Need to know.  
+
We need to keep the Power below one 1 Watt onto the PbF2 crystal. We can achieve this by setting the beam parameters of the 25-MeV LINAC (cf. http://www.iac.isu.edu/facilities/25MeV_hall.html).  Here we have:  P = E * I  [P] Watts, [E] MeV, and [I] uA
 +
 
 +
beam energy  20 MeV
 +
Rep Rate:    10 Hz
 +
I_peak:      50 mA/puls
 +
pulse width: 0.1 us
 +
 
 +
 
 +
We find.  Time on for one second is 1^{-5}s or 1 us. and Time off is 1 - 5us = 0.999999 s
 +
The average current
 +
 
 +
0.999999 * 0 + 0.000001*50mA = 0.05 uA.
 +
 
 +
P = (20 MeV) * (0.05 uA) = 1 Watt.
 +
 
 +
 
 +
 
  
 
   a. beam current and rep rate
 
   a. beam current and rep rate

Revision as of 18:39, 20 November 2007

Radiation exposure tests in December for Hall A.

We will have three days of beamtime (Dec 3-5). We will use the the 25-MeV linac to direct an electron beam upon the PbF2 crystals. The lead Flouride Crystals will be placed on a translator table so that each crystal can moved into the beam. The crystals are 30x30x185 mm^3 and are 1290g each.


We need to keep the Power below one 1 Watt onto the PbF2 crystal. We can achieve this by setting the beam parameters of the 25-MeV LINAC (cf. http://www.iac.isu.edu/facilities/25MeV_hall.html). Here we have: P = E * I [P] Watts, [E] MeV, and [I] uA

beam energy  20 MeV
Rep Rate:    10 Hz
I_peak:      50 mA/puls
pulse width: 0.1 us


We find. Time on for one second is 1^{-5}s or 1 us. and Time off is 1 - 5us = 0.999999 s The average current

0.999999 * 0 + 0.000001*50mA = 0.05 uA.

P = (20 MeV) * (0.05 uA) = 1 Watt.



 a. beam current and rep rate
 b. total time of exposure for each individual crystals

o Day 1: 5 hours of beamtime

 1a) a first irradiation(blocks 1~4) ~5 hours Day 1
 2a) do a transparency measurement (blocks 1~4) ~3 hours Day 1
 3a) do a block (blocks 1~4) curing exposure to a blue light 24 hours day 1 day 

o Day 2: 5 hours of beamtime

 1b) a first irradiation(blocks 5~8) ~5 hours Day 2
 2b) do a transparency measurement (blocks 5~8) ~3 hours Day 2
 3b) do a block (blocks 5~8 curing exposure to a blue light 24 hours Day 2 and Day 3


o Day 3: 5 hours + 5 hours of beamtime.

 4a) repeat a transparency measurement (blocks 1~4) ~3hours day 3
 5a) do a second irradiation (blocks 1~4) ~5 hours day 3
 3a) do a transparency measurement (blocks 1~4) ~3 hours day 3
 4b) repeat a transparency measurement (blocks 5~8) ~3 hours day 3
 5b) do a second irradiation (blocks 5~8) ~5 hours day 3 at the end of the day


Day 4 and Day 5 will be used to more curing and transparency measurements.



The specifics are listed below RunPlan:

PbFl_RadTests-12-3-07

PbFl_RadTests-12-4-07

PbFl_RadTests-12-5-07


References:

NIM articles:

File:Achenbach NIM A416 1998 pg357.pdf


Private Communications

File:FEL IrradiationTests 8-25-07.pdf
File:TransmissionTestProcedure J.Roche 10-15-07.pdf
File:CalorimeterHeatingEquations.pdf