Difference between revisions of "05/12/2011"
Jump to navigation
Jump to search
(Created page with '[https://wiki.iac.isu.edu/index.php/N%27s_detector_calibration Go Back] The powers are: PS1 and PS2: -1000 V (the old 2-channel TENNELEC TC952) n's: -1100 V (the new 2-chan…') |
|||
| Line 11: | Line 11: | ||
PS1 and PS2: -80 mV | PS1 and PS2: -80 mV | ||
n's: -200 mV | n's: -200 mV | ||
| − | |||
The discriminator widths are: | The discriminator widths are: | ||
| Line 20: | Line 19: | ||
| − | Let's do the random coincidence test: | + | Let's do the random coincidence test. PS1 and PS2 is about 30 cm away from each other. And let's look on the coincidence signal between PS1 and PS2: |
| + | {| border="1" cellspacing="0" style="text-align: center; width: 800px; height: 100px;" | ||
| + | ! scope="col" width="50" | | ||
| + | ! scope="col" width="50" | counts/sec (5 min test), no source | ||
| + | |- | ||
| + | !PS1+PS2 | ||
| + | |19.89<math>\pm</math>0.26 (total 5966 cnts) | ||
| + | |} | ||
Revision as of 18:12, 12 May 2011
The powers are:
PS1 and PS2: -1000 V (the old 2-channel TENNELEC TC952) n's: -1100 V (the new 2-channel C.A.E.N N471)
The thresholds are:
PS1 and PS2: -80 mV n's: -200 mV
The discriminator widths are:
PS1: -20 ns PS2: -10 ns n's: -20 ns
Let's do the random coincidence test. PS1 and PS2 is about 30 cm away from each other. And let's look on the coincidence signal between PS1 and PS2:
| counts/sec (5 min test), no source | |
|---|---|
| PS1+PS2 | 19.890.26 (total 5966 cnts) |