Difference between revisions of "05/12/2011"
Jump to navigation
Jump to search
Line 45: | Line 45: | ||
Let's do the '''2n correlation test''' with CF252 source. The voltages and thresholds are the same. Set up is like that: | Let's do the '''2n correlation test''' with CF252 source. The voltages and thresholds are the same. Set up is like that: | ||
+ | :the distance between PS1 and PS2 is | ||
+ | :the CF252 source in the middle | ||
+ | Let's do the '''random coincidence cosmic ray test'''. PS1 and PS2 is about 30 cm away from each other. And let's look on the coincidence signal between PS1 and PS2: | ||
+ | {| border="1" cellspacing="0" style="text-align: center; width: 800px; height: 100px;" | ||
+ | ! scope="col" width="50" | | ||
+ | ! scope="col" width="50" | total counts | ||
+ | ! scope="col" width="50" | time | ||
+ | ! scope="col" width="50" | counts/min | ||
+ | ! scope="col" width="50" | counts/sec | ||
+ | |- | ||
+ | !PS1 | ||
+ | |<math>255977\pm505.9</math>||10 min||<math>25597.7\pm50.6</math>||<math>426.63\pm0.84</math> | ||
+ | |- | ||
+ | !PS1 | ||
+ | |<math>421917\pm649.5</math>||10 min||<math>42191.7\pm64.9</math>||<math>703.19\pm1.08</math> | ||
+ | |- | ||
+ | !PS1+PS2 | ||
+ | |<math>1224\pm34.9</math>||10 min||<math>122.4\pm3.5</math>||<math>2.04\pm0.06</math> | ||
+ | |} |
Revision as of 20:30, 12 May 2011
The powers are:
PS1 and PS2: -1000 V (the old 2-channel TENNELEC TC952) n's: -1100 V (the new 2-channel C.A.E.N N471)
The thresholds are:
PS1 and PS2: -80 mV n's: -200 mV
The discriminator widths are:
PS1: -20 ns PS2: -10 ns n's: -20 ns
Let's do the random coincidence cosmic ray test. PS1 and PS2 is about 30 cm away from each other. And let's look on the coincidence signal between PS1 and PS2:
total counts | time | counts/min | counts/sec | |
---|---|---|---|---|
PS1 | 70 min | |||
PS1 | 70 min | |||
PS1+PS2 | 70 min |
And below my scope pictures of this rear event:
Let's do the 2n correlation test with CF252 source. The voltages and thresholds are the same. Set up is like that:
- the distance between PS1 and PS2 is
- the CF252 source in the middle
Let's do the random coincidence cosmic ray test. PS1 and PS2 is about 30 cm away from each other. And let's look on the coincidence signal between PS1 and PS2:
total counts | time | counts/min | counts/sec | |
---|---|---|---|---|
PS1 | 10 min | |||
PS1 | 10 min | |||
PS1+PS2 | 10 min |