Difference between revisions of "CLAS12 R1 S6 QA"
Line 139: | Line 139: | ||
|- | |- | ||
| 02/26/13 || 14:58 || S6 SL2 HVB7 || 536 || -700 || 1170 || 1.020 || 1.115 || | | 02/26/13 || 14:58 || S6 SL2 HVB7 || 536 || -700 || 1170 || 1.020 || 1.115 || | ||
− | + | |- | |
+ | | 02/26/13 || 15:51 || S6 SL2 HVB7 || 536 || -700 || 1180 || 1.00 || 1.090 || | ||
+ | |- | ||
+ | | 02/26/13 || 15:58 || S6 SL2 HVB7 || 536 || -700 || 1190 || 1.04 || 1.135 || | ||
+ | |- | ||
+ | | 02/26/13 || 16:35 || S6 SL2 HVB7 || 536 || -700 || 1200 || 1.01 || 1.09 || | ||
+ | |- | ||
+ | | 02/26/13 || 16:38 || S6 SL2 HVB7 || 536 || -700 || 1200 || Tripped || || | ||
Revision as of 17:55, 27 February 2013
HV Burn-in Procedure
1. Set current trip threshold to 10
A on the HV supply2. Set Voltage to 50 volts on all channels. Probe HVB to be sure that all sense, field, and guard wires have a voltage.
3. Using a ramp speed of 5 V/s, set the HV on the supply to the following:
G:F:S= 525,-700, 700
Monitor the current, make sure there is not an excessive current draw (1+
A)Final operating voltage settings range from
G:F:S = (525, -700, 1400) to (536, -700, 1500)
4. Increase the voltage on the sense wire in steps (step-size may change as needed, depending on current draw) and record the initial current draw when the voltage setting on the HV supply is reached on the wire. Allow the field and sense wire current draw to decrease to <1.0
A before increasing voltage.If the HV supply trips due to excessive current draw when increasing the voltage, record that the voltage setting caused a trip condition and change the voltage setting using a smaller step-size from the last working voltage that drew <1.0
A.As of 2/26/13, we have observed that initially starting at 700V on the sense wire, then jumping to 1000V does not result in excessive current. The "border region" of 1100-1200V and up may require smaller step sizes (20-50V at a time) to avoid exceeding 10
A current draw.SL2-HVB1
Guard current is typically below 0.01 microAmp
Date | Guard | Field | Sense | I_{Sense} | I_{Field} | Time |
Volts | Volts | Volts | A | A | ||
2/20/13 | 525 | -700 | 1400 | 0.5 | 0.7 | 22:50 |
SL2-HVB2
Guard current is typically below 0.01 microAmp
Date | Guard | Field | Sense | I_{Sense} | I_{Field} | Time |
Volts | Volts | Volts | A | A | ||
2/20/13 | 525 | -700 |
Sector 6 - Serial Numbers on STB & HVB | |||||
Top Endplate | Bottom Endplate | ||||
Board # (nose=1, tail=7) | SL1 (STB) | SL2 (HVB) | SL1 (HVB) | SL2 (STB) | |
1 | G1202270 | ? | ? | K1202516 | |
2 | G1202266 | ? | ? | J1202341 | |
3 | G1202273 | ? | ? | J1202339 | |
4 | G1202275 | ? | ? | K1201716 | |
5 | G1202271 | ? | ? | K1201717 | |
6 | G1202274 | ? | ? | K1201719 | |
7 | G1202272 | ? | ? | K1202346 |
HV Burn-in Log | ||||||||
Date | Time (24h) | HV Board ID | Guard (V) | Field (V) | Sense (V) | Sense ( | A)Field ( | A)comments |
02/23/13 | 14:00 | S6 SL2 HVB1 | 525 | -700 | 700 | 2.2 | ||
02/23/13 | 14:20 | S6 SL2 HVB1 | 525 | -700 | 1400 | 8.3 | ||
02/23/13 | 14:24 | S6 SL2 HVB1 | 525 | -700 | 1400 | 5.145 | ||
02/23/13 | 14:27 | S6 SL2 HVB1 | 500 | -500 | 500 | |||
02/23/13 | 14:28 | S6 SL2 HVB1 | 500 | ? | 700 | 0 | ||
02/23/13 | 14:41 | S6 SL2 HVB1 | 525 | -525 | 525 | |||
02/23/13 | 14:41 | S6 SL2 HVB1 | 525 | -700 | 700 | |||
02/23/13 | 14:49 | S6 SL2 HVB1 | 525 | -700 | 1400 | .5 | .7 | |
02/23/13 | 15:50 | S6 SL2 HVB1 | 525 | -700 | 1450 | .8 | .775 | |
02/23/13 | 15:52 | S6 SL2 HVB1 | 525 | -700 | 1400 | .465 | .465 | |
02/23/13 | 17:25 | S6 SL2 HVB1 | 536 | -700 | 1500 | .720 | ||
02/25/13 | 17:05 | S6 SL2 HVB7 | 536 | -500 | 550 | .08 | .075 | |
02/25/13 | 17:15 | S6 SL2 HVB7 | 536 | -700 | 700 | .08 | .01 | |
02/25/13 | 17:18 | S6 SL2 HVB7 | 536 | -700 | 1000 | .12 | .05 | |
02/25/13 | 17:21 | S6 SL2 HVB7 | 536 | -700 | 1000 | .345 | ||
02/25/13 | 17:24 | S6 SL2 HVB7 | 536 | -700 | 1200 | 8.9 | almost tripped | |
02/25/13 | 17:26 | S6 SL2 HVB7 | 536 | -700 | 1100 | 1.85 | ||
02/25/13 | 17:31 | S6 SL2 HVB7 | 536 | -700 | 1200 | 8.6 | Almost lowered voltage but it's decreasing | |
02/25/13 | 17:34 | S6 SL2 HVB7 | 536 | -700 | 1200 | 7.3 | ||
02/25/13 | 17:38 | S6 SL2 HVB7 | 536 | -700 | 1200 |
tripped | ||
02/25/13 | 17:38 | S6 SL2 HVB7 | 536 | -700 | 1150 | 2.85 | ||
02/25/13 | 17:55 | S6 SL2 HVB7 | 536 | -700 | 1150 | 1.04 | Powered down at end of day (18:00) | |
02/26/13 | 14:25 | S6 SL2 HVB7 | 536 | -500 | 500 | 0 | 0.01 | |
02/26/13 | 14:30 | S6 SL2 HVB7 | 536 | -700 | 700 | 0 | 0.035 | |
02/26/13 | 14:33 | S6 SL2 HVB7 | 536 | -700 | 1000 | 0.3 | 0.38 | |
02/26/13 | 14:33 | S6 SL2 HVB7 | 536 | -700 | 1100 | 0.7 | 0.785 | |
02/26/13 | 14:45 | S6 SL2 HVB7 | 536 | -700 | 1150 | 0.93 | 1.05 | |
02/26/13 | 14:54 | S6 SL2 HVB7 | 536 | -700 | 1200 |
trip | ||
02/26/13 | 14:58 | S6 SL2 HVB7 | 536 | -700 | 1170 | 1.020 | 1.115 | |
02/26/13 | 15:51 | S6 SL2 HVB7 | 536 | -700 | 1180 | 1.00 | 1.090 | |
02/26/13 | 15:58 | S6 SL2 HVB7 | 536 | -700 | 1190 | 1.04 | 1.135 | |
02/26/13 | 16:35 | S6 SL2 HVB7 | 536 | -700 | 1200 | 1.01 | 1.09 | |
02/26/13 | 16:38 | S6 SL2 HVB7 | 536 | -700 | 1200 | Tripped |
|
Pre-Potting Tension Error
MYSQL commands
mysql> show databases;
+--------------------+ | Database | +--------------------+ | information_schema | | mysql | | region1 | | test | +--------------------+ 4 rows in set (0.02 sec)
mysql> use region1;
Reading table information for completion of table and column names You can turn off this feature to get a quicker startup with -A Database changed
mysql> show tables;
+-------------------------------+ | Tables_in_region1 | +-------------------------------+ | R1S1_Tension_Trend_Test_Field | | R1S1_Tension_Trend_Test_Sense | | R1S4_Tension_Trend_Test_Field | | R1S4_Tension_Trend_Test_Sense | | R1S5_Tension_Trend_Test_Field | | R1S6_Tension_Trend_Test_Field | | R1S6_Tension_Trend_Test_Sense | | R1_SL1_S1_Field | | R1_SL1_S1_Sense | | R1_SL1_S4_Field | | R1_SL1_S4_Sense | | R1_SL1_S5_Field | | R1_SL1_S6_Field | | R1_SL1_S6_Sense | | R1_SL2_S1_Field | | R1_SL2_S1_Sense | | R1_SL2_S4_Field | | R1_SL2_S4_Sense | | R1_SL2_S5_Field | | R1_SL2_S6_Field | | R1_SL2_S6_Sense | | R1_particleCounts | | R1_work_log | +-------------------------------+ 23 rows in set (0.00 sec)
mysql> describe R1_SL1_S6_Sense;
+------------------+-------------+------+-----+-------------------+-------+ | Field | Type | Null | Key | Default | Extra | +------------------+-------------+------+-----+-------------------+-------+ | iCol | int(11) | NO | | NULL | | | iRow | int(11) | NO | | NULL | | | PeriodMeasured | float | YES | | NULL | | | Result | varchar(32) | YES | | NULL | | | TensionMeasured | float | YES | | NULL | | | CalculatedPeriod | float | YES | | NULL | | | SuperLayer | int(11) | NO | | NULL | | | Length | float | YES | | NULL | | | FreqMeasured | float | YES | | NULL | | | TensionError | float | YES | | NULL | | | TestLogic | int(11) | NO | | NULL | | | CalcFreq | float | YES | | NULL | | | HighCut | float | YES | | NULL | | | LowCut | float | YES | | NULL | | | Sector | int(11) | NO | | NULL | | | Date | varchar(32) | YES | | NULL | | | Mod_time | timestamp | NO | | CURRENT_TIMESTAMP | | +------------------+-------------+------+-----+-------------------+-------+ 17 rows in set (0.04 sec)
mysql> select TensionError from R1_SL1_S6_Sense INTO OUTFILE '/tmp/R1SL1S6Sense.dat';