconds, run-away current instability will usually start-up again. | ||
+ | |- | ||
+ | | 04/11/13 || 10:32 || 536 || 1130 || -750 || 0.080 || 0.170 || | ||
+ | |- | ||
+ | | 04/11/13 || 11:25 || 536 || 1130 || -750 || 0.080 || 0.170 ||HV tripped at about this time | ||
+ | |- | ||
+ | | 04/11/13 || 11:25 || 536 || 1120 || -750 || 0.080 || 0.170 || | ||
+ | |- | ||
+ | | 04/15/13 || 07:34 || 536 || 1068 || -750 || 0.080 || 0.170 || Left it this way over the weekend | ||
+ | |- | ||
+ | | 04/15/13 || 07:40 || 536 || 1110 || -750 || 0.000 || 0.080 || | ||
+ | |- | ||
+ | | 04/15/13 || 08:40 || 536 || 1130 || -750 || 0.000 || 0.090 || | ||
+ | |- | ||
+ | | 04/15/13 || 09:08 || 536 || 1140 || -750 || 0.000 || 0.090 || | ||
+ | |- | ||
+ | | 04/15/13 || 10:05 || 536 || 1150 || -750 || 0.000 || 0.090 || | ||
+ | |- | ||
+ | | 04/15/13 || 10:35 || 536 || 1150 || -750 || 0.005 || 0.100 || | ||
+ | |- | ||
+ | | 04/15/13 || 12:50|| 536 || 1160 || -750 || 0.015 || 0.10 || | ||
+ | |- | ||
+ | | 04/15/13 || 15:27|| 536 || 1170 || -750 || 0.020 || 0.12 || | ||
+ | |- | ||
+ | | 04/16/13 || 07:39|| 536 || 1195 || -750 || 0.005 || 0.100 || | ||
+ | |- | ||
+ | | 04/16/13 || 08:56|| 536 || 1195 || -750 || 0.01 || 0.100 || | ||
+ | |- | ||
+ | | 04/16/13 || 13:37|| 536 || 1195 || -750 || 0.00 || 0.090 || | ||
+ | |- | ||
+ | | 04/16/13 || 13:40|| 536 || 1205 || -750 || 0.005 || 0.100 || | ||
+ | |- | ||
+ | | 04/16/13 || 16:50|| 536 || 1205 || -750 || 5.0 || 5.1 || Current is fluctuating around 5, but not entirely stable. No observed large spikes above 6. | ||
+ | |- | ||
+ | | 04/17/13 || 07:35|| 536 || 1205 || -750 || 0 || 0 || found in tripped state. | ||
+ | |- | ||
+ | | 04/17/13 || 07:35|| 536 || 1020 || -750 || 3 || 3 || during rise to 1175 V, current appeared unstable so I backed off to 1020V where it oscillated between 3 and <1. Letting it settle here before ramping back up to 1205. | ||
+ | |- | ||
+ | | 04/17/13 || 12:46|| 536 || 1070 || -750 || 0.1 || 0.2 || increased Senve voltage by 20 volts until current fluctuations were seen on the field current | ||
+ | |- | ||
+ | | 04/17/13 || 14:48|| 536 || 1090 || -750 || 0.01 || 0.1 || | ||
+ | |- | ||
+ | | 04/17/13 || 14:48|| 536 || 1110 || -750 || 0.01 || 0.1 ||some 5uA current spikes are occurring, leave at this voltage overnight. | ||
+ | |- | ||
+ | | 04/18/13 || 10:50|| 536 || 1160 || -750 || 0.080 || 0.170 || Noticed some possible spiking, getting close to trip current at 1170, scaled back to 1160 and current stabilized within a minute. | ||
+ | |- | ||
+ | | 04/18/13 || 12:04|| 536 || 1170 || -750 || 0.080 || 0.170 || | ||
+ | |- | ||
+ | | 04/18/13 || 13:46|| 536 || 1180 || -750 || 0.080 || 0.170 || | ||
+ | |- | ||
+ | | 04/19/13 || 15:28|| 536 || 1020 || -750 || 0.080 || 0.170 || Found HV in tripped state, wouldn't restore to 1100 Volts without tripping so set to 1020 for weekend, current spikes began around 1050 volts | ||
+ | |- | ||
+ | | 04/19/13 || 20:20|| 536 || 1000 || -750 || 0.080 || 0.170 || Found HV in tripped state, restored to 1000 | ||
+ | |- | ||
+ | | 04/22/13 || 07:30|| 536 || 1000 || -750 || 0.000 || 0.035 || | ||
+ | |- | ||
+ | | 04/22/13 || 07:35|| 536 || 1100 || -750 || 0.000 || 0.045 || | ||
+ | |- | ||
+ | | 04/22/13 || 10:40|| 536 || 1150 || -750 || 0.000 || 0.050 || | ||
+ | |- | ||
+ | | 04/22/13 || 13:30|| 536 || 1150 || -750 || 0.000 || 0.045 || | ||
+ | |- | ||
+ | | 04/22/13 || 13:35|| 536 || 1200 || -750 || 0.000 || 0.070 || | ||
+ | |- | ||
+ | | 04/22/13 || 17:40|| 536 || 1140 || -750 || 0.000 || 0.050 || HV was found in a tripped state a few hours ago. Voltage was ramped up to 1050 a few minutes ago. Just ramped up to 1140 now. Current appears stable at 0.050 on field. | ||
+ | |- | ||
+ | | 04/23/13 || 07:30|| 536 || 1140 || -750 || 0.000 || 0.040 || | ||
+ | |- | ||
+ | | 04/23/13 || 07:35|| 536 || 1170 || -750 || 0.000 || 0.050 || | ||
+ | |- | ||
+ | | 04/23/13 || 10:00|| 536 || 1140 || -750 || 0.000 || 0.050 || Found in tripped state. Reset to previous stable voltage. Some current spiking from 0 - 3, sometimes up to 140 microamps for a split second in the first few minutes after reset. | ||
+ | |- | ||
+ | | 04/23/13 || 10:00|| 536 || 1150 || -750 || 0.000 || 0.045 || | ||
+ | |- | ||
+ | | 04/23/13 || 11:31|| 536 || 1180 || -750 || 0.000 || 0.045 || | ||
+ | |- | ||
+ | | 04/24/13 || 17:43|| 536 || 1400 || -750 || 6.2 || 6.3 || Kraken was removed from SL1 and placed on SL2 to test up to full operating voltage on a burned in SL. SL2 averaged 1.2 microamps after 24 hours of full voltage on with the Kraken. moved Kraken back to SL1. Able to get voltage up much higher now, and current spiking does not occur quite as readily to high levels (e.g., 140 microamps with an average reading of 3 or less microamps). Current trip level is set to 200 microamps on all wires. | ||
+ | |- | ||
+ | | 04/24/13 || 17:53|| 536 || 1050 || -750 || 3 || 3 || Found fields in tripped state. Reset and turned down sense voltage to 1050. Still observing some current instability a few minutes after setting the voltage to 1150 initially and turning down to 1050. | ||
+ | |- | ||
+ | | 04/25/13 || 07:35|| 536 || 1050 || -750 || 0.000 || 0.040 || | ||
+ | |- | ||
+ | | 04/25/13 || 07:38|| 536 || 1100 || -750 || 3 || 3 || Current instability started when taking voltage up to 1150. Dropped back to 1100, current is still somewhat high. | ||
+ | |- | ||
+ | | 04/25/13 || 11:00|| 536 || 1150 || -750 || 0.000 || 0.055 || | ||
+ | |- | ||
+ | | 04/25/13 || 12:33|| 536 || 1160 || -750 || 0.000 || 0.055 || | ||
+ | |- | ||
+ | | 04/25/13 || 16:30|| 536 || 1160 || -750 || 0.000 || 0.055 || | ||
+ | |- | ||
+ | | 04/25/13 || 16:35|| 536 || 1190 || -750 || 3.2 || 3.3 || Unstable current started when going up to 1200V. Averaged 3-13 microamps at 1200V. Average reading is closer to 3 at 1190V. | ||
+ | |- | ||
+ | | 04/25/13 || 16:35|| 536 || 1170 || -750 || 10 || 10 || Current went up to 12-20 microamps at 1190V after a couple minutes. Scaled back to 1170V. | ||
+ | |- | ||
+ | | 04/25/13 || 16:40|| 536 || 1180 || -750 || 0.000 || 0.080 || Current stabilized to almost 0 at 1170. Went back up to 1180. Slight rise in field current from 1160V. | ||
+ | |- | ||
+ | | 04/25/13 || 16:55|| 536 || 1190 || -750 || 0.000 || 0.095 || Temporary spiking above 0.095 on fields. Average reading is stable. | ||
+ | |- | ||
+ | | 04/25/13 || 17:07|| 536 || 1190 || -750 || 2.1 || 2.2 || Average reading is higher now. | ||
+ | |- | ||
+ | | 04/25/13 || 17:20|| 536 || 1190 || -750 || 10 || 10 || Average reading got higher, appeared unstable - was more around 15-16, occasional spikes to 20 and 70 microamps. Seems to be reaching a lower average now. | ||
+ | |- | ||
+ | | 04/25/13 || 17:23|| 536 || 1190 || -750 || 0.000 || 0.080 || Average reading is now stable. No observed spikes for a couple minutes. | ||
+ | |- | ||
+ | | 04/25/13 || 17:30|| 536 || 1190 || -750 || 0.000 || 0.075 || | ||
+ | |- | ||
+ | | 04/25/13 || 17:32|| 536 || 1195 || -750 || 0.000 || 0.080 || | ||
+ | |- | ||
+ | | 04/25/13 || 17:39|| 536 || 1195 || -750 || 0.000 || 0.085 || | ||
+ | |- | ||
+ | | 04/25/13 || 17:41|| 536 || 1200 || -750 || 9-30 || 9-30 || Reading is unstable, average ranges from 9-30 microamps. | ||
+ | |- | ||
+ | | 04/25/13 || 17:47|| 536 || 1200 || -750 || 0.000 || 0.085 || Reading has been stable for 3 minutes at 0.085-0.090 on fields. | ||
+ | |- | ||
+ | | 04/25/13 || 17:50|| 536 || 1205 || -750 || 0.000 || 0.090 || | ||
+ | |- | ||
+ | | 04/25/13 || 17:56|| 536 || 1205 || -750 || 0.000 || 0.090 || Reading has been stable for 6 minutes. Leaving at this voltage for the night. | ||
+ | |- | ||
+ | | 04/26/13 || 10:05|| 536 || 1240 || -750 || 0.000 || 0.09 || current was 0.5 , increased voltage from 1205 to 1220 in 3, 5 volt steps and 1V/s increases | ||
+ | |- | ||
+ | | 04/26/13 || 16:02|| 536 || 1230 || -750 || 0.000 || 0.09 || the HV did not trip but the current was fluctuating up to 2 muA so I lowered the HV to 1230 and the current went down to 0.09 on Field. | ||
+ | |- | ||
+ | | 04/27/13 || 15:40|| 536 || 1240 || -750 || 0.000 || 0.07 || I started seeing what may be signals on the scope at this HV but they look too narrow [[File:042713_S6_1240HV.png| 50 px]] | ||
+ | |- | ||
+ | | 04/29/13 || 07:40|| 536 || 1250 || -750 || 0.000 || 0.07 || HV stayed up 1240 over the weekend, increase another 10 volts. Signal rate went up and saw wider pulses[[File:042913_S6_1250HV.png| 50 px]] | ||
+ | |- | ||
+ | | 04/29/13 || 09:19|| 536 || 1260 || -750 || 0.000 || 0.08 || | ||
+ | |- | ||
+ | | 04/29/13 || 10:54|| 536 || 1270 || -750 || 0.000 || 0.09 || | ||
+ | |- | ||
+ | | 04/29/13 || 16:56|| 536 || 1500 || -750 || 11.30 || 11.4 || had to turn off power supply for a bit to work on soldering HVB-4, I tried to increase voltage to 1500 first to check if it would trip at 50 muA, The current was 11 uA and no current spike, lets see how long it will stay here. | ||
+ | |- | ||
+ | | 04/29/13 || 17:41|| 536 || 1500 || -750 || 9.8 || 9.9 || Guard Current 0muA, Field Current 9.55muA, Sense Current 9.43muA. | ||
+ | |- | ||
+ | | 04/30/13 || 07:43|| 536 || 1240 || -750 || 3 || 3|| Field HV was in tripped state, current fluctuations are large at 1240. | ||
+ | |- | ||
+ | | 04/30/13 || 09:54|| 536 || 1260 || -750 || .1 || .2|| | ||
+ | |- | ||
+ | | 04/30/13 || 10:44|| 536 || 1270 || -750 || .2 || .3|| | ||
+ | |- | ||
+ | | 04/30/13 || 11:43|| 536 || 1270 || -750 || .1 || .2|| | ||
+ | |- | ||
+ | | 04/30/13 || 11:43|| 536 || 1270 || -750 || .1 || .2|| | ||
+ | |- | ||
+ | | 04/30/13 || 17:53|| 536 || 1270 || -750 || .07 || .155|| | ||
+ | |- | ||
+ | | 05/01/13 || 10:33|| 536 || 1270 || -750 || 0.000 || 0.095|| | ||
+ | |- | ||
+ | | 05/01/13 || 10:54|| 536 || 1280 || -750 || 0.01 || 0.1|| | ||
+ | |- | ||
+ | | 05/01/13 || 11:38|| 536 || 1290 || -750 || 0.02 || 0.1|| | ||
+ | |- | ||
+ | | 05/01/13 || 16:35|| 536 || 1290 || -750 || 0.01 || 0.110|| | ||
+ | |- | ||
+ | | 05/02/13 || 09:27|| 536 || 1290 || -750 || 0.00 || 0.08|| | ||
+ | |- | ||
+ | | 05/02/13 || 09:33|| 536 || 1310 || -750 || 0.00 || 0.1|| | ||
+ | |- | ||
+ | | 05/02/13 || 14:10|| 536 || 1310 || -750 || 0.00 || 0.1|| | ||
+ | |- | ||
+ | | 05/02/13 || 14:30|| 536 || 1320 || -750 || 0.010 || 0.105|| | ||
+ | |- | ||
+ | | 05/02/13 || 14:34|| 536 || 1320 || -750 || 0.015 || 0.115|| | ||
+ | |- | ||
+ | | 05/02/13 || 14:56|| 536 || 1320 || -750 || 0.010 || 0.110|| | ||
+ | |- | ||
+ | | 05/02/13 || 14:58|| 536 || 1330 || -750 || 0.035 || 0.125|| Current is fluctuating by around 50% of this measurement. | ||
+ | |- | ||
+ | | 05/02/13 || 15:12|| 536 || 1330 || -750 || 0.030 || 0.125|| Current reading appears stable within 5% over a 1 minute period. | ||
+ | |- | ||
+ | | 05/02/13 || 15:12|| 536 || 1320 || -750 || 0.020 || 0.115|| Turned down voltage for the weekend. | ||
+ | |- | ||
+ | | 05/03/13 || 10:45|| 536 || 1300 || -750 || 0.00 || 0.08|| Found HV for Field wires in tripped state, lowered Sense voltage | ||
+ | |- | ||
+ | | 05/04/13 || 10:46|| 536 || 1300 || -750 || 0.00 || 0.06|| Found HV for Field wires in tripped state, lowered Sense voltage | ||
+ | |- | ||
+ | | 05/04/13 || 10:47|| 536 || 1320 || -750 || 0.00 || 0.07|| Found HV for Field wires in tripped state, lowered Sense voltage | ||
+ | |- | ||
+ | | 05/06/13 || 07:35|| 536 || 1320 || -750 || 0.00 || 0.07|| | ||
+ | |- | ||
+ | | 05/06/13 || 07:37|| 536 || 1280 || -750 || 0.00 || 0.07|| HV 110 voltage was interrupted and power supply was reset | ||
+ | |- | ||
+ | | 05/06/13 || 11:19|| 536 || 1350 || -750 || 0.04 || 0.13|| | ||
+ | |- | ||
+ | | 05/06/13 || 13:38|| 536 || 1350 || -750 || 0.07 || 0.16|| | ||
+ | |- | ||
+ | | 05/07/13 || 07:37|| 536 || 1360 || -750 || 0.1 || 0.2|| | ||
+ | |- | ||
+ | | 05/07/13 || 09:55|| 536 || 1360 || -750 || 0.13 || 0.22|| | ||
+ | |- | ||
+ | | 05/08/13 || 09:01|| 536 || 1360 || -750 || 0.04 || 0.13|| | ||
+ | |- | ||
+ | | 05/08/13 || 09:01|| 536 || 1380 || -750 || 0.1 || 0.2|| | ||
+ | |- | ||
+ | | 05/08/13 || 12:45|| 536 || 1380 || -750 || 0.12 || 0.22|| | ||
+ | |- | ||
+ | | 05/08/13 || 17:31|| 536 || 1380 || -750 || 0.25 || 0.34|| | ||
+ | |- | ||
+ | | 05/09/13 || 17:20|| 536 || 1380 || -750 || 0.2 || 0.3|| | ||
+ | |- | ||
+ | | 05/09/13 || 17:21|| 536 || 1500 || -750 || 2.3 || 2.4|| pictures have been taken, lets see if voltage will hold or trip | ||
+ | |- | ||
+ | | 05/09/13 || 17:26|| 536 || 1500 || -750 || 2.8 || 2.9|| no current spikes but the current did increase | ||
+ | |- | ||
+ | | 05/09/13 || 17:31|| 536 || 1500 || -750 || 2.7 || 2.8|| | ||
+ | |- | ||
+ | | 05/09/13 || 17:52|| 536 || 1500 || -750 || 2.4 || 2.5|| | ||
+ | |- | ||
+ | | 05/13/13 || 07:43|| 536 || 1500 || -750 || 0.67 || 0.77|| | ||
+ | |- | ||
+ | | 05/14/13 || 07:38|| 536 || 1500 || -750 || 0.35 || 0.44|| | ||
+ | |- |