Difference between revisions of "NaI JD-766 Plateau"
Jump to navigation
Jump to search
Line 20: | Line 20: | ||
| Run Number || Configuration || Discriminator Threshold (mV) || Cs-137 Button Source || Counts || Run Time (Sec) || Count Rate (Hz) | | Run Number || Configuration || Discriminator Threshold (mV) || Cs-137 Button Source || Counts || Run Time (Sec) || Count Rate (Hz) | ||
|- | |- | ||
− | | 1 || Unshielded detector at +700V || 05 || No || 958 || 60 || | + | | 1 || Unshielded detector at +700V || 05 || No || 958 || 60 || 15.97 |
|- | |- | ||
− | | 2 || Unshielded detector at +700V || 15 || No || 278 || 60 || | + | | 2 || Unshielded detector at +700V || 15 || No || 278 || 60 || 4.63 |
|- | |- | ||
− | | 3 || Unshielded detector at +700V || 35 || No || 107 || 60 || | + | | 3 || Unshielded detector at +700V || 35 || No || 107 || 60 || 1.78 |
|- | |- | ||
− | | 4 || Unshielded detector at +700V || 55 || No || 85 || 60 || | + | | 4 || Unshielded detector at +700V || 55 || No || 85 || 60 || 1.42 |
|- | |- | ||
− | | 5 || Unshielded detector at +700V || 75 || No || 80 || 60 || | + | | 5 || Unshielded detector at +700V || 75 || No || 80 || 60 || 1.33 |
|- | |- | ||
− | | 6 || Unshielded detector at +700V || 95 || No || 80 || 60 || | + | | 6 || Unshielded detector at +700V || 95 || No || 80 || 60 || 1.33 |
|- | |- | ||
− | | 7 || Unshielded detector at +700V || 115 || No || 75 || 60 || | + | | 7 || Unshielded detector at +700V || 115 || No || 75 || 60 || 1.25 |
|- | |- | ||
− | | 8 || Unshielded detector at +700V || 135 || No || 78 || 60 || | + | | 8 || Unshielded detector at +700V || 135 || No || 78 || 60 || |
|- | |- | ||
| 9 || Unshielded detector at +700V || 155 || No || 82 || 60 || | | 9 || Unshielded detector at +700V || 155 || No || 82 || 60 || |
Revision as of 20:18, 19 May 2021
JD-766 Cs-137 Button Source Plateau
Timing filter amp configuration
Course gain: x6
Fine gain: at minimum
Integrate (nS): 500nS
Diff (nS): Out
Invert/Noninverted: Non-Inv
Plateau
Run Number Configuration Discriminator Threshold (mV) Cs-137 Button Source Counts Run Time (Sec) Count Rate (Hz) 1 Unshielded detector at +700V 05 No 958 60 15.97 2 Unshielded detector at +700V 15 No 278 60 4.63 3 Unshielded detector at +700V 35 No 107 60 1.78 4 Unshielded detector at +700V 55 No 85 60 1.42 5 Unshielded detector at +700V 75 No 80 60 1.33 6 Unshielded detector at +700V 95 No 80 60 1.33 7 Unshielded detector at +700V 115 No 75 60 1.25 8 Unshielded detector at +700V 135 No 78 60 9 Unshielded detector at +700V 155 No 82 60 10 Unshielded detector at +700V 175 No 77 60 11 Unshielded detector at +700V 195 No 66 60 12 Unshielded detector at +700V 215 No 68 60 13 Unshielded detector at +700V 235 No 53 60 14 Unshielded detector at +700V 255 No 59 60
Run Number Configuration Discriminator Threshold (mV) Cs-137 Button Source Counts Run Time (Sec) Count Rate (Hz) 1 Unshielded detector at +750V 05 No 2211 60 2 Unshielded detector at +750V 15 No 1125 60 3 Unshielded detector at +750V 35 No 293 60 4 Unshielded detector at +750V 55 No 105 60 5 Unshielded detector at +750V 75 No 80 60 6 Unshielded detector at +750V 95 No 97 60 7 Unshielded detector at +750V 115 No 99 60 8 Unshielded detector at +750V 135 No 89 60 9 Unshielded detector at +750V 155 No 74 60 10 Unshielded detector at +750V 175 No 80 60 11 Unshielded detector at +750V 195 No 80 60 12 Unshielded detector at +750V 215 No 64 60 13 Unshielded detector at +750V 235 No 89 60 14 Unshielded detector at +750V 255 No 69 60
Run Number Configuration Discriminator Threshold (mV) Cs-137 Button Source Counts Run Time (Sec) Count Rate (Hz) 1 Unshielded detector at +800V 05 No 3866 60 2 Unshielded detector at +800V 15 No 2356 60 3 Unshielded detector at +800V 35 No 1091 60 4 Unshielded detector at +800V 55 No 384 60 5 Unshielded detector at +800V 75 No 244 60 6 Unshielded detector at +800V 95 No 167 60 7 Unshielded detector at +800V 115 No 100 60 8 Unshielded detector at +800V 135 No 88 60 9 Unshielded detector at +800V 155 No 89 60 10 Unshielded detector at +800V 175 No 86 60 11 Unshielded detector at +800V 195 No 90 60 12 Unshielded detector at +800V 215 No 81 60 13 Unshielded detector at +800V 235 No 95 60 14 Unshielded detector at +800V 255 No 89 60
Run Number Configuration Discriminator Threshold (mV) Cs-137 Button Source Counts Run Time (Sec) Count Rate (Hz) 1 Unshielded detector at +850V 05 No 6402 60 2 Unshielded detector at +850V 15 No 3867 60 3 Unshielded detector at +850V 35 No 2058 60 4 Unshielded detector at +850V 55 No 1233 60 5 Unshielded detector at +850V 75 No 787 60 6 Unshielded detector at +850V 95 No 391 60 7 Unshielded detector at +850V 115 No 256 60 8 Unshielded detector at +850V 135 No 193 60 9 Unshielded detector at +850V 155 No 151 60 10 Unshielded detector at +850V 175 No 97 60 11 Unshielded detector at +850V 195 No 105 60 12 Unshielded detector at +850V 215 No 97 60 13 Unshielded detector at +850V 235 No 87 60 14 Unshielded detector at +850V 255 No 101 60