Difference between revisions of "R1S6 SL1 BurnIn"
Jump to navigation
Jump to search
Line 29: | Line 29: | ||
| 04/09/13 || 10:12 || 536 || 900 || -750 || 5 || 4 || Found HV in tripped state, restored sense to 900 instead of 1050, it first tripped when attempting 900 Volts but staged 50 volts increases brought it back without tripping, I observe 30<math> \mu</math>A spikes at 900 Volts | | 04/09/13 || 10:12 || 536 || 900 || -750 || 5 || 4 || Found HV in tripped state, restored sense to 900 instead of 1050, it first tripped when attempting 900 Volts but staged 50 volts increases brought it back without tripping, I observe 30<math> \mu</math>A spikes at 900 Volts | ||
|- | |- | ||
− | | 04/09/13 || | + | | 04/09/13 || 11:00 || 536 || 890 || -750 || || || Found HV in tripped state, set HV to 890 any further and 30 <math>mu</math>A current spikes occurr, '''changed trip time from 0.1 to 1.1 sec''' |
Revision as of 16:57, 9 April 2013
Date | Time (24h) | Guard (V) | Sense (V) | Field (V) | Sense ( | A)Field ( | A)comments |
04/08/13 | 12:27 | 536 | 1100 | -750 | Tripped (trip is set to 10 | A). Initially settled around 4-5 A.||
04/08/13 | 12:36 | 536 | 950 | -750 | Tripped | ||
04/08/13 | 12:40 | 536 | 850 | -750 | Fluctuating from 1.5-3.0 | A on both sense and field.||
04/08/13 | 13:32 | 536 | 850 | -750 | Tripped | ||
04/08/13 | 13:40 | 536 | 800 | -750 | Set trip threshold to 15 | A. Current tripped within 5 minutes.||
04/08/13 | 13:45 | 536 | 750 | -750 | Sense/field fluctuating around 1.5 | A.||
04/08/13 | 14:40 | 536 | 750 | -750 | 2 | 2 | HV found in tripped state, set Imax to 100 | A, and restored HV.
04/09/13 | 07:34 | 536 | 750 | -750 | 0.1 | 0 | |
04/09/13 | 07:35 | 536 | 800 | -750 | 0.1 | 0 | |
04/09/13 | 07:35 | 536 | 900 | -750 | 0.2 | 0.3 | |
04/09/13 | 07:37 | 536 | 1000 | -750 | 1.5 | 1.6 | |
04/09/13 | 07:37 | 536 | 1050 | -750 | 6.2 | 6.3 | |
04/09/13 | 10:12 | 536 | 900 | -750 | 5 | 4 | Found HV in tripped state, restored sense to 900 instead of 1050, it first tripped when attempting 900 Volts but staged 50 volts increases brought it back without tripping, I observe 30 | A spikes at 900 Volts
04/09/13 | 11:00 | 536 | 890 | -750 | Found HV in tripped state, set HV to 890 any further and 30 | A current spikes occurr, changed trip time from 0.1 to 1.1 sec