==Testing procedure==
 +
 +
Things that may cause anomalous noise when looking for signals:
 +
 +
1. STB boards not securely attached to chamber. Make sure center screws are tightened to 14 in. lb's.
 +
2. Scope's termination set to 1Mohm. Set the scope to 50ohm termination.
 +
 