Difference between revisions of "R1S6 SL1 BurnIn"

From New IAC Wiki
Jump to navigation Jump to search
Line 29: Line 29:
 
| 04/09/13 || 10:12 ||  536 || 900 ||  -750 || 5 || 4 || Found HV in tripped state, restored sense to 900 instead of 1050, it first tripped when attempting 900 Volts but staged 50 volts increases brought it back without tripping, I observe 30<math> \mu</math>A spikes at 900 Volts
 
| 04/09/13 || 10:12 ||  536 || 900 ||  -750 || 5 || 4 || Found HV in tripped state, restored sense to 900 instead of 1050, it first tripped when attempting 900 Volts but staged 50 volts increases brought it back without tripping, I observe 30<math> \mu</math>A spikes at 900 Volts
 
|-
 
|-
| 04/09/13 || 11:00 ||  536 || 890 ||  -750 || 0.1 || 0 || Found HV in tripped state, set HV to 890 any further and 30 <math>mu</math>A current spikes occurr, '''changed trip time from 0.1 to 1.1 sec'''
+
| 04/09/13 || 11:00 ||  536 || 890 ||  -750 || 0.1 || 0 || Found HV in tripped state, set HV to 890 any further and 30 <math>\mu</math>A current spikes occurr, '''changed trip time from 0.1 to 1.1 sec'''
 
|-
 
|-
 
| 04/09/13 || 13:11 ||  536 || 900 ||  -750 || 0.1 || 0 || current is jumping more
 
| 04/09/13 || 13:11 ||  536 || 900 ||  -750 || 0.1 || 0 || current is jumping more

Revision as of 19:11, 9 April 2013

Date Time (24h) Guard (V) Sense (V) Field (V) Sense ([math]\mu[/math]A) Field ([math]\mu[/math]A) comments
04/08/13 12:27 536 1100 -750 Tripped (trip is set to 10 [math]\mu[/math]A). Initially settled around 4-5 [math]\mu[/math]A.
04/08/13 12:36 536 950 -750 Tripped
04/08/13 12:40 536 850 -750 Fluctuating from 1.5-3.0 [math]\mu[/math]A on both sense and field.
04/08/13 13:32 536 850 -750 Tripped
04/08/13 13:40 536 800 -750 Set trip threshold to 15 [math]\mu[/math]A. Current tripped within 5 minutes.
04/08/13 13:45 536 750 -750 Sense/field fluctuating around 1.5 [math]\mu[/math]A.
04/08/13 14:40 536 750 -750 2 2 HV found in tripped state, set Imax to 100 [math]\mu[/math]A, and restored HV.
04/09/13 07:34 536 750 -750 0.1 0
04/09/13 07:35 536 800 -750 0.1 0
04/09/13 07:35 536 900 -750 0.2 0.3
04/09/13 07:37 536 1000 -750 1.5 1.6
04/09/13 07:37 536 1050 -750 6.2 6.3
04/09/13 10:12 536 900 -750 5 4 Found HV in tripped state, restored sense to 900 instead of 1050, it first tripped when attempting 900 Volts but staged 50 volts increases brought it back without tripping, I observe 30[math] \mu[/math]A spikes at 900 Volts
04/09/13 11:00 536 890 -750 0.1 0 Found HV in tripped state, set HV to 890 any further and 30 [math]\mu[/math]A current spikes occurr, changed trip time from 0.1 to 1.1 sec
04/09/13 13:11 536 900 -750 0.1 0 current is jumping more