Difference between revisions of "CH HPGe Log"
Jump to navigation
Jump to search
(37 intermediate revisions by the same user not shown) | |||
Line 1: | Line 1: | ||
='''Pump downs'''= | ='''Pump downs'''= | ||
− | {| class="wikitable | + | {| class="wikitable mw-collapsible" border="2" style="text-align:center;" |cellpadding="20" cellspacing="1 |
|- | |- | ||
! Date of pump down !! Components !! Reason | ! Date of pump down !! Components !! Reason | ||
|- | |- | ||
− | | 05/2021 || | + | | 05/2021 || X-Cooler and Poptop || Detector had not been used in years. |
+ | |- | ||
+ | | 07/20/2022 || Poptop || Detector temp increased (~5 degree increase in ~14 days, according to thermistor). | ||
|- | |- | ||
− | | | + | | 11/09/2022 || Poptop || Detector resolution changed dramatically from 1.4KeV to 2.1KeV then to 5KeV with what looks to be double humped peaks. |
+ | |} | ||
+ | |||
+ | ='''Maintenance'''= | ||
+ | |||
+ | {| class="wikitable mw-collapsible" border="2" style="text-align:center;" |cellpadding="20" cellspacing="1 | ||
+ | |- | ||
+ | ! Date !! Components !! Reason !! Outcome | ||
+ | |- | ||
+ | | 11/09/2022 || X-Cooler II || Ortec sent new o-ring, sieve pill, and high vacuum grease to help try and remedy issues listed above under pumpdown for 11/09/2022. || | ||
+ | |- | ||
+ | | 11/10/2022 || Poptop || Detector thermal contacts were cleaned using small amount of acetone and q-tips. Left to air dry before reassembly. Old sieve pill looks discolored but will be saved and stored as they can be refurbished via baking at IAC. || | ||
|} | |} | ||
Line 14: | Line 27: | ||
Make sure detector high voltage is turned off | Make sure detector high voltage is turned off | ||
− | Unplug pre-amp power | + | Unplug (or turn off) pre-amp power |
Unscrew detector electronics housing (8 screws total) | Unscrew detector electronics housing (8 screws total) | ||
Line 26: | Line 39: | ||
Turn multimeter to ''resistance (Ω)'' setting | Turn multimeter to ''resistance (Ω)'' setting | ||
− | Measure resistance across the white/black wires | + | Measure resistance across the white/black wires that exit the white nylon 4 pinned terminal |
Check resistance in tables below for corresponding detector temperature | Check resistance in tables below for corresponding detector temperature | ||
Line 304: | Line 317: | ||
='''Thermistor Reading Log'''= | ='''Thermistor Reading Log'''= | ||
− | {| class="wikitable | + | {| class="wikitable mw-collapsible mw-collapsed" border="2" style="text-align:center;" |cellpadding="20" cellspacing="1 |
+ | |- | ||
+ | ! Date !! Time !! Thermistor Measurement (Ω) !! Detector Thermal Cycled Between Previous Measurement !! Notes | ||
+ | |- | ||
+ | | 04/13/22 || 11:45am || 241.2 || Yes || Detector casing cold to touch, has been cooling down >72hrs | ||
+ | |- | ||
+ | | 04/13/22 || 1:50pm || 229.3 || No || | ||
+ | |- | ||
+ | | 04/14/22 || 11:26am|| 176.0 || No || | ||
+ | |- | ||
+ | | 06/02/22 || 11:46am || 185.0 || No || Cooler was off for a few minutes prior due to detector being moved to other side of room | ||
+ | |- | ||
+ | | 06/16/22 || 10:00am || 176.8 || No || Measured before use of high voltage | ||
+ | |- | ||
+ | | 06/30/22 || 11:00am || 184.1 || No || Shutdown cable measurement of -11.97, change from normal of -11.87V | ||
+ | |- | ||
+ | | 07/03/22 || 12:45pm || 182.2 || No || | ||
+ | |- | ||
+ | | 07/06/22 || 11:40am || 181.7 || No || Detector warmed after this measurement in prep for pumpdown | ||
+ | |- | ||
+ | | 07/21/22 || 10:50am|| 177.2 || Yes || Poptop pumped down by IAC, chiller turned on 3:50pm 07/20/22 | ||
+ | |- | ||
+ | | 07/21/22 || 11:00am|| 176.2 || No || Repeat of above | ||
+ | |- | ||
+ | | 07/21/22 || 11:30am|| 174.7 || No || Repeat of above | ||
+ | |- | ||
+ | | 07/21/22 || 12:30pm|| 173.4 || No || Repeat of above | ||
+ | |- | ||
+ | | 07/21/22 || 1:22pm|| 172.0 || No || Repeat of above | ||
+ | |- | ||
+ | | 07/21/22 || 2:11pm|| 171.2 || No || Repeat of above | ||
+ | |- | ||
+ | | 07/21/22 || 3:30pm|| 170.3 || No || Repeat of above | ||
|- | |- | ||
− | + | | 07/25/22 || 10:10am|| 169.5 || No || Detector continued to cool over a weekend, shutdown cable = -11.99V | |
|- | |- | ||
− | | | + | | 10/27/22 || 10:55am|| 181.2 || No || Resolution of detector changed from ~1.4KeV FWHM to ~2.1KeV FWHM |
|- | |- | ||
− | | | + | | 11/01/22 || 10:48am|| 171.8 || Yes || Thermal cycle of detector in attempt to bring temp back down, attempt was successful. |
|- | |- | ||
− | | | + | | 11/03/22 || 12:16pm|| 175.1 || No || Cs137 peak was wide, poor energy resolution. HV/Preamp turned off, detector beginning to cool again. |
|- | |- | ||
− | | | + | | 11/03/22 || 1:00pm|| 170.8 || No || Preamp & HV off for 45 mins, detector cooled again. |
|- | |- | ||
− | | | + | | 11/06/22 || 12:58pm|| 184.2 || Yes || Thermal cycled detector due to >5KeV sigma in energy peaks. |
|- | |- | ||
− | | | + | | 11/07/22 || 7:45am|| 192.6 || No || Detector cooling for >36hrs. |
|- | |- | ||
− | | | + | | 11/10/22 || 3:55pm|| ---- || Yes || Began cooling detector at 3:55pm. |
|- | |- | ||
− | | | + | | 11/11/22 || 10:15am|| 168.6 || Yes || Maintenance on X-Cooler and pumpdown seems to have been successful. |
|- | |- | ||
|} | |} |
Latest revision as of 17:18, 11 November 2022
Pump downs
Date of pump down | Components | Reason |
---|---|---|
05/2021 | X-Cooler and Poptop | Detector had not been used in years. |
07/20/2022 | Poptop | Detector temp increased (~5 degree increase in ~14 days, according to thermistor). |
11/09/2022 | Poptop | Detector resolution changed dramatically from 1.4KeV to 2.1KeV then to 5KeV with what looks to be double humped peaks. |
Maintenance
Date | Components | Reason | Outcome |
---|---|---|---|
11/09/2022 | X-Cooler II | Ortec sent new o-ring, sieve pill, and high vacuum grease to help try and remedy issues listed above under pumpdown for 11/09/2022. | |
11/10/2022 | Poptop | Detector thermal contacts were cleaned using small amount of acetone and q-tips. Left to air dry before reassembly. Old sieve pill looks discolored but will be saved and stored as they can be refurbished via baking at IAC. |
How to Read HPGe Thermistor
Make sure detector high voltage is turned off
Unplug (or turn off) pre-amp power
Unscrew detector electronics housing (8 screws total)
Slide electronics housing towards the crystal end of the detector
Turn multimeter to resistance (Ω) setting
Measure resistance across the white/black wires that exit the white nylon 4 pinned terminal
Check resistance in tables below for corresponding detector temperature
Resistance and Corresponding Temperatures
100-109Ω
Thermistor Resistance (Ω) | Detector Temperature (Κ) |
---|---|
100 | 76.5780 |
101 | 77.0421 |
102 | 77.5063 |
103 | 77.9706 |
104 | 78.4352 |
105 | 78.9000 |
106 | 79.3649 |
107 | 79.8300 |
108 | 80.2952 |
109 | 80.7607 |
110-119Ω
Thermistor Resistance (Ω) | Detector Temperature (Κ) |
---|---|
110 | 81.2263 |
111 | 81.6921 |
112 | 82.1581 |
113 | 82.6242 |
114 | 83.0906 |
115 | 83.5571 |
116 | 84.0238 |
117 | 84.4906 |
118 | 84.9576 |
119 | 85.4248 |
120-129Ω
Thermistor Resistance (Ω) | Detector Temperature (Κ) |
---|---|
120 | 85.8922 |
121 | 86.3597 |
122 | 86.8275 |
123 | 87.2954 |
124 | 87.7634 |
125 | 88.2317 |
126 | 88.7001 |
127 | 89.1686 |
128 | 89.6374 |
129 | 90.1063 |
130-139Ω
Thermistor Resistance (Ω) | Detector Temperature (Κ) |
---|---|
130 | 90.5754 |
131 | 91.0446 |
132 | 91.5141 |
133 | 91.9837 |
134 | 92.4524 |
135 | 92.9234 |
136 | 93.3935 |
137 | 93.8637 |
138 | 94.3342 |
139 | 94.8048 |
140-149Ω
Thermistor Resistance (Ω) | Detector Temperature (Κ) |
---|---|
140 | 95.2755 |
141 | 95.7465 |
142 | 96.2176 |
143 | 96.6888 |
144 | 97.1603 |
145 | 97.6319 |
146 | 98.1036 |
147 | 98.5756 |
148 | 99.0477 |
149 | 99.5199 |
150-159Ω
Thermistor Resistance (Ω) | Detector Temperature (Κ) |
---|---|
150 | 99.9923 |
151 | 100.4649 |
152 | 100.9377 |
153 | 101.4106 |
154 | 101.8836 |
155 | 102.3569 |
156 | 102.8303 |
157 | 103.3038 |
158 | 103.7775 |
159 | 104.2514 |
160-169Ω
Thermistor Resistance (Ω) | Detector Temperature (Κ) |
---|---|
160 | 104.7254 |
161 | 105.1996 |
162 | 105.6740 |
163 | 106.1485 |
164 | 106.6232 |
165 | 107.0980 |
166 | 107.5730 |
167 | 108.0482 |
168 | 108.5235 |
169 | 108.9989 |
170-179Ω
Thermistor Resistance (Ω) | Detector Temperature (Κ) |
---|---|
170 | 109.4746 |
171 | 109.9503 |
172 | 110.4263 |
173 | 110.9024 |
174 | 111.3786 |
175 | 111.8550 |
176 | 112.3316 |
177 | 112.8083 |
178 | 113.2851 |
179 | 113.7622 |
180-189Ω
Thermistor Resistance (Ω) | Detector Temperature (Κ) |
---|---|
180 | 114.2393 |
181 | 114.7167 |
182 | 115.1941 |
183 | 115.6718 |
184 | 116.1496 |
185 | 116.6275 |
186 | 117.1056 |
187 | 117.5838 |
188 | 118.0622 |
189 | 118.5408 |
190-199Ω
Thermistor Resistance (Ω) | Detector Temperature (Κ) |
---|---|
190 | 119.0194 |
191 | 119.4983 |
192 | 119.9773 |
193 | 120.4564 |
194 | 120.9357 |
195 | 121.4152 |
196 | 121.8947 |
197 | 122.3745 |
198 | 122.8544 |
199 | 123.3344 |
Thermistor Reading Log
Date | Time | Thermistor Measurement (Ω) | Detector Thermal Cycled Between Previous Measurement | Notes |
---|---|---|---|---|
04/13/22 | 11:45am | 241.2 | Yes | Detector casing cold to touch, has been cooling down >72hrs |
04/13/22 | 1:50pm | 229.3 | No | |
04/14/22 | 11:26am | 176.0 | No | |
06/02/22 | 11:46am | 185.0 | No | Cooler was off for a few minutes prior due to detector being moved to other side of room |
06/16/22 | 10:00am | 176.8 | No | Measured before use of high voltage |
06/30/22 | 11:00am | 184.1 | No | Shutdown cable measurement of -11.97, change from normal of -11.87V |
07/03/22 | 12:45pm | 182.2 | No | |
07/06/22 | 11:40am | 181.7 | No | Detector warmed after this measurement in prep for pumpdown |
07/21/22 | 10:50am | 177.2 | Yes | Poptop pumped down by IAC, chiller turned on 3:50pm 07/20/22 |
07/21/22 | 11:00am | 176.2 | No | Repeat of above |
07/21/22 | 11:30am | 174.7 | No | Repeat of above |
07/21/22 | 12:30pm | 173.4 | No | Repeat of above |
07/21/22 | 1:22pm | 172.0 | No | Repeat of above |
07/21/22 | 2:11pm | 171.2 | No | Repeat of above |
07/21/22 | 3:30pm | 170.3 | No | Repeat of above |
07/25/22 | 10:10am | 169.5 | No | Detector continued to cool over a weekend, shutdown cable = -11.99V |
10/27/22 | 10:55am | 181.2 | No | Resolution of detector changed from ~1.4KeV FWHM to ~2.1KeV FWHM |
11/01/22 | 10:48am | 171.8 | Yes | Thermal cycle of detector in attempt to bring temp back down, attempt was successful. |
11/03/22 | 12:16pm | 175.1 | No | Cs137 peak was wide, poor energy resolution. HV/Preamp turned off, detector beginning to cool again. |
11/03/22 | 1:00pm | 170.8 | No | Preamp & HV off for 45 mins, detector cooled again. |
11/06/22 | 12:58pm | 184.2 | Yes | Thermal cycled detector due to >5KeV sigma in energy peaks. |
11/07/22 | 7:45am | 192.6 | No | Detector cooling for >36hrs. |
11/10/22 | 3:55pm | ---- | Yes | Began cooling detector at 3:55pm. |
11/11/22 | 10:15am | 168.6 | Yes | Maintenance on X-Cooler and pumpdown seems to have been successful. |