Difference between revisions of "Se170063 Thin Window Analysis"
(14 intermediate revisions by 2 users not shown) | |||
Line 12: | Line 12: | ||
||Original Window ||[[File:170063 PureSe 400 640Sec OGWindow.png|100px]] || [[File:170063 PureSe 1100 1360 Seconds OGWindow.png|100px]] || [[File:170063 PureSeSpec 1875 2150Sec OGWindow.png|100px]] || [[File:170063 PureSeSpec 2650 2930Sec OGWindow.png|100px]] || [[File:170063 PureSeSpec 3400 3690Sec OGWindow.png|100px]] || [[File:170063 PureSeSpec 4120 4400Sec OGWindow.png|100px]] || [[File:170063 PureSeSpec 4840 5130 OGWindow.png|100px]] | ||Original Window ||[[File:170063 PureSe 400 640Sec OGWindow.png|100px]] || [[File:170063 PureSe 1100 1360 Seconds OGWindow.png|100px]] || [[File:170063 PureSeSpec 1875 2150Sec OGWindow.png|100px]] || [[File:170063 PureSeSpec 2650 2930Sec OGWindow.png|100px]] || [[File:170063 PureSeSpec 3400 3690Sec OGWindow.png|100px]] || [[File:170063 PureSeSpec 4120 4400Sec OGWindow.png|100px]] || [[File:170063 PureSeSpec 4840 5130 OGWindow.png|100px]] | ||
|- | |- | ||
− | ||Expanded Window || [[File:170063 PureSe Spect 400 640Sec ExpWindow.png|100px]] || [[File:170063 | + | ||Expanded Window || [[File:170063 PureSe Spect 400 640Sec ExpWindow.png|100px]] || [[File:170063 PureSeSpec 1100 1360 ExpWindow 1 11.png|100px]] || [[File:170063 PureSeSpec 1875 2150 ExpWidnow 1 11.png|100px]] || [[File:170063 PureSeSpec 2650 2930 ExpWindow 1 11.png|100px]] ||[[File:170063 PureSeSpec 3400 3940 ExpWindow 1 11.png|100px]] || [[File:170063 PureSeSpec 4120 4400 ExpWindow 1 11.png|100px]] || [[File:170063 PureSeSpec 4840 5130 ExpWindow 1 11.png|100px]] |
|- | |- | ||
||Maximum of Histogram|| 260980 || 241756 || 220348 || 191476 || 165549 || 139528 || 130930 | ||Maximum of Histogram|| 260980 || 241756 || 220348 || 191476 || 165549 || 139528 || 130930 | ||
Line 54: | Line 54: | ||
||Original Window || [[File:170063 MixSpec 0 300 OGWindow.png|100px]] || [[File:170063 MixSpec 730 1020Sec OGWindow.png|100px]] || [[File:170063 MixSpec 1480 1775Sec OGWindow.png|100px]] || [[File:170063 MixSpec 2250 2550Sec OGWindow.png|100px]] || [[File:170063 MixSpec 3050 3300Sec OGWindow.png|100px]] || [[File:170063 MixSpec 3775 4050Sec OGWindow.png|100px]] || [[File:170063 MixSpec 4480 4770Sec OGWindow.png|100px]] | ||Original Window || [[File:170063 MixSpec 0 300 OGWindow.png|100px]] || [[File:170063 MixSpec 730 1020Sec OGWindow.png|100px]] || [[File:170063 MixSpec 1480 1775Sec OGWindow.png|100px]] || [[File:170063 MixSpec 2250 2550Sec OGWindow.png|100px]] || [[File:170063 MixSpec 3050 3300Sec OGWindow.png|100px]] || [[File:170063 MixSpec 3775 4050Sec OGWindow.png|100px]] || [[File:170063 MixSpec 4480 4770Sec OGWindow.png|100px]] | ||
|- | |- | ||
− | ||Expanded Window || [[File:170063 MixSpec 0 | + | ||Expanded Window || [[File:170063 MixSpec 0 300 ExpWindow 1 11.png|100px]] || [[File:170063 MixSpec 730 1020 ExpWindow 1 11.png|100px]] || [[File:170063 MixSpec 1480 1775 ExpWindow 1 11.png|100px]] || [[File:170063 MixSpec 2250 2550 ExpWindow 1 11.png|100px]] || [[File:170063 MixSpec 3050 3300 ExpWindow 1 11.png|100px]] || [[File:170063 MixSpec 3775 4050 ExpWindow 1 11.png|100px]] ||[[File:170063 MixSpec 4480 4770 ExpWindow 1 11.png|100px]] |
|- | |- | ||
||Maximum of Histogram|| 173289 || 155770 || 135740 || 124656 || 88582 || 80993 || 77781 | ||Maximum of Histogram|| 173289 || 155770 || 135740 || 124656 || 88582 || 80993 || 77781 | ||
Line 446: | Line 446: | ||
{| border="3" cellpadding="5" cellspacing="0" | {| border="3" cellpadding="5" cellspacing="0" | ||
− | || || Run 1 || Run 2 || Run 3 || Run 4 || Run 5 || Run 6 || | + | || || Run 1 || Run 2 || Run 3 || Run 4 || Run 5 || Run 6 || |
|- | |- | ||
− | || Original Measurement (Hz) || 1961.89 +/- 216.27 | + | || Original Measurement (Hz) || 1961.89 +/- 216.27 || 1713.22 +/- 175.05 || 1510.32 +/- 151.17 || 1274.03 +/- 130.1 || 1036.14 +/- 131.17 || 913.17 +/- 105.31 |
|- | |- | ||
− | ||n+1 Measurement (Hz) || 1713.22 +/- 175.05 | + | ||n+1 Measurement (Hz) || 1713.22 +/- 175.05 || 1510.32 +/- 151.17 || 1274.03 +/- 130.1 || 1036.14 +/- 131.17 || 913.17 +/- 105.31 || 793.67 +/- 90.19 |
|- | |- | ||
− | ||Time Correction Factor (Sec) ||720 | + | ||Time Correction Factor (Sec) ||720 || 790 || 780 || 760 || 710 ||730 |
|- | |- | ||
− | ||n+1 Measurement Time Corrected (Hz) ||1980.97 +/- 202.41 | + | ||n+1 Measurement Time Corrected (Hz) ||1980.97 +/- 202.41 || 1771.19 +/- 177.28 || 1491.08 +/- 152.26 || 1207.78 +/- 152.90 || 1053.76 +/- 121.52 || 919.56 +/- 104.50 |
|- | |- | ||
− | ||n Measurement/n+1 Corrected Measurement || 1.01 +/- 0. | + | ||n Measurement/n+1 Corrected Measurement || 0.99 +/- 0.15 || 0.97 +/- 0.14 || 1.01 +/- 0.14 || 1.05 +/- 0.17 || 0.98 +/- 0.17 || 0.99 +/- 0.16 |
|- | |- | ||
|} | |} | ||
Line 773: | Line 773: | ||
|- | |- | ||
||Data File Entry || 6.90 +/- 0.30 || 6.85 +/- 0.22 ||6.73 +/- 0.19 || 6.61 +/- 0.19 || 6.43 +/- 0.19 || 6.23 +/- 0.21 || 6.14 +/- 0.21 | ||Data File Entry || 6.90 +/- 0.30 || 6.85 +/- 0.22 ||6.73 +/- 0.19 || 6.61 +/- 0.19 || 6.43 +/- 0.19 || 6.23 +/- 0.21 || 6.14 +/- 0.21 | ||
+ | |- | ||
+ | |} | ||
+ | |||
+ | |||
+ | Below is a table to check the activities against later measurements by tracing the activity back in time and taking a ratio | ||
+ | |||
+ | {| border="3" cellpadding="5" cellspacing="0" | ||
+ | || || Run 1 || Run 2 || Run 3 || Run 4 || Run 5 || Run 6 | ||
+ | |- | ||
+ | || Original Measurement (Hz) || 994.73 +/- 304.327 || 948.20 +/- 211.33 || 839.55 +/- 163.38 || 738.99 +/- 137.82 || 619.70 +/- 119.20 || 502.25 +/- 106.43 | ||
+ | |- | ||
+ | ||n+1 Measurement (Hz) || 948.20 +/- 211.33 || 839.55 +/- 163.38 ||738.99 +/- 137.82 || 619.70 +/- 119.20 ||502.25 +/- 106.43 || 463.05 +/- 95.31 | ||
+ | |- | ||
+ | ||Time Correction Factor (Sec) ||720 || 755 || 775 || 750 || 750 || 720 | ||
+ | |- | ||
+ | ||n+1 Measurement Time Corrected (Hz) ||1096.39 +/- 244.36 || 977.63 +/- 190.25 || 864.01 +/- 161.14 || 720.90 +/- 138.67 || 584.27 +/- 123.81 || 535.42 +/- 110.21 | ||
+ | |- | ||
+ | ||n Measurement/n+1 Corrected Measurement || 0.91 +/- 0.33 || 0.97 +/- 0.29 || 0.97 +/- 0.26|| 1.02 +/- 0.27 || 1.06 +/- 0.3 || 0.94 +/- 0.28 | ||
|- | |- | ||
|} | |} | ||
==Plots== | ==Plots== | ||
+ | |||
+ | Below is a table for the plots of the cross sample ratios | ||
+ | |||
+ | {| border="3" cellpadding="5" cellspacing="0" | ||
+ | || || Run 1 || Run 2 || Run 3 || Run 4 || Run 5 || Run 6 || Run 7 | ||
+ | |- | ||
+ | || Mix Measurement (Hz) || 994.73 +/- 304.327 || 948.20 +/- 211.33 || 839.55 +/- 163.38 || 738.99 +/- 137.82 || 619.70 +/- 119.2 || 502.25 +/- 106.43 || 463.05 +/- 95.31 | ||
+ | |- | ||
+ | ||Original Pure Se Measurement (Hz) || 1961.89 +/- 216.27 || 1713.22 +/- 175.05 || 1510.32 +/- 151.17 || 1274.03 +/- 130.1 || 1036.14 +/- 131.17 || 913.17 +/- 105.31 || 793.67 +/- 90.19 | ||
+ | |- | ||
+ | ||Time Correction Factor (Sec) ||340 || 340 ||375 || 380 || 390 || 350 || 360 | ||
+ | |- | ||
+ | ||Time Corrected Mix Measurement (Hz) ||1065.33 +/- 325.93 || 1015.50 +/- 226.33 || 905.51 +/- 176.22 || 797.85 +/- 148.80 || 670.41 +/- 128.95 || 538.98 +/- 114.214 || 497.92 +/- 102.49 | ||
+ | |- | ||
+ | || Time Corrected Mix Measurement/Pure Se Measurement || 0.54 +/- 0.18 || 0.59 +/- 0.14 || 0.60 +/- 0.13 || 0.63 +/- 0.13 || 0.65 +/- 0.15 || 0.59 +/- 0.14 || 0.63 +/- 0.15 | ||
+ | |- | ||
+ | |} | ||
Below are the time vs. frequency plots | Below are the time vs. frequency plots | ||
− | [[File:170063 | + | [[File:170063 MixHLPlot 1 11 18.png|200px]] |
− | [[File:170063 | + | [[File:170063 PureSeHLPlot 1 11 18.png|200px]] |
+ | |||
+ | By exponentiating we find that the initial activity of the Mixture is 1056.48 +/- 2.91 Hz, while the pure sample has an initial activity of 2162.90 +/- 5.16 Hz which gives a ratio of | ||
+ | |||
+ | <math> 0.488 \pm 0.002 </math> | ||
To check consistency I would time correct measurements and take a ratio to compare, for example measurement 1 was taken and a 2nd measurement was taken some time later. So correct the rate of the second measurement back to the first measurement by using the radioactive decay equation, then take a ratio of these 2 numbers. | To check consistency I would time correct measurements and take a ratio to compare, for example measurement 1 was taken and a 2nd measurement was taken some time later. So correct the rate of the second measurement back to the first measurement by using the radioactive decay equation, then take a ratio of these 2 numbers. | ||
− | [[File:170063 | + | [[File:170063 Mix RateRatios 1 11 18.png|200px]] |
− | [[File:170063 | + | [[File:170063 PureSe RateRatio 1 11 18.png|200px]] |
Another check to do is to take the ratio of the mixture and the pure selenium sample (time corrected to the same time for each measurement) and see if the ratio here is relatively constant | Another check to do is to take the ratio of the mixture and the pure selenium sample (time corrected to the same time for each measurement) and see if the ratio here is relatively constant | ||
− | [[File:170063 | + | [[File:170063 CrossSample RateRatios 1 11 18.png|200px]] |
− | By exponentiating we find that the initial activity of the Mixture is | + | By exponentiating we find that the initial activity of the Mixture is 1082.15 +/- 174.89 Hz, while the pure sample has an initial activity of 2169.13 +/- 175.01 Hz which gives a ratio of |
− | <math> 0. | + | Thin Window Mean + 2 Sigma:<math> 0.50 \pm 0.09 </math> ,Original Window + 2 Sigma or original window |
We can also check the ratio of the nickel with the selenium mixture. The rate of the nickel foil was <math> 1.127931 \times 10^6 \pm 7.258 \times 10^3 Hz </math> | We can also check the ratio of the nickel with the selenium mixture. The rate of the nickel foil was <math> 1.127931 \times 10^6 \pm 7.258 \times 10^3 Hz </math> | ||
Line 808: | Line 847: | ||
308985.7143 +/- 882.6876 Hz for the pure selenium sample | 308985.7143 +/- 882.6876 Hz for the pure selenium sample | ||
− | |||
− | |||
− | |||
− |
Latest revision as of 19:19, 11 January 2018
In order to try to pin down the ratio of pure selenium to the selenium in the soil, try using a 2 channel window to find the signal. The uncertainty in the number of counts in the signal will be determined by subtracting the counts from the 2 channel window from the counts from a window expanded by 1 standard deviation, then divided by the original number of counts in the 2 channel window. Since the max value is fixed, it does not appear in the stats box, but I have included it in the table. It would seem that widening the gaussian function by half of its standard deviation yields results that agree up to 3 digits after the decimal with the result of widening the gaussian by its standard deviation. This would imply not much has changed between these methods.
Changes Made to table:
12/14: Error in signal changed to
(try to not underestimate error) This includes changing the expanded window to a window expanded off of the 2 channel windowBelow is the table for the Se-Ash Mixture
Below is a plot of the ratio of the activities where the pure selenium sample has been time corrected back to match the run time of the mixture.
Note that this does give the expected result that the ratio between the two samples is statistically the same as 0.5, which is the ratio of the masses. The only problem here is that the error is quite large, The average of the uncertainty divided by the mean is 39.6%.
Below are the half life plots for the Pure Selenium sample and the ash mixture
The slope of the mixture plot gives a half life of 63.48 +/- 26.85 minutes. This does overlap, but the error is extremely large.
The slope of the pure selenium sample plot gives a half life of 56.08 +/- 16.33 minutes. Again this does overlap, but the error is still quite large.
The ratio of the initial activities is 0.49 +/- 0.13. Now this does fall within 1 standard deviation of 0.5, which is the value we want.
Dead Time Corrected
I have shown that by tracing the activity back in time for the pure selenium sample that the activities are the same here LB Thesis Thin Window Analysis, but now I must make sure that this will actually give the ratio that is expected.
Begin with the third measurement that was taken for the pure selenium sample
Pure Se Analysis
This will be using a new method as of 12/14 to find the error in the signal. To do this, find the intrinsic error by
Where the expanded integral value is the integral from the stats box after the 2 channel window has been increased by 2 sigma. This number will then be multiplied by the background subtracted thin window signal. So the error in the signal is
First Pure Se Measurement
Below is the histogram of interest
From this we can see that
and the integrated background is given by
Now we can find the background subtracted signal
Now convert this into an activity by dividing by the runtime
Now we can find the true value using the integrated measurement
So now the true value for the activity is
Now we must correct for the dead time. The third run had an average count rate of 1387.06 Hz, which corresponds to a dead time of 3.06 +/- 0.30
So the real activity for the third measurement of the Pure Se sample is
Second Pure Se Measurement
Below is the histogram of interest
From this we can see that
and the integrated background is given by
Now we can find the background subtracted signal
Now convert this into an activity by dividing by the runtime
Now we can find the true value using the integrated measurement
So now the true value for the activity is
Now we must correct for the dead time. The third run had an average count rate of 1387.06 Hz, which corresponds to a dead time of 2.53 +/- 0.31
So the real activity for the third measurement of the Pure Se sample is
Third Pure Se Measurement
Below is the histogram of interest
From this we can see that
and the integrated background is given by
Now we can find the background subtracted signal
Now convert this into an activity by dividing by the runtime
Now we can find the true value using the integrated measurement
So now the true value for the activity is
Now we must correct for the dead time. The third run had an average count rate of 1387.06 Hz, which corresponds to a dead time of 2.53 +/- 0.3
So the real activity for the third measurement of the Pure Se sample is
Below is a table representing the numbers that will be put into a data file to be linearly fit (Also using the values from the thesis portion)
Fourth Pure Se Measurement
Now do the analysis for the 4th measurement taken on the pure selenium sample.
Below is the histogram of interest
From this we can see that
and the integrated background is given by
Now we can find the background subtracted signal
Now convert this into an activity by dividing by the runtime
Now we can find the true value using the integrated measurement
So now the true value for the activity is
Now we must correct for the dead time. The 4th run had an average count rate of 1274.32 Hz, which corr5esponds to a dead time of 2.26 +/- 0.33
So the real activity for the third measurement of the Pure Se sample is
Fifth Pure Se Measurement
Now do the analysis for the 5th measurement taken on the pure selenium sample.
Below is the histogram of interest
From this we can see that
and the integrated background is given by
Now we can find the background subtracted signal
Now convert this into an activity by dividing by the runtime
Now we can find the true value using the integrated measurement
So now the true value for the activity is
Now we must correct for the dead time. The 5th run had an average count rate of 1168.06 Hz, which corresponds to a dead time of 2.26 +/- 0.33
So the real activity for the third measurement of the Pure Se sample is
Sixth Pure Se Measurement
Now do the analysis for the 6th measurement taken on the pure selenium sample.
Below is the histogram of interest
From this we can see that
and the integrated background is given by
Now we can find the background subtracted signal
Now convert this into an activity by dividing by the runtime
Now we can find the true value using the integrated measurement
So now the true value for the activity is
Now we must correct for the dead time. The 6th run had an average count rate of 1168.06 Hz, which corresponds to a dead time of 1.96 +/- 0.26
So the real activity for the 6th measurement of the Pure Se sample is
Seventh Pure Se Measurement
Now do the analysis for the 7th measurement taken on the pure selenium sample.
Below is the histogram of interest
From this we can see that
and the integrated background is given by
Now we can find the background subtracted signal
Now convert this into an activity by dividing by the runtime
Now we can find the true value using the integrated measurement
So now the true value for the activity is
Now we must correct for the dead time. The 7th run had an average count rate of 1012 Hz, which corresponds to a dead time of 1.96 +/- 0.26
So the real activity for the 7th measurement of the Pure Se sample is
400 <t< 640 sec | 1100 < t < 1360 sec | 1875 < t < 2150 | 2650 < t < 2930 sec | 3400 < t < 3690 sec | 4120 < t < 4400 sec | 4840 < t < 5130 sec | |
.dat file entry | 7.58 +/- 0.11 | 7.45 +/- 0.10 | 7.32 +/- 0.10 | 7.15 +/- 0.10 | 6.94 +/- 0.13 | 6.82 +/- 0.11 | 6.68 +/- 0.11 |
Below is a table for the self comparison of the sample runs
Run 1 | Run 2 | Run 3 | Run 4 | Run 5 | Run 6 | ||
Original Measurement (Hz) | 1961.89 +/- 216.27 | 1713.22 +/- 175.05 | 1510.32 +/- 151.17 | 1274.03 +/- 130.1 | 1036.14 +/- 131.17 | 913.17 +/- 105.31 | |
n+1 Measurement (Hz) | 1713.22 +/- 175.05 | 1510.32 +/- 151.17 | 1274.03 +/- 130.1 | 1036.14 +/- 131.17 | 913.17 +/- 105.31 | 793.67 +/- 90.19 | |
Time Correction Factor (Sec) | 720 | 790 | 780 | 760 | 710 | 730 | |
n+1 Measurement Time Corrected (Hz) | 1980.97 +/- 202.41 | 1771.19 +/- 177.28 | 1491.08 +/- 152.26 | 1207.78 +/- 152.90 | 1053.76 +/- 121.52 | 919.56 +/- 104.50 | |
n Measurement/n+1 Corrected Measurement | 0.99 +/- 0.15 | 0.97 +/- 0.14 | 1.01 +/- 0.14 | 1.05 +/- 0.17 | 0.98 +/- 0.17 | 0.99 +/- 0.16 |
Se-Sage Mix Analysis
Activity of the First Measurement
Below is the histogram of interest
From this we can see that
and the integrated background is given by
Now we can find the background subtracted signal
Now convert this into an activity by dividing by the runtime
Now we can find the true value using the integrated measurement
So now the true value for the activity is
Now we must correct for the dead time. The 1st run had an average count rate of 2434.44 Hz, which corresponds to a dead time of 5.06 +/- 0.39
So the real activity for the 1st measurement of the Mixed sample is
Activity of the Second Measurement
Below is the histogram of interest
From this we can see that
and the integrated background is given by
Now we can find the background subtracted signal
Now convert this into an activity by dividing by the runtime
Now we can find the true value using the integrated measurement
So now the true value for the activity is
Now we must correct for the dead time. The 1st run had an average count rate of 1756.30 Hz, which corresponds to a dead time of 3.06 +/- 0.30
So the real activity for the 2nd measurement of the Mixed sample is
Activity of the Third Measurement
Below is the histogram of interest
From this we can see that
and the integrated background is given by
Now we can find the background subtracted signal
Now convert this into an activity by dividing by the runtime
Now we can find the true value using the integrated measurement
So now the true value for the activity is
Now we must correct for the dead time. The 1st run had an average count rate of 1401.98 Hz, which corresponds to a dead time of 2.53 +/- 0.31
So the real activity for the 2nd measurement of the Mixed sample is
Fourth Mix Sample Measurement
Below is the histogram of interest
From this we can see that
and the integrated background is given by
Now we can find the background subtracted signal
Now convert this into an activity by dividing by the runtime
Now we can find the true value using the integrated measurement
So now the true value for the activity is
Now we must correct for the dead time. The 1st run had an average count rate of 1401.98 Hz, which corresponds to a dead time of 2.53 +/- 0.31
>
So the real activity for the 2nd measurement of the Mixed sample is
Activity of the Fifth Measurement
Below is the histogram of interest
From this we can see that
and the integrated background is given by
Now we can find the background subtracted signal
Now convert this into an activity by dividing by the runtime
Now we can find the true value using the integrated measurement
So now the true value for the activity is
Now we must correct for the dead time. The 1st run had an average count rate of 1025.13 Hz, which corresponds to a dead time of 1.92 +/- 0.26
So the real activity for the 2nd measurement of the Mixed sample is
Sixth Mix Sample Measurement
Below is the histogram of interest
From this we can see that
and the integrated background is given by
Now we can find the background subtracted signal
Now convert this into an activity by dividing by the runtime
Now we can find the true value using the integrated measurement
So now the true value for the activity is
Now we must correct for the dead time. The 1st run had an average count rate of 898.19 Hz, which corresponds to a dead time of 1.65 +/- 0.34
So the real activity for the 2nd measurement of the Mixed sample is
Activity of the 7th Measurement
Below is the histogram of interest
From this we can see that
and the integrated background is given by
Now we can find the background subtracted signal
Now convert this into an activity by dividing by the runtime
Now we can find the true value using the integrated measurement
So now the true value for the activity is
Now we must correct for the dead time. The 1st run had an average count rate of 846.40 Hz, which corresponds to a dead time of 1.65 +/- 0.34
So the real activity for the 2nd measurement of the Mixed sample is
0<t<300 | 730 <t< 1020 | 1480<t<1775 | 2250<t<2550 | 3050<t<3300 | 3775<t<4050 | 4480<t<4770 | |
Data File Entry | 6.90 +/- 0.30 | 6.85 +/- 0.22 | 6.73 +/- 0.19 | 6.61 +/- 0.19 | 6.43 +/- 0.19 | 6.23 +/- 0.21 | 6.14 +/- 0.21 |
Below is a table to check the activities against later measurements by tracing the activity back in time and taking a ratio
Run 1 | Run 2 | Run 3 | Run 4 | Run 5 | Run 6 | |
Original Measurement (Hz) | 994.73 +/- 304.327 | 948.20 +/- 211.33 | 839.55 +/- 163.38 | 738.99 +/- 137.82 | 619.70 +/- 119.20 | 502.25 +/- 106.43 |
n+1 Measurement (Hz) | 948.20 +/- 211.33 | 839.55 +/- 163.38 | 738.99 +/- 137.82 | 619.70 +/- 119.20 | 502.25 +/- 106.43 | 463.05 +/- 95.31 |
Time Correction Factor (Sec) | 720 | 755 | 775 | 750 | 750 | 720 |
n+1 Measurement Time Corrected (Hz) | 1096.39 +/- 244.36 | 977.63 +/- 190.25 | 864.01 +/- 161.14 | 720.90 +/- 138.67 | 584.27 +/- 123.81 | 535.42 +/- 110.21 |
n Measurement/n+1 Corrected Measurement | 0.91 +/- 0.33 | 0.97 +/- 0.29 | 0.97 +/- 0.26 | 1.02 +/- 0.27 | 1.06 +/- 0.3 | 0.94 +/- 0.28 |
Plots
Below is a table for the plots of the cross sample ratios
Run 1 | Run 2 | Run 3 | Run 4 | Run 5 | Run 6 | Run 7 | |
Mix Measurement (Hz) | 994.73 +/- 304.327 | 948.20 +/- 211.33 | 839.55 +/- 163.38 | 738.99 +/- 137.82 | 619.70 +/- 119.2 | 502.25 +/- 106.43 | 463.05 +/- 95.31 |
Original Pure Se Measurement (Hz) | 1961.89 +/- 216.27 | 1713.22 +/- 175.05 | 1510.32 +/- 151.17 | 1274.03 +/- 130.1 | 1036.14 +/- 131.17 | 913.17 +/- 105.31 | 793.67 +/- 90.19 |
Time Correction Factor (Sec) | 340 | 340 | 375 | 380 | 390 | 350 | 360 |
Time Corrected Mix Measurement (Hz) | 1065.33 +/- 325.93 | 1015.50 +/- 226.33 | 905.51 +/- 176.22 | 797.85 +/- 148.80 | 670.41 +/- 128.95 | 538.98 +/- 114.214 | 497.92 +/- 102.49 |
Time Corrected Mix Measurement/Pure Se Measurement | 0.54 +/- 0.18 | 0.59 +/- 0.14 | 0.60 +/- 0.13 | 0.63 +/- 0.13 | 0.65 +/- 0.15 | 0.59 +/- 0.14 | 0.63 +/- 0.15 |
Below are the time vs. frequency plots
By exponentiating we find that the initial activity of the Mixture is 1056.48 +/- 2.91 Hz, while the pure sample has an initial activity of 2162.90 +/- 5.16 Hz which gives a ratio of
To check consistency I would time correct measurements and take a ratio to compare, for example measurement 1 was taken and a 2nd measurement was taken some time later. So correct the rate of the second measurement back to the first measurement by using the radioactive decay equation, then take a ratio of these 2 numbers.
Another check to do is to take the ratio of the mixture and the pure selenium sample (time corrected to the same time for each measurement) and see if the ratio here is relatively constant
By exponentiating we find that the initial activity of the Mixture is 1082.15 +/- 174.89 Hz, while the pure sample has an initial activity of 2169.13 +/- 175.01 Hz which gives a ratio of
Thin Window Mean + 2 Sigma:
,Original Window + 2 Sigma or original windowWe can also check the ratio of the nickel with the selenium mixture. The rate of the nickel foil was
So using the efficiency corrected rate for the soil, we find that the ratio is
The measurement was made 20cm away from the detector face, which has an efficiency of 0.70 +/- 0.0011, so the measurements with the efficiency taken into account are
150925.7143 +/- 392.5923 Hz for the mixture and
308985.7143 +/- 882.6876 Hz for the pure selenium sample