| + | ==Testing procedure== | ||
| + | |||
| + | Things that may cause "noise" that are actually due to your testing setup: | ||
| + | |||
| + | * STB boards not securely attached to chamber. Make sure center screws are tightened to 14 in. lb's. | ||
| + | |||
| + | * Scope's termination set to 1Mohm. Set the scope to 50ohm termination. | ||
| + | |||
| + | * Lower bandwidth of 1 Ghz scope to 200 Mhz. This attenuates peak height, but also rejects high frequency noise and reduces the noise level from 55-65 mV to ~11-15 mV. | ||
| + | |||
=Super Layer 1= | =Super Layer 1= | ||
==STB-1== | ==STB-1== | ||
| Line 59: | Line 69: | ||
=Super Layer 2= | =Super Layer 2= | ||
| − | |||
| − | |||
| − | |||
| − | |||
| − | |||
| − | |||
==STB-1== | ==STB-1== | ||
Revision as of 15:53, 2 April 2013
Testing procedure
Things that may cause "noise" that are actually due to your testing setup:
- STB boards not securely attached to chamber. Make sure center screws are tightened to 14 in. lb's.
- Scope's termination set to 1Mohm. Set the scope to 50ohm termination.
- Lower bandwidth of 1 Ghz scope to 200 Mhz. This attenuates peak height, but also rejects high frequency noise and reduces the noise level from 55-65 mV to ~11-15 mV.
Super Layer 1
STB-1
| Layer | Wire | Date | comments |
STB-2
| Layer | Wire | Date | comments |
STB-3
| Layer | Wire | Date | comments |
STB-4
| Layer | Wire | Date | comments |
STB-5
| Layer | Wire | Date | comments |
STB-6
| Layer | Wire | Date | comments |
STB-7
| Layer | Wire | Date | comments |
Super Layer 2
STB-1
| Layer | Wire | Date | comments |
STB-2
| Layer | Wire | Date | comments |
STB-3
| Layer | Wire | Date | comments |
STB-4
| Layer | Wire | Date | comments |
STB-5
| Layer | Wire | Date | comments |
STB-6
| Layer | Wire | Date | comments |
STB-7
| Layer | Wire | Date | comments |