Both PMT HV were set to 1100 volts and the LED threshold was at 20 mV
 
Both PMT HV were set to 1100 volts and the LED threshold was at 20 mV
  
==run 4518 terminstors in==
+
==run 4518 terminators in==
  
 
Run 4518 had the bottom PMT of Detector 4 as the trigger and the terminators in.
 
Run 4518 had the bottom PMT of Detector 4 as the trigger and the terminators in.
Line 26: Line 26:
  
 
  R1DC->Draw("(evt.TDC1190[1][2]-evt.TDC1190[2][2])/10>>(4096,-4096,4096)"
 
  R1DC->Draw("(evt.TDC1190[1][2]-evt.TDC1190[2][2])/10>>(4096,-4096,4096)"
==run 4519 terminstors out==
+
==run 4519 terminators out==
  
 
Run 4519 had the bottom PMT of Detector 4 as the trigger and the terminators out.
 
Run 4519 had the bottom PMT of Detector 4 as the trigger and the terminators out.
Line 32: Line 32:
 
The TOP PMT DID have mutliple peaks
 
The TOP PMT DID have mutliple peaks
  
 +
[[File:r4519_Det4Top_1-2hitdiff.png| 200 px]]
  
 
[[2nCor_44#Run_Plan]]
 
[[2nCor_44#Run_Plan]]

Revision as of 17:09, 10 April 2015

I did some TDC ring tests on Detector 4 by adding a 50 Ohm pass through terminator on the base of the top detector and another pass through terminator at the exit of the patch pannel


I needed recommission the DAQ TDC logic which seems to have become unplugged.

I added connectors so TDC1190[6] would have two hits separated by 400 ns

replay the data on daq2 with the command

CODA/CODAreader/ROOT_V5.30/R1DC/evio2nt -fr4516.dat > /dev/null

R1DC->Draw("(evt.TDC1190[1][3])");

R1DC->Draw("(evt.TDC1190[1][6]-evt.TDC1190[2][6])/10>>(4096,-4096,4096));


Both PMT HV were set to 1100 volts and the LED threshold was at 20 mV

run 4518 terminators in

Run 4518 had the bottom PMT of Detector 4 as the trigger and the terminators in.

The TOP PMT did not have mutliple peaks

R1DC->Draw("(evt.TDC1190[1][3]-evt.TDC1190[2][3])/10>>(4096,-4096,4096)");
R1DC->Draw("(evt.TDC1190[1][2]-evt.TDC1190[2][2])/10>>(4096,-4096,4096)"

run 4519 terminators out

Run 4519 had the bottom PMT of Detector 4 as the trigger and the terminators out.

The TOP PMT DID have mutliple peaks

R4519 Det4Top 1-2hitdiff.png