Difference between revisions of "R1 S6 SignalSearchLog"

From New IAC Wiki
Jump to navigation Jump to search
Line 15: Line 15:
 
|Layer || Wire || Channel || Date ||Comments
 
|Layer || Wire || Channel || Date ||Comments
 
|-
 
|-
|      ||      ||    U1  ||      ||This channel presents higher noise level than the other ones, which makes difficult to get a signal with amplitude below 30mV. This behaviour extends to all the first channels of each connector in the board. however, if a high enough trigger level is set in the scope(~40mV), it is always possible to get a signal in these channels.   
+
|      ||      ||    U1  ||      ||This channel presents higher noise level than the other ones, which makes difficult to get a signal with amplitude below 30mV. This behaviour extends to all the first channels of each connector in the board. however, if a high enough trigger level is set in the scope(~40mV), it is always possible to get a signal in these channels.[[File:Example.jpg|200px]]  
 
|-
 
|-
 
|      ||      ||  U24  ||      || Warm Wire: This is a channel that always triggers a fake signal.
 
|      ||      ||  U24  ||      || Warm Wire: This is a channel that always triggers a fake signal.

Revision as of 20:58, 9 May 2013

Testing procedure

Things that may cause "noise" that are actually due to your testing setup:

  • STB boards not securely attached to chamber. Make sure center screws (ones attaching directly to the chamber, not the nylon stand-offs) are tightened to 14 in. lb's.
  • Scope's termination set to 1Mohm. Set the scope to 50ohm termination.
  • Lower bandwidth of 1 Ghz scope to 200 Mhz. This attenuates peak height, but also rejects high frequency noise and reduces the noise level from 55-65 mV to ~11-15 mV.

Super Layer 1

STB-1 G1202220

Layer Wire Channel Date Comments
U1 This channel presents higher noise level than the other ones, which makes difficult to get a signal with amplitude below 30mV. This behaviour extends to all the first channels of each connector in the board. however, if a high enough trigger level is set in the scope(~40mV), it is always possible to get a signal in these channels.Example.jpg
U24 Warm Wire: This is a channel that always triggers a fake signal.
U38 Warm Wire.
U66 HOT Wire: This is a channel that always triggers a fake signal very often.
U87 HOT Wire.
U93 HOT Wire.

The rest of channels not mention here has a similar behaviour between them, with a noise level not greater than ~20mV.

STB-2 G1202266

Layer Wire Channel Date comments
U38 Hot wire.
U40 Warm wire.
U51 Hot wire.
U65 Hot wire.

STB-3 G1202273

Layer Wire Channel Date comments
U1 Hot wire.
U2 Hot wire.
U39 Hot wire.
U78 Hot wire.

STB-4 G1202275

Layer Wire Channel Date comments
U15 This Board presents a unusual noise in all the channels. The noise is a big positive spark, which makes difficult the recolection of signals. However, It was possible by setting a higher trigger level in the scope. U15 SV 1289 Big Positive Spark.png

STB-5

Layer Wire Date comments

STB-6

Layer Wire Date comments

STB-7

Layer Wire Date comments

Super Layer 2

STB-1

Layer Wire Date comments

STB-2

Layer Wire Date comments

STB-3

Layer Wire Date comments

STB-4

Layer Wire Date comments

STB-5

Layer Wire Date comments

STB-6

Layer Wire Date comments

STB-7

Layer Wire Date comments