Difference between revisions of "CLAS12 R1 S6 QA"

From New IAC Wiki
Jump to navigation Jump to search
Line 139: Line 139:
 
|-
 
|-
 
| 02/26/13 || 14:58 || S6 SL2 HVB7 || 536 || -700 || 1170 || 1.020 || 1.115 ||
 
| 02/26/13 || 14:58 || S6 SL2 HVB7 || 536 || -700 || 1170 || 1.020 || 1.115 ||
 
+
|-
 +
| 02/26/13 || 15:51 || S6 SL2 HVB7 || 536 || -700 || 1180 || 1.00 || 1.090 ||
 +
|-
 +
| 02/26/13 || 15:58 || S6 SL2 HVB7 || 536 || -700 || 1190 || 1.04 || 1.135 ||
 +
|-
 +
| 02/26/13 || 16:35 || S6 SL2 HVB7 || 536 || -700 || 1200 || 1.01 || 1.09 ||
 +
|-
 +
| 02/26/13 || 16:38 || S6 SL2 HVB7 || 536 || -700 || 1200 || Tripped ||  ||
  
  

Revision as of 17:55, 27 February 2013

HV Burn-in Procedure

1. Set current trip threshold to 10 [math]\mu[/math]A on the HV supply

2. Set Voltage to 50 volts on all channels. Probe HVB to be sure that all sense, field, and guard wires have a voltage.

3. Using a ramp speed of 5 V/s, set the HV on the supply to the following:

G:F:S= 525,-700, 700

Monitor the current, make sure there is not an excessive current draw (1+ [math]\mu[/math]A)

Final operating voltage settings range from

G:F:S = (525, -700, 1400) to (536, -700, 1500)

4. Increase the voltage on the sense wire in steps (step-size may change as needed, depending on current draw) and record the initial current draw when the voltage setting on the HV supply is reached on the wire. Allow the field and sense wire current draw to decrease to <1.0 [math]\mu[/math]A before increasing voltage.

If the HV supply trips due to excessive current draw when increasing the voltage, record that the voltage setting caused a trip condition and change the voltage setting using a smaller step-size from the last working voltage that drew <1.0 [math]\mu[/math]A.

As of 2/26/13, we have observed that initially starting at 700V on the sense wire, then jumping to 1000V does not result in excessive current. The "border region" of 1100-1200V and up may require smaller step sizes (20-50V at a time) to avoid exceeding 10 [math]\mu[/math]A current draw.

SL2-HVB1

Guard current is typically below 0.01 microAmp

Date Guard Field Sense I_{Sense} I_{Field} Time
Volts Volts Volts [math]\mu[/math] A [math]\mu[/math] A
2/20/13 525 -700 1400 0.5 0.7 22:50

SL2-HVB2

Guard current is typically below 0.01 microAmp

Date Guard Field Sense I_{Sense} I_{Field} Time
Volts Volts Volts [math]\mu[/math] A [math]\mu[/math] A
2/20/13 525 -700


Sector 6 - Serial Numbers on STB & HVB
Top Endplate Bottom Endplate
Board # (nose=1, tail=7) SL1 (STB) SL2 (HVB) SL1 (HVB) SL2 (STB)
1 G1202270 ? ? K1202516
2 G1202266 ? ? J1202341
3 G1202273 ? ? J1202339
4 G1202275 ? ? K1201716
5 G1202271 ? ? K1201717
6 G1202274 ? ? K1201719
7 G1202272 ? ? K1202346
HV Burn-in Log
Date Time (24h) HV Board ID Guard (V) Field (V) Sense (V) Sense ([math]\mu[/math]A) Field ([math]\mu[/math]A) comments
02/23/13 14:00 S6 SL2 HVB1 525 -700 700 2.2
02/23/13 14:20 S6 SL2 HVB1 525 -700 1400 8.3
02/23/13 14:24 S6 SL2 HVB1 525 -700 1400 5.145
02/23/13 14:27 S6 SL2 HVB1 500 -500 500
02/23/13 14:28 S6 SL2 HVB1 500 ? 700 0
02/23/13 14:41 S6 SL2 HVB1 525 -525 525
02/23/13 14:41 S6 SL2 HVB1 525 -700 700
02/23/13 14:49 S6 SL2 HVB1 525 -700 1400 .5 .7
02/23/13 15:50 S6 SL2 HVB1 525 -700 1450 .8 .775
02/23/13 15:52 S6 SL2 HVB1 525 -700 1400 .465 .465
02/23/13 17:25 S6 SL2 HVB1 536 -700 1500 .720
02/25/13 17:05 S6 SL2 HVB7 536 -500 550 .08 .075
02/25/13 17:15 S6 SL2 HVB7 536 -700 700 .08 .01
02/25/13 17:18 S6 SL2 HVB7 536 -700 1000 .12 .05
02/25/13 17:21 S6 SL2 HVB7 536 -700 1000 .345
02/25/13 17:24 S6 SL2 HVB7 536 -700 1200 8.9 almost tripped
02/25/13 17:26 S6 SL2 HVB7 536 -700 1100 1.85
02/25/13 17:31 S6 SL2 HVB7 536 -700 1200 8.6 Almost lowered voltage but it's decreasing
02/25/13 17:34 S6 SL2 HVB7 536 -700 1200 7.3
02/25/13 17:38 S6 SL2 HVB7 536 -700 1200
tripped
02/25/13 17:38 S6 SL2 HVB7 536 -700 1150 2.85
02/25/13 17:55 S6 SL2 HVB7 536 -700 1150 1.04 Powered down at end of day (18:00)
02/26/13 14:25 S6 SL2 HVB7 536 -500 500 0 0.01
02/26/13 14:30 S6 SL2 HVB7 536 -700 700 0 0.035
02/26/13 14:33 S6 SL2 HVB7 536 -700 1000 0.3 0.38
02/26/13 14:33 S6 SL2 HVB7 536 -700 1100 0.7 0.785
02/26/13 14:45 S6 SL2 HVB7 536 -700 1150 0.93 1.05
02/26/13 14:54 S6 SL2 HVB7 536 -700 1200

trip

02/26/13 14:58 S6 SL2 HVB7 536 -700 1170 1.020 1.115
02/26/13 15:51 S6 SL2 HVB7 536 -700 1180 1.00 1.090
02/26/13 15:58 S6 SL2 HVB7 536 -700 1190 1.04 1.135
02/26/13 16:35 S6 SL2 HVB7 536 -700 1200 1.01 1.09
02/26/13 16:38 S6 SL2 HVB7 536 -700 1200 Tripped


Pre-Potting Tension Error

SL1S6 Sense PrePot.pngSL2S6 Sense PrePot.png

SL1S6 Field PrePot.pngSL2S6 Field PrePot.png

MYSQL commands

mysql> show databases;

+--------------------+
| Database           |
+--------------------+
| information_schema | 
| mysql              | 
| region1            | 
| test               | 
+--------------------+
4 rows in set (0.02 sec)


mysql> use region1;

Reading table information for completion of table and column names
You can turn off this feature to get a quicker startup with -A

Database changed


mysql> show tables;

+-------------------------------+
| Tables_in_region1             |
+-------------------------------+
| R1S1_Tension_Trend_Test_Field | 
| R1S1_Tension_Trend_Test_Sense | 
| R1S4_Tension_Trend_Test_Field | 
| R1S4_Tension_Trend_Test_Sense | 
| R1S5_Tension_Trend_Test_Field | 
| R1S6_Tension_Trend_Test_Field | 
| R1S6_Tension_Trend_Test_Sense | 
| R1_SL1_S1_Field               | 
| R1_SL1_S1_Sense               | 
| R1_SL1_S4_Field               | 
| R1_SL1_S4_Sense               | 
| R1_SL1_S5_Field               | 
| R1_SL1_S6_Field               | 
| R1_SL1_S6_Sense               | 
| R1_SL2_S1_Field               | 
| R1_SL2_S1_Sense               | 
| R1_SL2_S4_Field               | 
| R1_SL2_S4_Sense               | 
| R1_SL2_S5_Field               | 
| R1_SL2_S6_Field               | 
| R1_SL2_S6_Sense               | 
| R1_particleCounts             | 
| R1_work_log                   | 
+-------------------------------+
23 rows in set (0.00 sec)

mysql> describe R1_SL1_S6_Sense;

+------------------+-------------+------+-----+-------------------+-------+
| Field            | Type        | Null | Key | Default           | Extra |
+------------------+-------------+------+-----+-------------------+-------+
| iCol             | int(11)     | NO   |     | NULL              |       | 
| iRow             | int(11)     | NO   |     | NULL              |       | 
| PeriodMeasured   | float       | YES  |     | NULL              |       | 
| Result           | varchar(32) | YES  |     | NULL              |       | 
| TensionMeasured  | float       | YES  |     | NULL              |       | 
| CalculatedPeriod | float       | YES  |     | NULL              |       | 
| SuperLayer       | int(11)     | NO   |     | NULL              |       | 
| Length           | float       | YES  |     | NULL              |       | 
| FreqMeasured     | float       | YES  |     | NULL              |       | 
| TensionError     | float       | YES  |     | NULL              |       | 
| TestLogic        | int(11)     | NO   |     | NULL              |       | 
| CalcFreq         | float       | YES  |     | NULL              |       | 
| HighCut          | float       | YES  |     | NULL              |       | 
| LowCut           | float       | YES  |     | NULL              |       | 
| Sector           | int(11)     | NO   |     | NULL              |       | 
| Date             | varchar(32) | YES  |     | NULL              |       | 
| Mod_time         | timestamp   | NO   |     | CURRENT_TIMESTAMP |       | 
+------------------+-------------+------+-----+-------------------+-------+
17 rows in set (0.04 sec)


mysql> select TensionError from R1_SL1_S6_Sense INTO OUTFILE '/tmp/R1SL1S6Sense.dat';