Difference between revisions of "Dose Calculations and 137Cs Shot Log"

From New IAC Wiki
Jump to navigation Jump to search
Line 31: Line 31:
 
Total dose calculated by <math>\int\limits_{t_0}^{t_f}\dot R\ dt </math>
 
Total dose calculated by <math>\int\limits_{t_0}^{t_f}\dot R\ dt </math>
  
In these calculations,  
+
In these calculations, <math>\Gamma = 0.35 \frac {(m^2)(R)}{(Ci)(hr)} </math>
  
 
==Nanodot OSL Quality Control==
 
==Nanodot OSL Quality Control==

Revision as of 22:20, 12 October 2017

Source Calibration and Dose Rate Data

Date Initial Source Strength (Ci) Current Source Strength (Ci) Distance to Faceplate (cm) Exposure Rate (Roentgen/hr)°
6/27/17 20 9.3 30.5 17.691
9/11/17 20 9.3 30.5 17.606

°Calculated by TSO using [math]\dot R = \frac {\Gamma A }{ D^2} [/math] where [math]\boldsymbol{\Gamma} = 3.3 \frac {(cm^2)(R)}{(mCi)(hr)} [/math] , [math]\mathbf{A} = 9.3 Ci [/math] , [math]\mathbf{D}=(\text{Distance to faceplate} + 11.2cm) [/math]

There is an adjustment distance of 11.2cm which accounts for the distance from the faceplate to the [math] ^{137}Cs[/math] source

137Cs Shot Log

Dot OSLs

Date Run Number Run Time (s) Distance to Faceplate Number of OSLs in Stack Dose Rate (R/hr) Total Exposed Dose (R) Measured Dose Adjusted Dose
9/8/17 1 5660 30cm 11 18.7188 108536 - -
9/8/17 1 5660 30cm 11 18.7188 108536 - -
9/26/17 2 4500 30cm 11 18.7188 - - -
9/26/17 2 4500 30cm 11 18.7188 - - -

Total dose calculated by [math]\int\limits_{t_0}^{t_f}\dot R\ dt [/math]

In these calculations, [math]\Gamma = 0.35 \frac {(m^2)(R)}{(Ci)(hr)} [/math]

Nanodot OSL Quality Control

Date Run Number Run Time (s) Distance to Faceplate Number of OSLs Total Exposed Dose (R) Measured Dose Adjusted Dose
10/12/17 1 284.04 30.5cm 3 5242.02 - -

Total dose calculated by [math]\int\limits_{t_0}^{t_f}\dot R\ dt [/math]



Reproducibility with 137Cs