Difference between revisions of "Dose Calculations and 137Cs Shot Log"

From New IAC Wiki
Jump to navigation Jump to search
Line 80: Line 80:
 
|-
 
|-
 
|}
 
|}
 
QC OSLs Averages
 
OSL # Listed Dose (mRad) Measured Dose (mRad) Average Dose (mRad)
 
DN08264870R Unexposed 69.36198 \
 
DN082701640 Unexposed 77.25963 74.74153667
 
DN08269802P Unexposed 77.603 /
 
DN10712168C 1000 1235.238 \
 
DN10309170Q 1000 1223.106 1226.119667
 
DN102095379 1000 1220.015 /
 
 
  
 +
{| border="2" style="text-align:center;" |cellpadding="20" cellspacing="0
 +
|-
 +
|OSL # || Listed Dose (mRad) || Measured Dose (mRad) || Average Dose (mRad)
 +
|-
 +
|DN08264870R || Unexposed || 69.36198 || \
 +
|-
 +
|DN082701640 || Unexposed || 77.25963 || 74.74153667
 +
|-
 +
|DN08269802P || Unexposed || 77.603 || /
 +
|-
 +
|DN10712168C || 1000 || 1235.238 || \
 +
|-
 +
|DN10309170Q || 1000 || 1223.106 || 1226.119667
 +
|-
 +
|DN102095379 || 1000 || 1220.015 | /
 +
|-
 +
|}
  
  

Revision as of 19:11, 16 October 2017

I need to change the exposed doses by changing from seconds to hours. Unit problems.

Source Calibration and Dose Rate Data

Date Initial Source Strength (Ci) Current Source Strength (Ci) Distance to Faceplate (cm) Exposure Rate (Roentgen/hr)°
6/27/17 20 9.3 30.5 17.691
9/11/17 20 9.3 30.5 17.606

°Calculated by TSO using [math]\dot R = \frac {\Gamma A }{ D^2} [/math] where [math]\boldsymbol{\Gamma} = 3.3 \frac {(cm^2)(R)}{(mCi)(hr)} = .35 \frac {(m^2)(R)}{(Ci)(hr)} [/math] , [math]\mathbf{A} = 9.3 Ci [/math] , [math]\mathbf{D}=(\text{Distance to faceplate} + 11.2cm) [/math]

There is an adjustment distance of 11.2cm which accounts for the distance from the faceplate to the [math] ^{137}Cs[/math] source

137Cs Shot Log

Dot OSLs

Date Run Number Run Time (s) Distance to Faceplate Number of OSLs in Stack Dose Rate (R/hr) Total Exposed Dose (R)
9/8/17 1 5660 30cm 11 18.7188 108536
9/8/17 1 5660 30cm 11 18.7188 108536
9/26/17 2 4500 30cm 11 18.7188 84234.8
9/26/17 2 4500 30cm 11 18.7188 84234.8

Total dose calculated by [math]\int\limits_{t_0}^{t_f}\dot R\ dt [/math]

In these calculations, [math]\Gamma = 0.35 \frac {(m^2)(R)}{(Ci)(hr)} [/math]

Nanodot OSL Quality Control

Date Run Number Run Time (s) Distance to Faceplate Number of OSLs Total Exposed Dose (R)
10/12/17 1 284.04 30.5cm 3 5242.02

Calibration OSLs

OSL # Listed Dose (mRad) Measured Dose (mRad) Average Dose (mRad)
DN08276339J Unexposed 71.19332 \
DN08276683L Unexposed 70.39211 72.10899
DN08258554P Unexposed 74.74154 /
DN10806640H 500 687.3246 \
DN10608908Z 500 681.0294 683.2422333
DN10806620J 500 681.3727 /
DN105057582 3000 3563.558 \
DN104085766 3000 3522.352 3567.258333
DN10608909X 3000 3615.865 /
DN108066085 50000 55966.55 \
DN103091798 50000 55420.36 55622.12
DN104086665 50000 55479.45 /
DN10606601L 100000 111610.5 \
DN106066095 100000 114271.2 114191.3333
DN10712163M 100000 116692.3 /
OSL # Listed Dose (mRad) Measured Dose (mRad) Average Dose (mRad)
DN08264870R Unexposed 69.36198 \
DN082701640 Unexposed 77.25963 74.74153667
DN08269802P Unexposed 77.603 /
DN10712168C 1000 1235.238 \
DN10309170Q 1000 1223.106 1226.119667
DN102095379 1000 /


Total dose calculated by [math]\int\limits_{t_0}^{t_f}\dot R\ dt [/math]



Reproducibility with 137Cs