Difference between revisions of "05-2-2017"
Jump to navigation
Jump to search
(10 intermediate revisions by the same user not shown) | |||
Line 1: | Line 1: | ||
[[2nCor 44| go_back]] | [[2nCor 44| go_back]] | ||
+ | '''The He-3 detector output from the spec amp to the ADC had the wrong polarity''' | ||
''' Electronics map for the two auxiliary scintillators ''' | ''' Electronics map for the two auxiliary scintillators ''' | ||
Line 11: | Line 12: | ||
|} | |} | ||
− | |||
− | |||
− | |||
He-3 timing is in patch panel 223A16, & energy is 223A17. | He-3 timing is in patch panel 223A16, & energy is 223A17. | ||
Line 25: | Line 23: | ||
Small aux. scintillators have a CFD threshold of 0.550 V for run 6274 and later. | Small aux. scintillators have a CFD threshold of 0.550 V for run 6274 and later. | ||
+ | |||
+ | |||
+ | =run list= | ||
+ | |||
{| border="1" | {| border="1" | ||
− | ! Run # !!Charge/pulse !!Target!! Configuration !! Individual histos + rates !! overlayed histos !! | + | ! Run # !!Charge/pulse !!Target!! Configuration !! Individual histos + rates !! overlayed histos !! neutron coincidences per pulse !! n-coin error !! neutron coincidences per pulse per Amp !! doubles/singles |
|- | |- | ||
− | | 6272|| 18 pV*s || D2O'''|| 7.5 MeV , 3 ns FWHM pulse width , 1.3 A Peak , 180.4 Hz , || || | + | | 6272|| 18 pV*s || D2O'''|| 7.5 MeV , 3 ns FWHM pulse width , 1.3 A Peak , 180.4 Hz , || || || singles: 3.63E-03 doubles: 6.69E-06 || singles: +- 4% doubles: +-100% || singles/A: 2.59E-03 doubles/A: 4.78E-06 || doubles/single: 1.85E-03 |
|- | |- | ||
| 6273|| 21 pV*s || ''' 0.5mm x 2cm x 4cm Th #170113000''' || 7.5 MeV , 3 ns FWHM pulse width , 1.3 A Peak , 180.4 Hz , small Scint at .725 m and 1.45 m, teed off Beam current monitor changing pulse size from 21 pV*s to 14 pV*s , inserted beam current into PADC ch 6 20 minutes into run|| || || | | 6273|| 21 pV*s || ''' 0.5mm x 2cm x 4cm Th #170113000''' || 7.5 MeV , 3 ns FWHM pulse width , 1.3 A Peak , 180.4 Hz , small Scint at .725 m and 1.45 m, teed off Beam current monitor changing pulse size from 21 pV*s to 14 pV*s , inserted beam current into PADC ch 6 20 minutes into run|| || || | ||
Line 43: | Line 45: | ||
| 6279||12 pV*s || 1" diam DU 111108007 || '''9.0 MeV''' , 3 ns FWHM pulse width , 1.3 A Peak , 180.4 Hz , small Scint at .725 m and 1.45 m, added 2" lead on beam right side of magnet || || [[File:R6276-R6279Overlay.png|200px]] || | | 6279||12 pV*s || 1" diam DU 111108007 || '''9.0 MeV''' , 3 ns FWHM pulse width , 1.3 A Peak , 180.4 Hz , small Scint at .725 m and 1.45 m, added 2" lead on beam right side of magnet || || [[File:R6276-R6279Overlay.png|200px]] || | ||
|- | |- | ||
− | | 6280||11.5 pV*s || 1" diam DU 111108007 || '''10.6 MeV''' , 3 ns FWHM pulse width , 1.4 A Peak , 180.4 Hz , small Scint at .725 m and 1.45 m, added 2" lead on beam right side of magnet || || [[File:R6279-R6280Overlay.png|200px]] || singles: | + | | 6280||11.5 pV*s || 1" diam DU 111108007 || '''10.6 MeV''' , 3 ns FWHM pulse width , 1.4 A Peak , 180.4 Hz , small Scint at .725 m and 1.45 m, added 2" lead on beam right side of magnet || || [[File:R6279-R6280Overlay.png|200px]] ||singles: 3.96E-02 doubles: 7.49E-04 || singles: +- 1% doubles: +-8% || singles/A: 2.83E-02 doubles/A: 5.35E-04 || doubles/single: 1.89E-02 |
|- | |- | ||
− | | 6282||11.5 pV*s || '''0.5 mm x 4 cm x 2 cm DU-238 R45020''' || 10.6 MeV , 3 ns FWHM pulse width , 1.4 A Peak , 180.4 Hz , small Scint at .725 m and 1.45 m, added 2" lead on beam right side of magnet || || [[File:R6280-R6282Overlay.png|200px]] || | + | | 6282||11.5 pV*s || '''0.5 mm x 4 cm x 2 cm DU-238 R45020''' || 10.6 MeV , 3 ns FWHM pulse width , 1.4 A Peak , 180.4 Hz , small Scint at .725 m and 1.45 m, added 2" lead on beam right side of magnet || || [[File:R6280-R6282Overlay.png|200px]] || singles: 4.87E-03 doubles: 2.47E-05 || singles: +- 1% doubles: +-24% || singles/A: 3.48E-03 doubles/A: 1.77E-05 || doubles/single: 5.07E-03 |
|- | |- | ||
| 6283||11.5 pV*s || 0.5 mm x 4 cm x 2 cm DU-238 R45020 || '''HV tripped around pulse 498,000''' ; 10.6 MeV , 3 ns FWHM pulse width , 1.4 A Peak , 180.4 Hz , '''Removed small Scint at .725 m Kept Scint in at 1.45 m''', || || Histograms only includes pulses with HV on. [[File:R6283.png|200px]] || | | 6283||11.5 pV*s || 0.5 mm x 4 cm x 2 cm DU-238 R45020 || '''HV tripped around pulse 498,000''' ; 10.6 MeV , 3 ns FWHM pulse width , 1.4 A Peak , 180.4 Hz , '''Removed small Scint at .725 m Kept Scint in at 1.45 m''', || || Histograms only includes pulses with HV on. [[File:R6283.png|200px]] || | ||
Line 55: | Line 57: | ||
+ | =Analysis= | ||
+ | [[Analysis_of_may_2017_runs]] | ||
Latest revision as of 00:12, 16 June 2017
The He-3 detector output from the spec amp to the ADC had the wrong polarity
Electronics map for the two auxiliary scintillators
Phi Angle (Degrees) | PMT distance | Signal cable label | HV cable label | HV channel on Lecroy Mainframe | CFD CH | VNIM-ECL channel | ROOT variable |
---|---|---|---|---|---|---|---|
~282 | m | 102B1.2 | shv1.2 | 5:L1.2 | 1 | (left) B-0 | evt.TDC[1][25] |
~282 | m | 282T1.7 | shv1.7 | 5:L1.7 | 2 | (left) B-1 | evt.TDC[1][26] |
He-3 timing is in patch panel 223A16, & energy is 223A17.
High voltage & discriminator settings:
30T: 1450;30 , 30B:1450;30 , 54T: 1500;30 , 54B: 1500;30 , 78T:1500;30 , 78B: 1525;30 , 126T: 1500;30 , 126B: 1500;30 , 150T: 1500;40 , 150B: 1500;40 , 210T: 1500;40 , 210B: 1500;40 , 234T: 1500;30 , 234B: 1500;30 , 258T: 1350;30, 258B: 1450;30 , 306T: 1400;30 , 306B: 1400;30 , 330T: 1450;30 , 330B: 1500;30
Radiator 1.5" Al.
Slits currently at 500 => 19 mm, at 75 =0, at 1000=38 mm, readback number is linear with slit width.
Small aux. scintillators have a CFD threshold of 0.550 V for run 6274 and later.
run list
Analysis