Difference between revisions of "CHIPS Run Feb-2008"

From New IAC Wiki
Jump to navigation Jump to search
Line 27: Line 27:
  
 
[[Image:BeamProfile_2-26-08_3.jpg | 400 px]]
 
[[Image:BeamProfile_2-26-08_3.jpg | 400 px]]
 +
 +
==Beam on ==
 +
 +
3:20:
 +
 +
Turned beam on zero degree port and NaI lights up whan positioned just after the beam pipe window.
 +
 +
Must be gammas  Rate = 10 count/sec.  Anode amplified (x64) then discriminated and put into a scaler. Threshold  on discriminator is 2 Volts and results in zero counts when beam off.  NaI PMT at 1500 volts.

Revision as of 22:27, 26 February 2008

2-25-08

Configure 25 Mev Beam line

NaI Detector

PMT HV 1500 Co-60 1microCi source: 1MeV band at about 55mV
Co60 NaI 1500HV.jpg

PMT HV 1500 no source, noisey tube
Background NaI 1500HV.png

Beam Profile

1.5 inch FWHM beam profile measured at 8 mA peak current. The units on the beam scan is in inches. so the grid is 2inch by 2 inch (5cm by 5cm). Frequency is 60 Hz.

BeamProfile 2-26-08 1.jpg

_________________________________ 4 mA peak current


400 px

_______________________________________________

400 px

Beam on

3:20:

Turned beam on zero degree port and NaI lights up whan positioned just after the beam pipe window.

Must be gammas Rate = 10 count/sec. Anode amplified (x64) then discriminated and put into a scaler. Threshold on discriminator is 2 Volts and results in zero counts when beam off. NaI PMT at 1500 volts.