This is a technique for efficiency test of the neutron detector used in the 2n correlation experiment. The neutron source was Cf-252 and fission trigger was made out of the NaI(Tl) detector with calibrated ADC and known spectrum of Cf-252 neutrons and photons. Old design of the neutron detector (w/o PMT holders). | This is a technique for efficiency test of the neutron detector used in the 2n correlation experiment. The neutron source was Cf-252 and fission trigger was made out of the NaI(Tl) detector with calibrated ADC and known spectrum of Cf-252 neutrons and photons. Old design of the neutron detector (w/o PMT holders). | ||
Line 7: | Line 16: | ||
[[File:technique1_1.pdf]] | [[File:technique1_1.pdf]] | ||
+ | Some scope pictures showing DAQ timing setup. | ||
+ | |||
+ | [[File:Scope_pic_1.png | 300 px]] | ||
+ | |||
+ | [[File:Scope_pic_2.png | 300 px]] | ||
+ | ==New detector design== | ||
Final procedure with the CFD threshold adjusted. Dr. Dale's design of the PMT holders. | Final procedure with the CFD threshold adjusted. Dr. Dale's design of the PMT holders. | ||
− | [[File: | + | [[File:technique2_2.pdf]] |
+ | |||
+ | Here is some CFD S/N calibration curve for Det_E. | ||
+ | |||
+ | [[File:CFD_calibration_DetE.pdf]] | ||
+ | |||
+ | ==Detector efficiency measurement vis D2 photodisintegration== | ||
+ | |||
+ | [[File:technique_diff_xsect2.pdf]] |
Latest revision as of 18:45, 28 February 2013
Cf-252 TOF calibration
CFD threshold adjustment w/ Cf-252 source
Old detector design
This is a technique for efficiency test of the neutron detector used in the 2n correlation experiment. The neutron source was Cf-252 and fission trigger was made out of the NaI(Tl) detector with calibrated ADC and known spectrum of Cf-252 neutrons and photons. Old design of the neutron detector (w/o PMT holders).
One more stage of the efficiency procedure measurements with improved BKG studying. Old design of the neutron detector (w/o PMT holders).
Some scope pictures showing DAQ timing setup.
New detector design
Final procedure with the CFD threshold adjusted. Dr. Dale's design of the PMT holders.
Here is some CFD S/N calibration curve for Det_E.