Difference between revisions of "05/13/2011 coincidence test"

From New IAC Wiki
Jump to navigation Jump to search
Line 43: Line 43:
 
|}
 
|}
  
the scope picture of coincidence event like:
+
 
 +
The scope picture of PS1+PS2 coincidence event like:
 +
[[File:Cosmic coinc PS1 PS2.png |100px]]

Revision as of 21:49, 13 May 2011

Go Back

Set up like that:

Detector1.png

the distance between PS and n's PMT is abot 80 cm


The voltages and thresholds are as below. Note: I increased my thresholds for PS1 and PS2 from 80 mV up to 100 mV according to mu previous calibration threshold line.

PMT voltage, V threshold, mV discr. width, ns
PS1 -1500 -80 100
PS1 -1500 -80 100
n's -1500 -200 10


Let's do first the coincidence test PS1+PS2

counts time counts/sec
PS1 2823[math]\pm[/math] 20 min [math]\pm[/math]
PS2 3091[math]\pm[/math] 20 min [math]\pm[/math]
PS1+PS2 93[math]\pm[/math] 20 min [math]\pm[/math]


The scope picture of PS1+PS2 coincidence event like: Cosmic coinc PS1 PS2.png