Difference between revisions of "TDC Calibration"

From New IAC Wiki
Jump to navigation Jump to search
Line 11: Line 11:
 
Range of all three TDCs was set to 50nsx100.  
 
Range of all three TDCs was set to 50nsx100.  
  
=ORTEC 567-02 and ADC 1D=
+
=ORTEC TAC 567-02 and ADC 1D=
  
 
[[Image:567_02.jpg |500px]]
 
[[Image:567_02.jpg |500px]]
  
=ORTEC 567-01 and ADC 1A=
+
=ORTEC TAC 567-01 and ADC 1A=
  
 
[[Image:567_01.jpg |500px]]
 
[[Image:567_01.jpg |500px]]
  
=ORTEC 566-01 and  ADC 1C=
+
=ORTEC TAC 566-01 and  ADC 1C=
  
 
[[Image:566_01.jpg |500px]]
 
[[Image:566_01.jpg |500px]]

Revision as of 03:30, 7 November 2008

We used:

-detector A to calibrate all the TDCs;

-LeCroy octal discriminator to produce standard negative pulses;

-passive delay line to calibrate TDCs in the region < 100 ns;

-delay amplifier TENNELEC TC215 to calibrate TDCs in the region > 250 ns;

Range of all three TDCs was set to 50nsx100.

ORTEC TAC 567-02 and ADC 1D

567 02.jpg

ORTEC TAC 567-01 and ADC 1A

567 01.jpg

ORTEC TAC 566-01 and ADC 1C

566 01.jpg