| 04/10/13 || 07:40 ||  536 || 970 ||  -750 || 0.115 || 0.005 || Found fields in tripped state, but not sense or guard. Turned on again, current was relatively low.
 
| 04/10/13 || 07:40 ||  536 || 970 ||  -750 || 0.115 || 0.005 || Found fields in tripped state, but not sense or guard. Turned on again, current was relatively low.
 +
|-
 +
| 04/10/13 || 07:50 ||  536 || 1080 ||  -750 || 1.1 || 1.0 || Went up in voltage in 10V increments at 1V/s ramp, monitoring change in average current at each step after voltage settled. Last increment to 1080 saw some surge currents up to 30 <math>\mu</math>A, but average current is around 1 <math>\mu</math>A.

Revision as of 13:48, 10 April 2013

Date Time (24h) Guard (V) Sense (V) Field (V) Sense ([math]\mu[/math]A) Field ([math]\mu[/math]A) comments
04/08/13 12:27 536 1100 -750 Tripped (trip is set to 10 [math]\mu[/math]A). Initially settled around 4-5 [math]\mu[/math]A.
04/08/13 12:36 536 950 -750 Tripped
04/08/13 12:40 536 850 -750 Fluctuating from 1.5-3.0 [math]\mu[/math]A on both sense and field.
04/08/13 13:32 536 850 -750 Tripped
04/08/13 13:40 536 800 -750 Set trip threshold to 15 [math]\mu[/math]A. Current tripped within 5 minutes.
04/08/13 13:45 536 750 -750 Sense/field fluctuating around 1.5 [math]\mu[/math]A.
04/08/13 14:40 536 750 -750 2 2 HV found in tripped state, set Imax to 100 [math]\mu[/math]A, and restored HV.
04/09/13 07:34 536 750 -750 0.1 0
04/09/13 07:35 536 800 -750 0.1 0
04/09/13 07:35 536 900 -750 0.2 0.3
04/09/13 07:37 536 1000 -750 1.5 1.6
04/09/13 07:37 536 1050 -750 6.2 6.3
04/09/13 10:12 536 900 -750 5 4 Found HV in tripped state, restored sense to 900 instead of 1050, it first tripped when attempting 900 Volts but staged 50 volts increases brought it back without tripping, I observe 30[math] \mu[/math]A spikes at 900 Volts
04/09/13 11:00 536 890 -750 0.1 0 Found HV in tripped state, set HV to 890 any further and 30 [math]\mu[/math]A current spikes occurr, changed trip time from 0.1 to 1.1 sec
04/09/13 13:11 536 900 -750 0.1 0 current is jumping more
04/09/13 13:11 536 890 -750 0.1 0 found HV in tripped state, went back to 890
04/09/13 17:22 536 890 -750 0.05 0.00
04/09/13 17:30 536 970 -750 1.3 1.2 Changed voltage in steps of 10V at a time. Changed ramp rate to 1V/s on sense. Observed some surge currents after voltage stabilized at around 30 - 40 [math]\mu[/math]A for a few seconds.
04/10/13 07:40 536 970 -750 0.115 0.005 Found fields in tripped state, but not sense or guard. Turned on again, current was relatively low.
04/10/13 07:50 536 1080 -750 1.1 1.0 Went up in voltage in 10V increments at 1V/s ramp, monitoring change in average current at each step after voltage settled. Last increment to 1080 saw some surge currents up to 30 [math]\mu[/math]A, but average current is around 1 [math]\mu[/math]A.