Difference between revisions of "MiniDC PreampOutput 10 10 07"
Jump to navigation
Jump to search
Line 1: | Line 1: | ||
− | + | Drift chamber High voltage settings | |
− | |||
{| border="1" |cellpadding="20" cellspacing="0 | {| border="1" |cellpadding="20" cellspacing="0 | ||
|colspan= "2" | HV settings | |colspan= "2" | HV settings | ||
Line 12: | Line 11: | ||
|Guard || 800 | |Guard || 800 | ||
|} | |} | ||
+ | |||
+ | =Scope pictures triggering on sense wire C2 = | ||
+ | I set the scope threshold to below -14 mV based on previous measurements. | ||
+ | |||
+ | |||
+ | =Scope pictures triggering on preamp output (pin 2) = | ||
+ | |||
+ | I found that a trigger threshold of -100 mV discriminated well between noise and pulses consistent with previous ionization events. | ||
[http://iac.isu.edu/mediawiki/index.php/Mini-prototype Back to Mini prototype page] | [http://iac.isu.edu/mediawiki/index.php/Mini-prototype Back to Mini prototype page] |
Revision as of 03:31, 12 October 2007
Drift chamber High voltage settings
HV settings | |
Wire | Volts |
Sense | 1800 |
Field | -800 |
Guard | 800 |
Scope pictures triggering on sense wire C2
I set the scope threshold to below -14 mV based on previous measurements.
Scope pictures triggering on preamp output (pin 2)
I found that a trigger threshold of -100 mV discriminated well between noise and pulses consistent with previous ionization events.