[[File:Qweak_R1_ChargeCollector_w_VFATLabels.png | 200 px]]
  
 
The Qweak charge collector has two layers of copper strips. The bottom layer is 400 <math>\mu m</math> wide and is in the shape of concentric circles centered around the beam pipe.  The strips start at a radius of 7 cm and extend out to 31.4 cm.  This should cover the expected angular range of the elastically scattered electrons in the Qweak experiment.  The top layer contains copper strips that are 100 <math>\mu m</math> wide and correspond to the "y" direction in the monte carlo.  The coordinate system is a hybrid between cartesian and spherical.
 
The Qweak charge collector has two layers of copper strips. The bottom layer is 400 <math>\mu m</math> wide and is in the shape of concentric circles centered around the beam pipe.  The strips start at a radius of 7 cm and extend out to 31.4 cm.  This should cover the expected angular range of the elastically scattered electrons in the Qweak experiment.  The top layer contains copper strips that are 100 <math>\mu m</math> wide and correspond to the "y" direction in the monte carlo.  The coordinate system is a hybrid between cartesian and spherical.
Line 8: Line 10:
 
The output connectors should be mounted on the backside (below the bottom layer) and several theodolite targets are also etched to help with detector alignment.
 
The output connectors should be mounted on the backside (below the bottom layer) and several theodolite targets are also etched to help with detector alignment.
  
Files: [http://brems.iac.isu.edu/~julian/Qweak_9.3/ Gerber files, Text files, PDF files and EPS files]
+
Files: [http://brems.iac.isu.edu/~julian/NEW_QWEAK/new_qweak8.5/ Gerber files, Text files, PDF files and EPS files]
  
 +
[[Image:Qweak_bottomLayer_V8.5.pdf]]
  
 
==Bottom Copper Layer==
 
==Bottom Copper Layer==
Line 36: Line 39:
  
 
[[Image:pad_connector.jpg | 500px]]
 
[[Image:pad_connector.jpg | 500px]]
 +
[[Image:pad_connectorCERNProblem.jpg | 500px]]
  
  
 
[[Image:pad_connector_zoom.jpg | 500px]]
 
[[Image:pad_connector_zoom.jpg | 500px]]
 +
 +
==CERN production==
 +
 +
 +
[[Image:CERN_Qweak_Charge_Collector_1.jpg | 500px]][[Image:CERN_Qweak_Charge_Collector_2.jpg | 500px]]
  
 
= Preliminary design 6/07 =
 
= Preliminary design 6/07 =
Line 51: Line 60:
 
[[Image:BottomHalfofChargeColecotr.jpg | 500 px]]<br>
 
[[Image:BottomHalfofChargeColecotr.jpg | 500 px]]<br>
  
==A close up view of the 130 pin pads which will be used to mount connectors for the [http://www.iac.isu.edu/mediawiki/index.php/VFAT_readout_electronics VFAT cards].==
+
==A close up view of the 130 pin pads which will be used to mount connectors for the [[VFAT_readout_electronics]].==
 +
 
 +
[https://panasonic-denko.co.jp/ac/e/control/connector/base-base/p5ks/number/index.jsp  P5KS 130 pin 6.5 mm Panasonic connectors]
 +
 
 +
[https://panasonic-denko.co.jp/ac/e/control/connector/base-base/specdetail/index.jsp?part_no=AXK5SA3277YG Socket]
 +
 
 +
[https://panasonic-denko.co.jp/ac/e/control/connector/base-base/specdetail/index.jsp?part_no=AXK6SA3677YG Header]
 +
 
 +
[https://panasonic-denko.co.jp/ac/e_download/control/connector/base-base/catalog/con_eng_p5k.pdf SPEC sheet]
  
 
We need to have plated holes which allow the copper lines to pass through to the very last layer where the copper pads for the 130 pin connector are.
 
We need to have plated holes which allow the copper lines to pass through to the very last layer where the copper pads for the 130 pin connector are.
Line 61: Line 78:
 
The main web page for the connectors is at
 
The main web page for the connectors is at
 
[http://www.erni.com/smcfront.htd?&pb-id=wdf2a56609138ecf524dd04bd6e12bcefda8624884a] <br>
 
[http://www.erni.com/smcfront.htd?&pb-id=wdf2a56609138ecf524dd04bd6e12bcefda8624884a] <br>
 +
 +
[[Image:ChargeColecotr_coordinates.png | 600 px]]
 +
 +
{| border="3"  cellpadding="20" cellspacing="0"
 +
| Point || Coordinates (mm)
 +
|-
 +
|A || 60.000; 0
 +
|-
 +
|B || 26.400; 16.500
 +
|-
 +
|C ||  3.450; 88.500
 +
|-
 +
|D ||  -17.850; 246.750
 +
|-
 +
|E ||  60.000; 295.050
 +
|-
 +
|F ||  138.000; 246.900
 +
|-
 +
|G || 116.550; 88.350
 +
|-
 +
|H || 93.450; 16.350
 +
|}
 +
 +
  
 
[[Taget3001_GerberFileProblem]]
 
[[Taget3001_GerberFileProblem]]
[http://www.ibfriedrich.com/english/index.htm  Target 3001 design software]
+
 
 +
=Software tools=
 +
[http://www.ibfriedrich.com/english/index.htm  Target 3001 design software]<br>
 +
[http://www.downstreamtech.com/CAM350.html CAM350]
 +
 
 +
 
 +
 
 +
[http://www.iac.isu.edu/mediawiki/index.php/Ionization_Chamber Go back]

Latest revision as of 15:30, 22 January 2010

The charge collector for the Qweak chambers will use concentric strips aligned with the electron scattering polar angle ([math]\theta[/math]: the Z-axis is pointing along the beam direction), shown as the top layer below in read, and radial strips (shown as the bottom layer in blue). The radial strip are not aligned with phi due to the expected phi symmetry. Six, 130 pin output connectors will be used. Two of the pins are ground and 128 carry signals from individual strips.


Latest Design 12/07

Qweak R1 ChargeCollector w VFATLabels.png

The Qweak charge collector has two layers of copper strips. The bottom layer is 400 [math]\mu m[/math] wide and is in the shape of concentric circles centered around the beam pipe. The strips start at a radius of 7 cm and extend out to 31.4 cm. This should cover the expected angular range of the elastically scattered electrons in the Qweak experiment. The top layer contains copper strips that are 100 [math]\mu m[/math] wide and correspond to the "y" direction in the monte carlo. The coordinate system is a hybrid between cartesian and spherical.

The output connectors should be mounted on the backside (below the bottom layer) and several theodolite targets are also etched to help with detector alignment.

Files: Gerber files, Text files, PDF files and EPS files

File:Qweak bottomLayer V8.5.pdf

Bottom Copper Layer

BOTTOM copper.jpg


Internal radius: 69.7mm

External radius: 313.9mm

Circular lines width: 0.4mm

Top Copper Layer

line width: 0.1mm


TOP copper.jpg

PDF file: Media:QweakGEMchargeCollectorTop1.pdf

Gerber file: Media:Qweak_19.8.7.Top.txt

Connector pads view

Pad connector.jpg Pad connectorCERNProblem.jpg


Pad connector zoom.jpg

CERN production

CERN Qweak Charge Collector 1.jpgCERN Qweak Charge Collector 2.jpg

Preliminary design 6/07

The PCboard design software had trouble displaying the entire readout board at once such that you could see enough detail. The two pictures below represent the Top and Bottom halfs of the readout board. A view of the entire board can be seen using a Gerber file viewer. The picture of the top and Bottom layers can be found here ReadoutBoardGerberFiles along with links to the gerber files themselves.

The Top Half of the Charge Collector showing all six output conenctors

TopHalfofChargeColecotr.jpg

This is the bottom half of the charge collector showing only the bottom 2 output connectors

BottomHalfofChargeColecotr.jpg

A close up view of the 130 pin pads which will be used to mount connectors for the VFAT_readout_electronics.

P5KS 130 pin 6.5 mm Panasonic connectors

Socket

Header

SPEC sheet

We need to have plated holes which allow the copper lines to pass through to the very last layer where the copper pads for the 130 pin connector are.


ConnectorPads.jpg


The main web page for the connectors is at [1]

ChargeColecotr coordinates.png

Point Coordinates (mm)
A 60.000; 0
B 26.400; 16.500
C 3.450; 88.500
D -17.850; 246.750
E 60.000; 295.050
F 138.000; 246.900
G 116.550; 88.350
H 93.450; 16.350


Taget3001_GerberFileProblem

Software tools

Target 3001 design software
CAM350