+ | =Method used at CERN= | ||
− | + | #When the leakage current is above 10 nA and below 200 nA when the GEM HV is 600 Volts, the GEMS are cleaned. | |
+ | #First try high pressure (2000 PSI) distilled water (10 Mohms quality) spray. CERN has what looks like a power washer and they spay it against a tub wall in a workshop. | ||
+ | #After spraying with distilled water they bake it at 80 celsius in an oven for 1 hour | ||
+ | #If this doesn't work they will repeat the above after immersing the foil in a degreaser bath at 60 degrees celsius for 1 minute. Then rinse with a non pressure (not high pressure) spray of DI. | ||
+ | |||
+ | Ask RUI which degreaser he uses | ||
When the leakage current is below 200 nA we just clean the GEMs with sprayed DI water (DI water 10Mohms quality) | When the leakage current is below 200 nA we just clean the GEMs with sprayed DI water (DI water 10Mohms quality) | ||
− | + | The above info came from Alexis Rodrigues (Alexis.Rodrigues@cern.ch) . Alexis did not recommend the alcohol method. | |
+ | |||
+ | |||
+ | ==Power Washers== | ||
+ | |||
+ | Power washer similar to what RUI is using | ||
+ | |||
+ | http://www.amazon.com/Earthwise-PWO2200-Electric-Pressure-Washer/dp/B002MAPNEE | ||
+ | |||
+ | the smaller model below may be better for the LDS | ||
+ | |||
+ | http://www.amazon.com/AR-AR118-Electric-Pressure-Washer/dp/B003B6KQZE/ref=pd_sbs_ol_16 | ||
+ | |||
+ | manual | ||
+ | |||
+ | http://www.arblueclean.com/uploadfiles/image/file/AR118%20Operating%20Manual(1).pdf | ||
+ | |||
+ | ==Clean room Ovens== | ||
+ | |||
+ | Terra Universal: Class 100 Cleanroom Oven Cat# 3700-70 $8,992, 2.8 cu. ft. | ||
+ | |||
+ | |||
+ | |||
+ | |||
+ | |||
+ | http://www.memmert.com/products/cleanroom-drying-oven/cleanroomdryingoven-overview/ | ||
+ | |||
+ | |||
+ | http://www.memmert.com/en/products/cleanroom-drying-oven/cleanroomdryingoven-models/models/UFP-500-RR/ | ||
+ | |||
+ | |||
+ | http://www.memmert.com/en/products/cleanroom-drying-oven/cleanroomdryingoven-models/models/UFP-500-RR/ | ||
+ | |||
+ | |||
+ | http://www.blue-m.com/products/productdetails.aspx?tid=1&pscid=47&plid=85&src=gaw&gclid=CPi77Nukh6sCFYTt7QodQnEgMQ | ||
+ | |||
+ | |||
+ | http://www.tedpella.com/ovens_html/ovens.htm#anchor1388 | ||
+ | |||
+ | http://www.carbolite.us/cleanroomovens.html | ||
+ | |||
+ | http://www.despatch.com/industrial-ovens/benchtop-ovens-and-lab-ovens.aspx | ||
+ | |||
+ | ==Warming tank== | ||
+ | |||
+ | A degreaser solution is stored at 60 celsius so foils can be dipped for 1 minute to clean the copper. | ||
+ | |||
+ | |||
+ | check TerraUniversal Ultrasonic Cleaner with heating | ||
+ | |||
+ | 2.5 gallon, cat #2619-00 $964. | ||
+ | |||
+ | Need a stainless steel tank to hold the solution and keep it warm | ||
+ | |||
+ | http://www.platingsales.com/electropolish.html | ||
+ | |||
+ | ==Deionizers== | ||
+ | |||
+ | http://www.portablewaterdi.com/deionizers.php | ||
+ | |||
+ | ==1600 PSI Electric Pressure Washer== | ||
+ | |||
+ | http://www.sears.com/shc/s/p_10153_12605_07175265000P?prdNo=4&blockNo=4&blockType=L4 | ||
+ | |||
+ | =Brookhaven= | ||
+ | The basic procedure followed for cleaning GEM foils suggested by Bob Azmoun, a researcher in the PHENIX Group at Brookhaven National Lab. [[Media:Georgia_3GEM_Detector_UR_Report_F04-1.pdf]], [[Media:UG_Research_Report_David_Fall_04-1.pdf]]. | ||
Line 19: | Line 90: | ||
One should keep in mind that after this process the GEM impedance should drop dramatically, due to the conductivity of the remaining alcohol. However, usually after 5hrs of purging with dry gas, the impedance returns to its original value. " | One should keep in mind that after this process the GEM impedance should drop dramatically, due to the conductivity of the remaining alcohol. However, usually after 5hrs of purging with dry gas, the impedance returns to its original value. " | ||
+ | |||
+ | |||
[https://wiki.iac.isu.edu/index.php/GEMS] | [https://wiki.iac.isu.edu/index.php/GEMS] |
Latest revision as of 17:28, 10 October 2011
Method used at CERN
- When the leakage current is above 10 nA and below 200 nA when the GEM HV is 600 Volts, the GEMS are cleaned.
- First try high pressure (2000 PSI) distilled water (10 Mohms quality) spray. CERN has what looks like a power washer and they spay it against a tub wall in a workshop.
- After spraying with distilled water they bake it at 80 celsius in an oven for 1 hour
- If this doesn't work they will repeat the above after immersing the foil in a degreaser bath at 60 degrees celsius for 1 minute. Then rinse with a non pressure (not high pressure) spray of DI.
Ask RUI which degreaser he uses
When the leakage current is below 200 nA we just clean the GEMs with sprayed DI water (DI water 10Mohms quality)
The above info came from Alexis Rodrigues (Alexis.Rodrigues@cern.ch) . Alexis did not recommend the alcohol method.
Power Washers
Power washer similar to what RUI is using
http://www.amazon.com/Earthwise-PWO2200-Electric-Pressure-Washer/dp/B002MAPNEE
the smaller model below may be better for the LDS
http://www.amazon.com/AR-AR118-Electric-Pressure-Washer/dp/B003B6KQZE/ref=pd_sbs_ol_16
manual
http://www.arblueclean.com/uploadfiles/image/file/AR118%20Operating%20Manual(1).pdf
Clean room Ovens
Terra Universal: Class 100 Cleanroom Oven Cat# 3700-70 $8,992, 2.8 cu. ft.
http://www.memmert.com/products/cleanroom-drying-oven/cleanroomdryingoven-overview/
http://www.tedpella.com/ovens_html/ovens.htm#anchor1388
http://www.carbolite.us/cleanroomovens.html
http://www.despatch.com/industrial-ovens/benchtop-ovens-and-lab-ovens.aspx
Warming tank
A degreaser solution is stored at 60 celsius so foils can be dipped for 1 minute to clean the copper.
check TerraUniversal Ultrasonic Cleaner with heating
2.5 gallon, cat #2619-00 $964.
Need a stainless steel tank to hold the solution and keep it warm
http://www.platingsales.com/electropolish.html
Deionizers
http://www.portablewaterdi.com/deionizers.php
1600 PSI Electric Pressure Washer
http://www.sears.com/shc/s/p_10153_12605_07175265000P?prdNo=4&blockNo=4&blockType=L4
Brookhaven
The basic procedure followed for cleaning GEM foils suggested by Bob Azmoun, a researcher in the PHENIX Group at Brookhaven National Lab. Media:Georgia_3GEM_Detector_UR_Report_F04-1.pdf, Media:UG_Research_Report_David_Fall_04-1.pdf.
The procedure, which is done under a laminar flow hood, is outlined below:
1.Spray down the foils (always at “grazing incidence”) with dry N2
2.Spray down the foils with ethyl-alcohol until foils are completely drenched, using a Windex-type dispenser spray
3.Before the ethanol has a chance to air dry (which could absorb particulate deposits from the air onto the GEM surface), quickly spray down the foils again with dry N2 (again, always at “grazing incidence”)
One should keep in mind that after this process the GEM impedance should drop dramatically, due to the conductivity of the remaining alcohol. However, usually after 5hrs of purging with dry gas, the impedance returns to its original value. "