CH HPGe Log: Difference between revisions
Jump to navigation
Jump to search
| (86 intermediate revisions by the same user not shown) | |||
| Line 1: | Line 1: | ||
='''Pump downs'''= | |||
='''100-109Ω'''= | {| class="wikitable mw-collapsible" border="2" style="text-align:center;" |cellpadding="20" cellspacing="1 | ||
|- | |||
! Date of pump down !! Components !! Reason | |||
|- | |||
| 05/2021 || X-Cooler and Poptop || Detector had not been used in years. | |||
|- | |||
| 07/20/2022 || Poptop || Detector temp increased (~5 degree increase in ~14 days, according to thermistor). | |||
|- | |||
| 11/09/2022 || Poptop || Detector resolution changed dramatically from 1.4KeV to 2.1KeV then to 5KeV with what looks to be double humped peaks. | |||
|} | |||
='''Maintenance'''= | |||
{| class="wikitable mw-collapsible" border="2" style="text-align:center;" |cellpadding="20" cellspacing="1 | |||
|- | |||
! Date !! Components !! Reason !! Outcome | |||
|- | |||
| 11/09/2022 || X-Cooler II || Ortec sent new o-ring, sieve pill, and high vacuum grease to help try and remedy issues listed above under pumpdown for 11/09/2022. || | |||
|- | |||
| 11/10/2022 || Poptop || Detector thermal contacts were cleaned using small amount of acetone and q-tips. Left to air dry before reassembly. Old sieve pill looks discolored but will be saved and stored as they can be refurbished via baking at IAC. || | |||
|} | |||
='''How to Read HPGe Thermistor'''= | |||
Make sure detector high voltage is turned off | |||
Unplug (or turn off) pre-amp power | |||
Unscrew detector electronics housing (8 screws total) | |||
[[File:CH HPGe Electronics Casing Screws.jpeg|thumb|none|Arrows pointing to screws on detector electronics housing]] | |||
Slide electronics housing towards the crystal end of the detector | |||
[[File:CH HPGe Electronics Casing Slide.jpeg|thumb|none|Slide the sectioned area in the direction of the arrow to expose the thermistor terminals]] | |||
Turn multimeter to ''resistance (Ω)'' setting | |||
Measure resistance across the white/black wires that exit the white nylon 4 pinned terminal | |||
Check resistance in tables below for corresponding detector temperature | |||
='''Resistance and Corresponding Temperatures'''= | |||
=='''100-109Ω'''== | |||
{| class="wikitable sortable mw-collapsible mw-collapsed" border="2" style="text-align:center;" |cellpadding="20" cellspacing="1 | {| class="wikitable sortable mw-collapsible mw-collapsed" border="2" style="text-align:center;" |cellpadding="20" cellspacing="1 | ||
| Line 27: | Line 72: | ||
|} | |} | ||
='''110-119Ω'''= | =='''110-119Ω'''== | ||
{| class="wikitable sortable mw-collapsible mw-collapsed" border="2" style="text-align:center;" |cellpadding="20" cellspacing="1 | {| class="wikitable sortable mw-collapsible mw-collapsed" border="2" style="text-align:center;" |cellpadding="20" cellspacing="1 | ||
| Line 54: | Line 99: | ||
|} | |} | ||
='''120-129Ω'''= | =='''120-129Ω'''== | ||
{| class="wikitable sortable mw-collapsible mw-collapsed" border="2" style="text-align:center;" |cellpadding="20" cellspacing="1 | |||
|- | |||
! Thermistor Resistance (Ω) !! Detector Temperature (Κ) | |||
|- | |||
| 120 || 85.8922 | |||
|- | |||
| 121 || 86.3597 | |||
|- | |||
| 122 || 86.8275 | |||
|- | |||
| 123 || 87.2954 | |||
|- | |||
| 124 || 87.7634 | |||
|- | |||
| 125 || 88.2317 | |||
|- | |||
| 126 || 88.7001 | |||
|- | |||
| 127 || 89.1686 | |||
|- | |||
| 128 || 89.6374 | |||
|- | |||
| 129 || 90.1063 | |||
|} | |||
=='''130-139Ω'''== | |||
{| class="wikitable sortable mw-collapsible mw-collapsed" border="2" style="text-align:center;" |cellpadding="20" cellspacing="1 | {| class="wikitable sortable mw-collapsible mw-collapsed" border="2" style="text-align:center;" |cellpadding="20" cellspacing="1 | ||
| Line 60: | Line 132: | ||
! Thermistor Resistance (Ω) !! Detector Temperature (Κ) | ! Thermistor Resistance (Ω) !! Detector Temperature (Κ) | ||
|- | |- | ||
| | | 130 || 90.5754 | ||
|- | |- | ||
| | | 131 || 91.0446 | ||
|- | |- | ||
| | | 132 || 91.5141 | ||
|- | |- | ||
| | | 133 || 91.9837 | ||
|- | |- | ||
| | | 134 || 92.4524 | ||
|- | |- | ||
| | | 135 || 92.9234 | ||
|- | |- | ||
| | | 136 || 93.3935 | ||
|- | |- | ||
| | | 137 || 93.8637 | ||
|- | |- | ||
| | | 138 || 94.3342 | ||
|- | |- | ||
| | | 139 || 94.8048 | ||
|} | |} | ||
=''' | =='''140-149Ω'''== | ||
{| class="wikitable sortable mw-collapsible mw-collapsed" border="2" style="text-align:center;" |cellpadding="20" cellspacing="1 | {| class="wikitable sortable mw-collapsible mw-collapsed" border="2" style="text-align:center;" |cellpadding="20" cellspacing="1 | ||
|- | |- | ||
! Thermistor Resistance ( | ! Thermistor Resistance (Ω) !! Detector Temperature (Κ) | ||
|- | |- | ||
| | | 140 || 95.2755 | ||
|- | |- | ||
| | | 141 || 95.7465 | ||
|- | |- | ||
| | | 142 || 96.2176 | ||
|- | |- | ||
| | | 143 || 96.6888 | ||
|- | |- | ||
| | | 144 || 97.1603 | ||
|- | |- | ||
| | | 145 || 97.6319 | ||
|- | |- | ||
| | | 146 || 98.1036 | ||
|- | |- | ||
| | | 147 || 98.5756 | ||
|- | |- | ||
| | | 148 || 99.0477 | ||
|- | |- | ||
| | | 149 || 99.5199 | ||
|} | |} | ||
=''' | =='''150-159Ω'''== | ||
{| class="wikitable sortable mw-collapsible mw-collapsed" border="2" style="text-align:center;" |cellpadding="20" cellspacing="1 | {| class="wikitable sortable mw-collapsible mw-collapsed" border="2" style="text-align:center;" |cellpadding="20" cellspacing="1 | ||
|- | |- | ||
! Thermistor Resistance ( | ! Thermistor Resistance (Ω) !! Detector Temperature (Κ) | ||
|- | |- | ||
| | | 150 || 99.9923 | ||
|- | |- | ||
| | | 151 || 100.4649 | ||
|- | |- | ||
| | | 152 || 100.9377 | ||
|- | |- | ||
| | | 153 || 101.4106 | ||
|- | |- | ||
| | | 154 || 101.8836 | ||
|- | |- | ||
| | | 155 || 102.3569 | ||
|- | |- | ||
| | | 156 || 102.8303 | ||
|- | |- | ||
| | | 157 || 103.3038 | ||
|- | |- | ||
| | | 158 || 103.7775 | ||
|- | |- | ||
| | | 159 || 104.2514 | ||
|} | |} | ||
=''' | =='''160-169Ω'''== | ||
{| class="wikitable sortable mw-collapsible mw-collapsed" border="2" style="text-align:center;" |cellpadding="20" cellspacing="1 | {| class="wikitable sortable mw-collapsible mw-collapsed" border="2" style="text-align:center;" |cellpadding="20" cellspacing="1 | ||
|- | |- | ||
! Thermistor Resistance ( | ! Thermistor Resistance (Ω) !! Detector Temperature (Κ) | ||
|- | |- | ||
| | | 160 || 104.7254 | ||
|- | |- | ||
| | | 161 || 105.1996 | ||
|- | |- | ||
| | | 162 || 105.6740 | ||
|- | |- | ||
| | | 163 || 106.1485 | ||
|- | |- | ||
| | | 164 || 106.6232 | ||
|- | |- | ||
| | | 165 || 107.0980 | ||
|- | |- | ||
| | | 166 || 107.5730 | ||
|- | |- | ||
| | | 167 || 108.0482 | ||
|- | |- | ||
| | | 168 || 108.5235 | ||
|- | |- | ||
| | | 169 || 108.9989 | ||
|} | |} | ||
=''' | =='''170-179Ω'''== | ||
{| class="wikitable sortable mw-collapsible mw-collapsed" border="2" style="text-align:center;" |cellpadding="20" cellspacing="1 | {| class="wikitable sortable mw-collapsible mw-collapsed" border="2" style="text-align:center;" |cellpadding="20" cellspacing="1 | ||
|- | |- | ||
! Thermistor Resistance ( | ! Thermistor Resistance (Ω) !! Detector Temperature (Κ) | ||
|- | |- | ||
| | | 170 || 109.4746 | ||
|- | |- | ||
| | | 171 || 109.9503 | ||
|- | |- | ||
| | | 172 || 110.4263 | ||
|- | |- | ||
| | | 173 || 110.9024 | ||
|- | |- | ||
| | | 174 || 111.3786 | ||
|- | |- | ||
| | | 175 || 111.8550 | ||
|- | |- | ||
| | | 176 || 112.3316 | ||
|- | |- | ||
| | | 177 || 112.8083 | ||
|- | |- | ||
| | | 178 || 113.2851 | ||
|- | |- | ||
| | | 179 || 113.7622 | ||
|} | |} | ||
=''' | =='''180-189Ω'''== | ||
{| class="wikitable sortable mw-collapsible mw-collapsed" border="2" style="text-align:center;" |cellpadding="20" cellspacing="1 | {| class="wikitable sortable mw-collapsible mw-collapsed" border="2" style="text-align:center;" |cellpadding="20" cellspacing="1 | ||
|- | |- | ||
! Thermistor Resistance ( | ! Thermistor Resistance (Ω) !! Detector Temperature (Κ) | ||
|- | |- | ||
| | | 180 || 114.2393 | ||
|- | |- | ||
| | | 181 || 114.7167 | ||
|- | |- | ||
| | | 182 || 115.1941 | ||
|- | |- | ||
| | | 183 || 115.6718 | ||
|- | |- | ||
| | | 184 || 116.1496 | ||
|- | |- | ||
| | | 185 || 116.6275 | ||
|- | |- | ||
| | | 186 || 117.1056 | ||
|- | |- | ||
| | | 187 || 117.5838 | ||
|- | |- | ||
| | | 188 || 118.0622 | ||
|- | |- | ||
| | | 189 || 118.5408 | ||
|} | |} | ||
=''' | =='''190-199Ω'''== | ||
{| class="wikitable sortable mw-collapsible mw-collapsed" border="2" style="text-align:center;" |cellpadding="20" cellspacing="1 | {| class="wikitable sortable mw-collapsible mw-collapsed" border="2" style="text-align:center;" |cellpadding="20" cellspacing="1 | ||
|- | |- | ||
! Thermistor Resistance ( | ! Thermistor Resistance (Ω) !! Detector Temperature (Κ) | ||
|- | |- | ||
| | | 190 || 119.0194 | ||
|- | |- | ||
| | | 191 || 119.4983 | ||
|- | |- | ||
| | | 192 || 119.9773 | ||
|- | |- | ||
| | | 193 || 120.4564 | ||
|- | |- | ||
| | | 194 || 120.9357 | ||
|- | |- | ||
| | | 195 || 121.4152 | ||
|- | |- | ||
| | | 196 || 121.8947 | ||
|- | |- | ||
| | | 197 || 122.3745 | ||
|- | |- | ||
| | | 198 || 122.8544 | ||
|- | |- | ||
| | | 199 || 123.3344 | ||
|} | |} | ||
='''Thermistor Reading Log'''= | |||
{| class="wikitable mw-collapsible mw-collapsed" border="2" style="text-align:center;" |cellpadding="20" cellspacing="1 | |||
|- | |||
! Date !! Time !! Thermistor Measurement (Ω) !! Detector Thermal Cycled Between Previous Measurement !! Notes | |||
|- | |||
| 04/13/22 || 11:45am || 241.2 || Yes || Detector casing cold to touch, has been cooling down >72hrs | |||
|- | |||
| 04/13/22 || 1:50pm || 229.3 || No || | |||
|- | |||
| 04/14/22 || 11:26am|| 176.0 || No || | |||
|- | |||
| 06/02/22 || 11:46am || 185.0 || No || Cooler was off for a few minutes prior due to detector being moved to other side of room | |||
|- | |||
| 06/16/22 || 10:00am || 176.8 || No || Measured before use of high voltage | |||
|- | |||
| 06/30/22 || 11:00am || 184.1 || No || Shutdown cable measurement of -11.97, change from normal of -11.87V | |||
|- | |||
| 07/03/22 || 12:45pm || 182.2 || No || | |||
|- | |||
| 07/06/22 || 11:40am || 181.7 || No || Detector warmed after this measurement in prep for pumpdown | |||
|- | |||
| 07/21/22 || 10:50am|| 177.2 || Yes || Poptop pumped down by IAC, chiller turned on 3:50pm 07/20/22 | |||
|- | |||
| 07/21/22 || 11:00am|| 176.2 || No || Repeat of above | |||
|- | |||
| 07/21/22 || 11:30am|| 174.7 || No || Repeat of above | |||
|- | |||
| 07/21/22 || 12:30pm|| 173.4 || No || Repeat of above | |||
|- | |||
| 07/21/22 || 1:22pm|| 172.0 || No || Repeat of above | |||
|- | |||
| 07/21/22 || 2:11pm|| 171.2 || No || Repeat of above | |||
|- | |||
| 07/21/22 || 3:30pm|| 170.3 || No || Repeat of above | |||
|- | |||
| 07/25/22 || 10:10am|| 169.5 || No || Detector continued to cool over a weekend, shutdown cable = -11.99V | |||
|- | |||
| 10/27/22 || 10:55am|| 181.2 || No || Resolution of detector changed from ~1.4KeV FWHM to ~2.1KeV FWHM | |||
|- | |||
| 11/01/22 || 10:48am|| 171.8 || Yes || Thermal cycle of detector in attempt to bring temp back down, attempt was successful. | |||
|- | |||
| 11/03/22 || 12:16pm|| 175.1 || No || Cs137 peak was wide, poor energy resolution. HV/Preamp turned off, detector beginning to cool again. | |||
|- | |||
| 11/03/22 || 1:00pm|| 170.8 || No || Preamp & HV off for 45 mins, detector cooled again. | |||
|- | |||
| 11/06/22 || 12:58pm|| 184.2 || Yes || Thermal cycled detector due to >5KeV sigma in energy peaks. | |||
|- | |||
| 11/07/22 || 7:45am|| 192.6 || No || Detector cooling for >36hrs. | |||
|- | |||
| 11/10/22 || 3:55pm|| ---- || Yes || Began cooling detector at 3:55pm. | |||
|- | |||
| 11/11/22 || 10:15am|| 168.6 || Yes || Maintenance on X-Cooler and pumpdown seems to have been successful. | |||
|- | |||
|} | |||
---- | ---- | ||
[[PAS HPGe Detector]] | [[PAS HPGe Detector]] | ||
Latest revision as of 17:18, 11 November 2022
Pump downs
| Date of pump down | Components | Reason |
|---|---|---|
| 05/2021 | X-Cooler and Poptop | Detector had not been used in years. |
| 07/20/2022 | Poptop | Detector temp increased (~5 degree increase in ~14 days, according to thermistor). |
| 11/09/2022 | Poptop | Detector resolution changed dramatically from 1.4KeV to 2.1KeV then to 5KeV with what looks to be double humped peaks. |
Maintenance
| Date | Components | Reason | Outcome |
|---|---|---|---|
| 11/09/2022 | X-Cooler II | Ortec sent new o-ring, sieve pill, and high vacuum grease to help try and remedy issues listed above under pumpdown for 11/09/2022. | |
| 11/10/2022 | Poptop | Detector thermal contacts were cleaned using small amount of acetone and q-tips. Left to air dry before reassembly. Old sieve pill looks discolored but will be saved and stored as they can be refurbished via baking at IAC. |
How to Read HPGe Thermistor
Make sure detector high voltage is turned off
Unplug (or turn off) pre-amp power
Unscrew detector electronics housing (8 screws total)

Slide electronics housing towards the crystal end of the detector

Turn multimeter to resistance (Ω) setting
Measure resistance across the white/black wires that exit the white nylon 4 pinned terminal
Check resistance in tables below for corresponding detector temperature
Resistance and Corresponding Temperatures
100-109Ω
| Thermistor Resistance (Ω) | Detector Temperature (Κ) |
|---|---|
| 100 | 76.5780 |
| 101 | 77.0421 |
| 102 | 77.5063 |
| 103 | 77.9706 |
| 104 | 78.4352 |
| 105 | 78.9000 |
| 106 | 79.3649 |
| 107 | 79.8300 |
| 108 | 80.2952 |
| 109 | 80.7607 |
110-119Ω
| Thermistor Resistance (Ω) | Detector Temperature (Κ) |
|---|---|
| 110 | 81.2263 |
| 111 | 81.6921 |
| 112 | 82.1581 |
| 113 | 82.6242 |
| 114 | 83.0906 |
| 115 | 83.5571 |
| 116 | 84.0238 |
| 117 | 84.4906 |
| 118 | 84.9576 |
| 119 | 85.4248 |
120-129Ω
| Thermistor Resistance (Ω) | Detector Temperature (Κ) |
|---|---|
| 120 | 85.8922 |
| 121 | 86.3597 |
| 122 | 86.8275 |
| 123 | 87.2954 |
| 124 | 87.7634 |
| 125 | 88.2317 |
| 126 | 88.7001 |
| 127 | 89.1686 |
| 128 | 89.6374 |
| 129 | 90.1063 |
130-139Ω
| Thermistor Resistance (Ω) | Detector Temperature (Κ) |
|---|---|
| 130 | 90.5754 |
| 131 | 91.0446 |
| 132 | 91.5141 |
| 133 | 91.9837 |
| 134 | 92.4524 |
| 135 | 92.9234 |
| 136 | 93.3935 |
| 137 | 93.8637 |
| 138 | 94.3342 |
| 139 | 94.8048 |
140-149Ω
| Thermistor Resistance (Ω) | Detector Temperature (Κ) |
|---|---|
| 140 | 95.2755 |
| 141 | 95.7465 |
| 142 | 96.2176 |
| 143 | 96.6888 |
| 144 | 97.1603 |
| 145 | 97.6319 |
| 146 | 98.1036 |
| 147 | 98.5756 |
| 148 | 99.0477 |
| 149 | 99.5199 |
150-159Ω
| Thermistor Resistance (Ω) | Detector Temperature (Κ) |
|---|---|
| 150 | 99.9923 |
| 151 | 100.4649 |
| 152 | 100.9377 |
| 153 | 101.4106 |
| 154 | 101.8836 |
| 155 | 102.3569 |
| 156 | 102.8303 |
| 157 | 103.3038 |
| 158 | 103.7775 |
| 159 | 104.2514 |
160-169Ω
| Thermistor Resistance (Ω) | Detector Temperature (Κ) |
|---|---|
| 160 | 104.7254 |
| 161 | 105.1996 |
| 162 | 105.6740 |
| 163 | 106.1485 |
| 164 | 106.6232 |
| 165 | 107.0980 |
| 166 | 107.5730 |
| 167 | 108.0482 |
| 168 | 108.5235 |
| 169 | 108.9989 |
170-179Ω
| Thermistor Resistance (Ω) | Detector Temperature (Κ) |
|---|---|
| 170 | 109.4746 |
| 171 | 109.9503 |
| 172 | 110.4263 |
| 173 | 110.9024 |
| 174 | 111.3786 |
| 175 | 111.8550 |
| 176 | 112.3316 |
| 177 | 112.8083 |
| 178 | 113.2851 |
| 179 | 113.7622 |
180-189Ω
| Thermistor Resistance (Ω) | Detector Temperature (Κ) |
|---|---|
| 180 | 114.2393 |
| 181 | 114.7167 |
| 182 | 115.1941 |
| 183 | 115.6718 |
| 184 | 116.1496 |
| 185 | 116.6275 |
| 186 | 117.1056 |
| 187 | 117.5838 |
| 188 | 118.0622 |
| 189 | 118.5408 |
190-199Ω
| Thermistor Resistance (Ω) | Detector Temperature (Κ) |
|---|---|
| 190 | 119.0194 |
| 191 | 119.4983 |
| 192 | 119.9773 |
| 193 | 120.4564 |
| 194 | 120.9357 |
| 195 | 121.4152 |
| 196 | 121.8947 |
| 197 | 122.3745 |
| 198 | 122.8544 |
| 199 | 123.3344 |
Thermistor Reading Log
| Date | Time | Thermistor Measurement (Ω) | Detector Thermal Cycled Between Previous Measurement | Notes |
|---|---|---|---|---|
| 04/13/22 | 11:45am | 241.2 | Yes | Detector casing cold to touch, has been cooling down >72hrs |
| 04/13/22 | 1:50pm | 229.3 | No | |
| 04/14/22 | 11:26am | 176.0 | No | |
| 06/02/22 | 11:46am | 185.0 | No | Cooler was off for a few minutes prior due to detector being moved to other side of room |
| 06/16/22 | 10:00am | 176.8 | No | Measured before use of high voltage |
| 06/30/22 | 11:00am | 184.1 | No | Shutdown cable measurement of -11.97, change from normal of -11.87V |
| 07/03/22 | 12:45pm | 182.2 | No | |
| 07/06/22 | 11:40am | 181.7 | No | Detector warmed after this measurement in prep for pumpdown |
| 07/21/22 | 10:50am | 177.2 | Yes | Poptop pumped down by IAC, chiller turned on 3:50pm 07/20/22 |
| 07/21/22 | 11:00am | 176.2 | No | Repeat of above |
| 07/21/22 | 11:30am | 174.7 | No | Repeat of above |
| 07/21/22 | 12:30pm | 173.4 | No | Repeat of above |
| 07/21/22 | 1:22pm | 172.0 | No | Repeat of above |
| 07/21/22 | 2:11pm | 171.2 | No | Repeat of above |
| 07/21/22 | 3:30pm | 170.3 | No | Repeat of above |
| 07/25/22 | 10:10am | 169.5 | No | Detector continued to cool over a weekend, shutdown cable = -11.99V |
| 10/27/22 | 10:55am | 181.2 | No | Resolution of detector changed from ~1.4KeV FWHM to ~2.1KeV FWHM |
| 11/01/22 | 10:48am | 171.8 | Yes | Thermal cycle of detector in attempt to bring temp back down, attempt was successful. |
| 11/03/22 | 12:16pm | 175.1 | No | Cs137 peak was wide, poor energy resolution. HV/Preamp turned off, detector beginning to cool again. |
| 11/03/22 | 1:00pm | 170.8 | No | Preamp & HV off for 45 mins, detector cooled again. |
| 11/06/22 | 12:58pm | 184.2 | Yes | Thermal cycled detector due to >5KeV sigma in energy peaks. |
| 11/07/22 | 7:45am | 192.6 | No | Detector cooling for >36hrs. |
| 11/10/22 | 3:55pm | ---- | Yes | Began cooling detector at 3:55pm. |
| 11/11/22 | 10:15am | 168.6 | Yes | Maintenance on X-Cooler and pumpdown seems to have been successful. |