Difference between revisions of "MiniDC PreampOutput 10 10 07"
Jump to navigation
Jump to search
Line 30: | Line 30: | ||
[http://iac.isu.edu/mediawiki/index.php/Mini-prototype Back to Mini prototype page] | [http://iac.isu.edu/mediawiki/index.php/Mini-prototype Back to Mini prototype page] | ||
+ | |||
+ | |||
+ | [[Image:MINIDC_Sense-Preamp_Output_10-10-07_AmpTrig1.jpg | 200 px]] | ||
+ | |||
+ | [[Image:MINIDC_Sense-Preamp_Output_10-10-07_AmpTrig2.jpg | 200 px]] | ||
+ | |||
+ | [[Image:MINIDC_Sense-Preamp_Output_10-10-07_AmpTrig3.jpg | 200 px]] | ||
+ | |||
+ | [[Image:MINIDC_Sense-Preamp_Output_10-10-07_AmpTrig4.jpg | 200 px]] | ||
+ | |||
+ | [[Image:MINIDC_Sense-Preamp_Output_10-10-07_AmpTrig5.jpg | 200 px]] | ||
+ | |||
+ | [[Image:MINIDC_Sense-Preamp_Output_10-10-07_AmpTrig6.jpg | 200 px]] | ||
+ | |||
+ | [[Image:MINIDC_Sense-Preamp_Output_10-10-07_AmpTrig7.jpg | 200 px]] | ||
+ | |||
+ | [[Image:MINIDC_Sense-Preamp_Output_10-10-07_AmpTrig8.jpg | 200 px]] |
Revision as of 00:37, 16 October 2007
Drift chamber High voltage settings
HV settings | |
Wire | Volts |
Sense | 1800 |
Field | -800 |
Guard | 800 |
I should have set the Field and Guard wires to 900 volts according to Mac's e-mail reporting his Garfield simulation results. Next time.
Scope pictures triggering on sense wire C2
I set the scope threshold to below -14 mV based on previous measurements.
Scope pictures triggering on preamp output (pin 2)
I found that a trigger threshold of -100 mV discriminated well between noise and pulses consistent with previous ionization events.