HRRL 03-15-2011: Difference between revisions

From New IAC Wiki
Jump to navigation Jump to search
Setisadi (talk | contribs)
Setisadi (talk | contribs)
Line 134: Line 134:
x-projection:
x-projection:


emit=0.110 +- 0.001 mm*mrad, emit_norm=2.158 +- 0.029 mm*mrad  
emit=0.110 +- 0.001 mm*mrad, emit_norm=2.158 +- 0.014 mm*mrad  


beta=0.439 +- 0.003, alpha=1.439 +-0.009
beta=0.439 +- 0.001, alpha=0.985 +-0.004


[[File:HRRL_05-15-3011-Scan_Q4_176mA_Par_Fit_y.png | 300px]]
[[File:HRRL_05-15-3011-Scan_Q4_176mA_Par_Fit_y.png | 300px]]
Line 148: Line 148:
===88 mA===
===88 mA===


[[File:HRRL_05-15-3011-Scan_Q4_176mA_Par_Fit_x.png | 300px]]
[[File:HRRL_05-15-3011-Scan_Q4_88mA_Par_Fit_x.png | 300px]]






[[File:HRRL_05-15-3011-Scan_Q4_176mA_Par_Fit_y.png | 300px]]
[[File:HRRL_05-15-3011-Scan_Q4_88mA_Par_Fit_y.png | 300px]]
 
 


===44mA===
===44mA===


[[File:HRRL_05-15-3011-Scan_Q4_176mA_Par_Fit_x.png | 300px]]
[[File:HRRL_05-15-3011-Scan_Q4_44mA_Par_Fit_x.png | 300px]]






[[File:HRRL_05-15-3011-Scan_Q4_176mA_Par_Fit_y.png | 300px]]
[[File:HRRL_05-15-3011-Scan_Q4_44mA_Par_Fit_y.png | 300px]]
 
y-projection:
 
emit=0.920 +- 0.023 mm*mrad, emit_norm=17.998 +- 0.442 mm*mrad
 
beta=0.630 +- 0.010, alpha=-1.956 +-0.031


===22 mA===
===22 mA===


[[File:HRRL_05-15-3011-Scan_Q4_176mA_Par_Fit_x.png | 300px]]
[[File:HRRL_05-15-3011-Scan_Q4_22mA_Par_Fit_x.png | 300px]]






[[File:HRRL_05-15-3011-Scan_Q4_176mA_Par_Fit_y.png | 300px]]
[[File:HRRL_05-15-3011-Scan_Q4_22mA_Par_Fit_y.png | 300px]]


===11 mA===
===11 mA===


[[File:HRRL_05-15-3011-Scan_Q4_176mA_Par_Fit_x.png | 300px]]
[[File:HRRL_05-15-3011-Scan_Q4_11mA_Par_Fit_x.png | 300px]]
 




[[File:HRRL_05-15-3011-Scan_Q4_11mA_Par_Fit_y.png | 300px]]


[[File:HRRL_05-15-3011-Scan_Q4_176mA_Par_Fit_y.png | 300px]]
y-projection:
 
emit=0.085 +- 0.005 mm*mrad, emit_norm=1.656 +- 0.098 mm*mrad
 
beta=0.321 +- 0.006, alpha=-1.349 +-0.029


=Summary=
=Summary=

Revision as of 16:46, 16 March 2011

Quad scan

1.) deliver 165 mA peak current to FC with all Quads off Ouput corrector off (H2=0,V2=0) 2.) Use Q1-Q3 to make smallest spot on the OTR

3.) Use Output corrector to center beam spot onto OTR. Record transmission.

Quad scan using Q4

Do quick quad scan with analog camera


Install digital and shielding

Quad current STOR image
File:Image.png

Scan Results

(Normalized Emittance)=(Relativistic Factor) *(beta) * (Geometrical Emittance)

[math]\displaystyle{ \epsilon_{nor} = \gamma_{rel} \beta \epsilon_{geo} }[/math]

[math]\displaystyle{ \beta=v/c }[/math] is almost one for 10 MeV

Scan_Quad1

163mA


x-projection:

emit=0.236 +- 0.001 mm*mrad, emit_norm=4.622 +- 0.018 mm*mrad

beta=0.574 +- 0.001, alpha=0.895 +-0.002


y-projection:

emit=0.299 +- 0.003 mm*mrad, emit_norm=5.846 +- 0.054 mm*mrad

beta=0.315 +- 0.001, alpha=-0.742 +-0.004

80mA


x-projection:

emit=0.110 +- 0.001 mm*mrad, emit_norm=2.151 +- 0.014 mm*mrad

beta=0.378 +- 0.001, alpha=0.957 +-0.003



y-projection:

emit=0.281 +- 0.002 mm*mrad, emit_norm=5.491 +- 0.030 mm*mrad

beta=0.546 +- 0.001, alpha=-0.873 +-0.003


40mA


x-projection:

emit=0.267 +- 0.002 mm*mrad, emit_norm=5.224 +- 0.033 mm*mrad

beta=0.449 +- 0.001, alpha=1.142 +-0.004




y-projection:

emit=0.731 +- 0.004 mm*mrad, emit_norm=14.303 +- 0.076 mm*mrad

beta=0.382 +- 0.001, alpha=-0.613 +-0.002


20mA


x-projection:

emit=0.245 +- 0.002 mm*mrad, emit_norm=4.804 +- 0.031 mm*mrad

beta=0.463 +- 0.001, alpha=1.139 +-0.004



y-projection:

emit=0.699 +- 0.003 mm*mrad, emit_norm=13.673 +- 0.063 mm*mrad

beta=0.459 +- 0.001, alpha=-0.601 +-0.002



Scan_Quad4

176 mA

x-projection:

emit=0.110 +- 0.001 mm*mrad, emit_norm=2.158 +- 0.014 mm*mrad

beta=0.439 +- 0.001, alpha=0.985 +-0.004

y-projection:

emit=0.161 +- 0.001 mm*mrad, emit_norm=3.150 +- 0.012 mm*mrad

beta=0.734 +- 0.001, alpha=-0.607 +-0.002

88 mA



44mA


y-projection:

emit=0.920 +- 0.023 mm*mrad, emit_norm=17.998 +- 0.442 mm*mrad

beta=0.630 +- 0.010, alpha=-1.956 +-0.031

22 mA


11 mA


y-projection:

emit=0.085 +- 0.005 mm*mrad, emit_norm=1.656 +- 0.098 mm*mrad

beta=0.321 +- 0.006, alpha=-1.349 +-0.029

Summary

(Normalized Emittance)=(Relativistic Factor) *(beta) * (Geometrical Emittance)

[math]\displaystyle{ \epsilon_{x}^{\mbox{Norm}} = \gamma_{rel} \beta \epsilon_{x}^{\mbox{geom}} }[/math]

[math]\displaystyle{ \beta=v/c }[/math] is almost one for 10 MeV


Peak Beam current [math]\displaystyle{ \epsilon_{nx} }[/math] [math]\displaystyle{ \epsilon_{ny} }[/math]
(mA) (mm*mrad) mm*mrad
0.102 29 72
20 4.8 13.7
40 5.2 14.3
80 2.2 5.5
163 4.6 5.8