Difference between revisions of "LB DetLimits Thesis"
Jump to navigation
Jump to search
Line 46: | Line 46: | ||
|- | |- | ||
|} | |} | ||
+ | |||
+ | |||
+ | Below are plots of the activity ratio of the Se/Soil mixture and specific activity of the pure selenium pellet as a reference material | ||
+ | |||
+ | [[File:A Mix Div a Pure NormCorrected Se75 NoFitRestrict.png|200px]] |
Revision as of 20:36, 27 July 2018
50% Excluded
For this analysis, begin by using the first measurement of the Se-75 line and using the standard exponential decay equation to correct it and its error back to the time of beam off. Below is a table of the Se-75 (120d) corrected measurements as well as the front nickel foils and the Mn-54 data for each measurement. Note the Mn-54 analysis was weighted by the mass of the soil.
Sample | Activated (total beam time) | Counted | Fully Thesis Corrected Rate @ 1st Measurement (Hz) | |
10% Se/Soil Mixture | 5/23/17 and 5/24/17 (83 min) | 07/18/17 (16:25:44) | 3109.68 | 95.394280.08 | 131.29
10% Se/Soil Pure Witness selenium | 5/23/17 and 5/24/17 (83 min) | 8/1/17 (16:05:51) | 4161.29 | 127.636210.50 | 190.48
10% Se/Soil Front Inner Ni Foil | 5/23/17 (1h) | 5/23/17 (15:52:21) | 385955 | 23488.8425064.36 | 25868.95
10% Se/Soil Mixture Mn-54 | 5/23/17 (1h) | 07/18/17 (16:25:44) | 15.63 | 0.5617.67 | 0.63
1% Sample Se/Soil Mixture | 5/23/17 (1h) | 6/23/17 (16:39:03) | 2233.33 | 68.522675.24 | 82.08
1% Sample Se/Soil Pure Witness selenium | 5/23/17 (1h) | 7/20/17 (16:39:46) | 3840.86 | 117.815379.43 | 165.00
1% Sample Se/Soil Front Inner Ni Foil | 5/23/17 (1h) | 5/23/17 (17:59:39) | 261461 | 16045.7293170.78 | 17991.71
1% Sample Se/Soil Mixture Mn-54 | 5/23/17 (1h) | 6/23/17 (16:39:03) | 9.38 | 0.3910.05 | 0.42
0.1% Sample Se/Soil Mixture | 5/23/17 (30 min) | 6/19/17 (18:25:18) | 849.46 | 26.07993.59 | 30.49
0.1% Sample Se/Soil Pure Witness Selenium | 5/23/17 (30 min) | 07/25/17 (16:32:52) | 834.41 | 25.611201.58 | 36.88
0.1% Sample Se/Soil Front Inner Ni Foil | 5/23/17 (30 min) | 5/24/17 (14:25:39 | 118412 | 7280.06197670.18 | 12152.98
0.1% Sample Se/Soil Mn-54 | 5/23/17 (30 min) | 6/19/17 (18:25:18) | 9.38 | 0.339.96 | 0.35
Mn-54 Efficiency
A calibrated Mn-54 source was used to find the efficiency of an 834 keV line. The source was serial #J4-348, which had an activity of 9.882
on 8/01/12, so the activity on 4/18/17 was
Now converting to Hz gives
Source | Energy (keV) | Position | Expected Rate (Hz) | ROOT Window (keV) | HpGe Rate (Hz) | HpGe Detector A Efficiency (%) |
Mn-54 (J4-348) | 834 | 10cm (Det A) | 8138.05 | 3\%[829,839] | 25.87 | 0.280.32 | 0.01
Below are plots of the activity ratio of the Se/Soil mixture and specific activity of the pure selenium pellet as a reference material